|
|
||||||||
J Neurophysiol (December 1, 2002). 10.1152/jn.01050.2001
Submitted on 26 December 2001
Accepted on 12 August 2002
1Department of Mechanical and Aerospace Engineering and Center for Biomedical Engineering, University of California, Irvine, 92697-3975; 2Department of Physiological Sciences and Brain Research Institute, University of California, Los Angeles, 90095-1760; and 3Department of Kinesiology and Nutritional Science, California State University, Los Angeles, California 90032-8162
| |
ABSTRACT |
|---|
|
|
|---|
Timoszyk, W. K., R. D. de Leon, N. London, R. R. Roy, V. R. Edgerton, and D. J. Reinkensmeyer. The Rat Lumbosacral Spinal Cord Adapts to Robotic Loading Applied During Stance. J. Neurophysiol. 88: 3108-3117, 2002. Load-related afferent information modifies the magnitude and timing of hindlimb muscle activity during stepping in decerebrate animals and spinal cord-injured humans and animals, suggesting that the spinal cord mediates load-related locomotor responses. In this study, we found that stepping on a treadmill by adult rats that received complete, midthoracic spinal cord transections as neonates could be altered by loading the hindlimbs using a pair of small robotic arms. The robotic arms applied a downward force to the lower shanks of the hindlimbs during the stance phase and measured the position of the lower shank during stepping. No external force was applied during the swing phase of the step. When applied bilaterally, this stance force field perturbed the hindlimb trajectories so that the ankle position was shifted downward during stance. In response to this perturbation, both the stance and step cycle durations decreased. During swing, the hindlimb initially accelerated toward the normal, unperturbed swing trajectory and then tracked the normal trajectory. Bilateral loading increased the magnitude of the medial gastrocnemius electromyographic (EMG) burst during stance and increased the amplitude of the semitendinosus and rectus femoris EMG bursts. When the force field was applied unilaterally, stance duration decreased in the loaded hindlimb, while swing duration was decreased in the contralateral hindlimb, thereby preserving interlimb coordination. These results demonstrate the feasibility of using robotic devices to mechanically modulate afferent input to the injured spinal cord during weight-supported locomotion. In addition, these results indicate that the lumbosacral spinal cord responds to load-related input applied to the lower shank during stance by modifying step timing and muscle activation patterns, while preserving normal swing kinematics and interlimb coordination.
| |
INTRODUCTION |
|---|
|
|
|---|
Load-related
sensory information plays an important role in shaping the muscle
activity patterns during locomotion (Dietz and Duysens
2000
; Duysens et al. 2000
; Harkema et al.
1997
; McCrea 2001
; Pearson 1993
).
In both humans and cats, with and without intact cortical control,
loading of the limb or whole body modulates locomotor activity in two
ways. First, loading modulates the magnitude of extensor muscle
activity during stance. For example, unloading a single hindlimb of a
decerebrate cat during treadmill stepping by allowing the hindlimb to
step into a hole reduces average ankle extensor amplitude during the
500-ms period after toe-down by 70% (Hiebert and Pearson
1999
). Reloading of the ankle extensors with a cable-driven
mechanism while the cats stepped into a hole restores the activity of
those muscles to near normal levels. Mechanically unloading ankle
extensors in humans with a cable-driven device also substantially
reduces extensor muscle activity (Sinkjaer et al. 2000
),
while adding load at the center of mass during stance enhances extensor
muscle activity (Stephens and Yang 1999
). In spinal
cord-injured humans, reducing body weight support while stepping at
selected speeds on a treadmill can significantly increase the magnitude
of ankle plantarflexor electromyographic (EMG) amplitudes, suggesting
that the human lumbosacral spinal cord responds to the loading
conditions by generating efferent output for stepping (Dietz et
al. 1998
; Harkema et al. 1997
; Kojima et
al. 1999
). Consistent with these findings, electrical
stimulation of load-related afferents in decerebrate cats during stance
powerfully enhances ipsilateral extensor EMG activity (see review
Pearson et al. 1998
). Second, loading alters the timing
of the transition between extensor and flexor muscle activity, with the
onset of swing typically being delayed when the load is increased. For
example, adding load to human infants during stepping prolongs stance
(Pang and Yang 2000
; Yang et al. 1998
),
although a similar prolongation was not seen in adults (Stephens
and Yang 1999
). Electrical stimulation of load-related
afferents consistently prolongs extensor EMG activity during stance
(Pearson et al. 1998
). Such load-related responses have
been hypothesized to play a key functional role in locomotor control in
that they dynamically adjust the magnitude and timing of muscular
activity to compensate for step-by-step variations in loading.
The presence of load-related locomotor responses in decerebrate
animals, spinal-injured humans, and human infants suggests that the
spinal cord itself plays a significant role in mediating the responses.
The purpose of this study was to characterize the capacity of the
lumbosacral spinal cord, isolated from supraspinal control, to respond
to load-related sensory input. Specifically, we applied loads to the
lower shank of the hindlimbs of rats that received complete
midthoracic spinal cord transections as neonates during bipedal
stepping on a treadmill to determine if the spinal cord could adapt to
a repetitively applied load on the hindlimbs. A unique feature of the
experiment, in addition to defining load-related responses in animals
spinally transected as neonates, was the use of small, lightweight
robot arms to apply loads just above the ankle of the hindlimbs as well
as to record the hindlimb trajectories during stepping
(Reinkensmeyer et al. 2000
; London et al.
2000; Edgerton et al. 2001
; Timoszyk et
al. 2002
). This technique allowed us to quantify the effects of
load-related sensory input on a broad array of features of stepping
including step timing, hindlimb kinematics, and interlimb coordination
across thousands of steps. Our findings indicate that the lumbosacral
spinal cord responds to additional loading by modifying step timing and
muscle activation patterns, while actively maintaining near-normal
swing kinematics and interlimb coordination.
| |
METHODS |
|---|
|
|
|---|
Animals
The experiments were performed on nine rats that
were completely transected at a midthoracic level 5 days after birth.
The transected rat pups were returned to their mothers until they reached 21 days of age. At the time of the first and second series of
experiments (see following text), the rats were 2 to 3 months old and
could perform alternating, weight-bearing, bipedal stepping on the
plantar surfaces of the paws on a treadmill. Six months after these
experiments, two rats were implanted surgically with fine wire
intramuscular electrodes bilaterally for a third series of experiments.
The medial gastrocnemius (MG, a knee flexor and ankle plantarflexor)
and tibialis anterior (TA, an ankle dorsiflexor) were implanted in one
rat and the semitendinosus (ST, a hip extensor and knee flexor) and
rectus femoris (RF, a hip flexor and knee extensor) in the other rat as
described previously (de Leon et al. 1994
; Roy et
al. 1991
). EMG recordings were made after 21 days of recovery
from the implant surgery. The surgical procedures and maintenance of
the spinal rats followed the protocols detailed for cats in Roy
et al. (1992)
. All procedures used in this study were approved
by the Animal Use Committee at University of California at Los Angeles
and followed the American Physiological Society Animal Care Guidelines.
Robotic apparatus
The rats stepped bipedally on a motorized,
variable-speed treadmill with a robot arm attached to each of the lower
shanks just above the ankle (Fig. 1)
(Edgerton et al. 2001
; Reinkensmeyer et al.
2000
; Timoszyk et al. 2002
). The robots were
attached with padded neoprene straps (3 × 30 mm) that were
wrapped around the lower shanks. Both ends of the straps were then
inserted into alligator clips mounted at the end of the robot arms.
Each clip rotated freely in the parasagittal plane through a revolute,
ball-bearing joint.
|
The robotic arms were PHANToM 1.0 haptic interfaces (SensAble Technologies, Inc., Woburn, MA). These arms are small, cable-driven mechanical linkages that provide high-fidelity position sensing and force control with three degrees of freedom. The robot manufacturer's software development kit, the General Haptic Open Software Toolkit (GHOST SDK 2.1), was used to implement control laws for the robots.
The static friction of the robot arms in the horizontal and vertical planes was approximately 0.14 and 0.05 N, respectively. To reduce the effects of this friction, the robots applied a force equal to 80% of the estimated friction in the direction of motion of the robot endpoint during all experiments. The robots were mechanically counterbalanced so that they did not gravitationally load the hindlimbs.
A counterbalanced body weight support system was used to partially unload the hindlimbs and to stabilize the rats during stepping (Fig. 1). The rats were attached via a harness to one end of a lever arm 30 cm from the pivot, and a counterweight was positioned at the other end of the lever arm. The amount of body weight support could be adjusted by changing the mass of the counterweight or by moving it relative to the pivot. The harness could be rotated about a ball joint and locked to allow adjustment of the rats' orientation relative to the treadmill.
Experimental protocol
To load the rats' hindlimbs during stance, the robot arms were programmed to apply a downward force at the ankle cuff that was proportional to the backward velocity of the robot tip (Fig. 2). The force was applied only when the hindlimb was moving backward. The advantage of using such a velocity-dependent force field to load the hindlimb was that the downward force was applied primarily during the stance phase yet was initiated and terminated smoothly as the hindlimb reversed horizontal direction at the beginning and end of stance.
|
Three series of experiments were performed using this force field perturbation. In the first series, the force field was applied to both hindlimbs (number of rats = 8, gain = 4.0 mN/mm/s, mean downward force = 8% of body weight). In the second series, the same force field was applied to a single hindlimb (side randomized, number of rats = 9). The third series was performed following the EMG electrode implantations: the bilateral force field experiment (first series), with the addition of the EMG recordings, was repeated in two rats.
Each rat was trained to step on the treadmill with the robots attached for approximately 3 min/day, for 3 days before each experiment. Because the rats received spinal cord transections as neonates, this training was sufficient for the rats to recover the ability to perform partial body weight-supported stepping. During the period between the second and third series of experiments, the rats were not trained. To facilitate the best stepping in each rat during the experiments, an experienced animal trainer adjusted the amount of body weight support and the orientation of the harness. The amount of support provided was 75 ± 8% (mean ± SD) of the body weight. The treadmill speed was set at 11 cm/s.
To assess the effects of the stance force field, the force field was applied and removed alternately while the rats stepped. The force field was activated or deactivated every 25 steps with a software switch until 10 bouts in the force-field-off condition and 10 bouts in the force-field-on condition were recorded (a total of 500 steps). For the experiments in the third series when EMG was recorded, five force-field-on and five force-field-off bouts were recorded. In all of the experiments, if the rat stopped stepping for more than 5 consecutive seconds, the experimenter briefly raised the rat using the weight-support lever (Fig. 1) and then lowered the rat to the treadmill. Typically, this would induce the rat to reinitiate stepping.
Data collection and analysis
The three-dimensional coordinates of the attachment point of the
robot arms to the shanks and the force applied by each robot arm to the
shank were sampled at 100 Hz. EMG recordings were sampled at 2 kHz.
During all experiments, video recordings of the hindlimb movements were
taken in the sagittal plane at 30 frames/s. Robotic, EMG, and video
recordings were synchronized by sampling the video time code generator
output. Details of the video and EMG data acquisition and analyses have
been described previously (see de Leon et al. 1998
).
Step detector
To identify individual steps from the robotic motion capture data, the following algorithm was used (Fig. 3). First, the horizontal trajectory of each hindlimb (Fig. 3, solid line) was low-pass filtered (forward and backward to prevent phase lag) with a ninth-order Butterworth filter with a cutoff frequency of 1.5 Hz (roughly 1.5 times the mean step frequency) (Fig. 3A, dashed line). This filtering procedure removed high-frequency variations in the horizontal trajectory, typically leaving a single minimum point for each step. These local minima were identified (Fig. 3A, squares), and the minimum of the unfiltered, horizontal trajectory within a 500-ms interval centered about each filtered minimum was defined as the beginning of a step cycle (Fig. 3, A-D, circles). The change from positive to negative horizontal velocity was defined as the beginning of stance (Fig. 3, A-D, ×). Thus the swing and stance durations of each identified step were calculated as the duration of time during which the horizontal velocity was positive or negative, respectively. The length of each step was calculated as the difference between the maximum and minimum of the horizontal trajectory of the robot tip within a step (Fig. 3D). Note that, following the step identification procedure, the unfiltered position data were used to calculate the step parameters, as well as for all other analyses and presentations.
|
The identified step durations and step lengths were approximately normally distributed. Because the step detection algorithm did not distinguish actual steps from short flexion movements, large withdrawal movements, or step resetting by the experimenter, very short and very long steps, as well as steps with very small step lengths, were discarded from further analysis. The step duration (mean ± SD) was 980 ± 270 ms and the results presented below excluded steps whose durations fell above or below two SDs from the mean. Step length was 38.0 ± 17.6 mm with step lengths shorter than two SDs from the mean excluded from the analyses. Steps that did not contain either a swing or a stance phase with at least a duration of 90 ms were also discarded. Based on the above criteria, 7.6 ± 4.1% (mean ± SD) and 8.3 ± 1.7% of steps were omitted from each rat in the force-field-on and force-field-off conditions from bilateral force field experiments, and 5.6 ± 1.2 and 5.3 ± 0.8% were omitted from unilateral force field experiments. An analysis was performed to check the sensitivity of the results to the percentage of discarded steps based on step cycle duration only. No changes in statistical significance were observed when between 0 and 60% of the shortest and longest steps were discarded.
Mean trajectory/velocity profiles
To calculate mean step trajectory profiles, each identified step was normalized with respect to time, and the normalized trajectory was resampled with 100 points (cubic spline interpolation). The mean step trajectory then was calculated as the ensemble average of the resampled steps. This resampling procedure allowed the individual step trajectories to be averaged even though the trajectories contained a varying number of points. This procedure yielded trajectories that accurately represented the spatial path of the hindlimbs. Mean step trajectories were calculated for the force-field-on and force-field-off conditions for each rat and averaged for each condition across rats.
The velocity of the shank attachment point was calculated by taking the finite difference of the raw position trajectories. Mean step velocity profiles then were calculated with the same resampling method used to calculate the mean step trajectories. Note that the velocity profiles were resampled after finite differentiation of the position trajectories to preserve the measured velocity information despite the time normalization.
Step length, step cycle duration, swing duration, and stance duration were quantified for both the force-field-on and force-field-off conditions for both hindlimbs of each rat. The mean of each of these parameters then was calculated across all rats for each hindlimb and for each condition.
EMG analysis
EMG activity that corresponded to weight bearing, plantar surface stepping was analyzed in two rats. For these rats, an observer blinded to the robotic motion analysis reviewed video data to identify stepping bouts during which alternating plantar stepping occurred. Data depicting poor stepping were discarded. For the rat with hip/knee muscle (RF and ST) electrodes implanted, 20% of the data were eliminated and for the rat with knee/ankle muscle (MG and TA) electrodes 12% of the data were eliminated.
Onsets of EMG bursts were defined as the times at which the rectified
and filtered (cutoff 400 Hz) EMG signal exceeded the threshold voltage
(mean of the baseline + 3 SD) for 5 ms (cf. Harkema et al.
1997
), and ends of bursts were the times at which the signal
remained below the threshold for 60 ms (Fig.
4). If three or more burst onsets were
detected within a single step segment, the bursts were combined by
decrementing the minimum time the EMG signal needed to remain below
threshold to result in a burst end (the resulting mean minimum time
across all analyzed bursts was 57.2 ± 10.0 ms SD), until two or
less bursts remained. Sometimes small, short, apparently spurious
bursts were detected by this algorithm. To eliminate these bursts, the
integral of the EMG signal with respect to time of each burst was
calculated, and bursts whose integral was below 0.002 v/s were
discarded. This threshold value was determined by examining a histogram
of the burst integrals, which had a bimodal distribution with the lower
values corresponding to the small, spurious bursts. The duration of
each burst was defined as the time from the burst onset to the first
burst end. The magnitude of each burst was defined as the magnitude of
the EMG signal at the burst peak. Figure 4 provides a detailed example
of the results of this burst detection procedure.
|
As described previously (Nicolopoulos-Stournaras and Iles
1984
), two bursts were sometimes seen for the biarticular RF
and ST muscles during a single step cycle. One burst occurred near toe-off in association with knee flexion, whereas the other burst occurred near toe-down in association with hip extension. If the peak
of a burst occurred closer to the end of stance it was classified as a
"toe-off" burst, and if it occurred closer to the beginning of stance it was classified as a "toe-down" burst. Toe-off and toe-down bursts were analyzed separately.
Statistics
To determine the effects of the robotic loading, the mean values of the timing and kinematic parameters were first determined across all steps within each exposure (sample size = 25 steps per exposure). The means of those values were then calculated across exposures to the force-field-on and force-field-off conditions (sample size = 10 exposures for each condition). This calculation yielded two numbers for each rat hindlimb: a parameter value for the force-field-on condition and a value for the force-field-off condition. Paired, two-tailed t-tests were performed on the force-field-on and force-field-off values (sample size = number of rats = 8 for each hindlimb) to test the hypothesis that their mean difference was equal to zero. For the EMG analysis, a paired t-test across field exposures (sample size = 5) was used to compare burst parameters for each of the two rats implanted with EMG electrodes.
| |
RESULTS |
|---|
|
|
|---|
The mean force applied bilaterally to the shank during stance (averaged across rats) was 21 ± 5 gmF or about 6% of body weight. When this force was applied, the position of the ankle was lowered during stance (Figs. 5A and 6A). All rats continued to step consistently when the perturbation was imposed to one or both hindlimbs. The total number of steps analyzed for the eight rats for the bilateral and unilateral fields was 7,682 and 8,455, respectively.
|
Imposing an additional load at the lower shank during stance decreased stance duration while maintaining a normal swing trajectory and duration
In response to the bilateral perturbation, step cycle duration decreased significantly in both hindlimbs (Fig. 7). The decrease in step duration was primarily due to a shorter stance duration since swing duration was unaffected. Correspondingly, step length, measured from the toe-down to the toe-off position on the treadmill by video analysis, decreased significantly (Fig. 6D). The step length measured by the robots at the shank (as opposed to the toes) was relatively unchanged (Fig. 5A) because application of the force field caused the ankle to dorsiflex (Fig. 6B) and the lower shank to move downward as well as backward (Fig. 6A). Consequently, the lower shank moved farther backward along the treadmill during stance compared with the distance traveled backward by the toes (Fig. 6D). The backward speed of the lower shank increased during stance (Fig. 5C) due to this rotation (Fig. 6A), even though the treadmill speed was constant.
|
During swing, no force was applied to the hindlimb, yet modifications were apparent in the spatial trajectory and velocity of the hindlimb as it moved forward. Specifically, the arc-length traveled by the lower shank during swing increased when the force field was applied because the hindlimb initiated swing from a lower position (Fig. 5A). Despite this increase in distance traveled during swing, the duration (Fig. 7) and spatial trajectory (Fig. 5A) of the swing were relatively unchanged. In other words, after the beginning of swing, the hindlimb quickly returned to the unperturbed swing path. To accomplish this, the vertical velocity of the hindlimbs was increased during early swing (Fig. 5, B and D).
|
An analysis was performed to determine how quickly the spinal control of stepping responded to the application and removal of the force field. When the force field was applied, step cycle duration decreased significantly on the first step to approximately its mean force-field-on value (Fig. 8A). In contrast, when the force field was turned off, step cycle duration did not increase significantly until the third step (Fig. 8B). Thus the spinal cord responded immediately to application of the force but responded over the course of a few steps to its removal. An analysis also was performed to determine whether the response to the force field varied with repeated exposures. Using a linear regression approach, no significant relationships were found for step cycle or stance duration as a function of the number of exposures to the force field (data not shown).
|
Hindlimb muscle activation patterns were modified by the force field application
Figure 4 shows ankle, knee, and hip muscle EMG recordings (filtered and rectified) from two rats with and without bilateral application of the force field. For the rat implanted with MG/TA electrodes, the peak MG burst amplitude increased by 102% in the right hindlimb and 23% in the left hindlimb in the force-field-on compared with the force-field-off condition (Fig. 9A). No significant differences in burst amplitude were observed for the TA (data not shown). Neither MG nor TA burst duration changed significantly with application of the force field.
|
With application of the bilateral stance field, the peak EMG amplitude of the ST at toe-down increased significantly (33%) in the right hindlimb (Fig. 9B) but was unchanged in the left hindlimb. The magnitude of the right hindlimb RF bursts also increased significantly (348%) with application of the force field. The ST in both hindlimbs seldom exhibited EMG bursts when the force field was off (about one burst per 25 steps at toe-down and none at toe-off), but, when the bilateral force field was on, the frequency of occurrence of toe-down bursts increased to about five per 25 steps and the occurrence of toe-off bursts increased to about one burst per 20 steps (Fig. 9D). The duration of the toe-off bursts of the RF in the right hindlimb increased significantly (from 322 to 914 ms) with application of the bilateral force field. No significant changes in burst duration for the ST muscle were observed.
Both the ipsilateral and contralateral hindlimbs adapted to the unilateral force field
When the force field was applied to only one hindlimb, the downward perturbation during stance was experienced only by the hindlimb to which the force field was applied (Fig. 10, A and B). However, the perturbation resulted in a small but significant decrease in step cycle duration (Fig. 11). Similar to the response observed with the application of the bilateral force field, stance duration decreased in the perturbed hindlimb with application of the unilateral force field (Fig. 11A). This adaptation was accompanied by a decrease in step cycle duration in the contralateral, nonperturbed hindlimb, primarily due to a decrease in swing duration (Fig. 11B). The combination of these changes was sufficient to maintain the interlimb coordination during stepping.
|
|
| |
DISCUSSION |
|---|
|
|
|---|
In this study loads that were applied just above the ankle during stance reduced stance duration and step length while enhancing the level of extensor muscle activity in spinal transected rats. In addition, the normal swing trajectory was preserved even though the loading substantially altered the hindlimb configuration from which swing was initiated. When loading was applied unilaterally the stance phase was shortened on the perturbed side, but interlimb coordination was preserved by reducing the swing duration of the unperturbed hindlimb. These results indicate that the spinal cord coordinated multiple joints and both limbs depending on the sensed state of the limbs, thereby sustaining locomotion in the varying dynamic environment.
Elevated loading during stance increased EMG activity but shortened the step cycle duration
The observation that robotic loading enhanced the magnitude of MG
EMG bursts during stance is consistent with many previous findings. For
example, imposed loading enhances extensor muscle activation in cats
and humans (Guertin et al. 1995
; Harkema et al.
1997
; Hiebert and Pearson 1999
; Sinkjaer
et al. 2000
; Stein et al. 2000
). These
load-related responses apparently function to sustain stepping in the
presence of variations in hindlimb loading.
The downward load applied to the lower shank by the robots decreased
both stance and step cycle durations, demonstrating, as has been shown
previously (Duysens et al. 2000
), that load-related afferent input alters the control of step timing by the spinal cord.
However, some previous studies have found that stance is prolonged if
stance forces are sustained during the end of stance (Duysens
and Pearson 1980
; Hiebert et al. 1994
). Similar
effects have been observed if load-related afferents from the ankle
plantarflexors are electrically stimulated at selected points in the
step cycle of decerebrate rats (Fouad and Pearson 1997
)
and cats (Pearson et al. 1998
). Furthermore, step cycle
and stance durations were longer when some loading of the limbs was
allowed in human infants (Yang et al. 1998
). Other
studies have found different effects on step timing when a load was
applied. For example, pulling downward on the hips of adults
(Stephens and Yang 1999
) or adding mass to the ankles of
running humans (Cavanagh and Kram 1989
) did not change
stride length or step cycle duration. Adding weights to the feet of
walking humans did not change stride length or step cycle duration
(Bonnard and Pailhous 1991
), although it shortened bilateral stance duration and lengthened swing duration. When the mass
was added to the waist, however, bilateral stance duration was
lengthened and swing duration was shortened.
The differences in the results between the present and some of the previous studies of loading may be due to different stimulation methods, experimental models, and species studied (i.e., decerebrated, acute spinal, chronic spinal, intact, neonatal, adult, rat, cat, or human). Alternately, it may be that the robotic loading technique used in this study generated a novel sensory input pattern to the spinal cord. This latter suggestion implies that the differences observed in the duration of stance with load reflect differences in the sensory inputs derived from the varied experimental designs used to study this issue.
For example, the loading technique used in the present study produced
local pressure at the attachment point to the lower shank, which could
have induced flexor reflexes. Thus the step timing changes might be
explained by an increased occurrence of flexor reflexes that
abbreviated stance. However, it is unlikely that stimulation of a gross
flexor reflex would create nearly the same swing trajectory and
interlimb coordination as when the reflexes did not occur. For example,
in designing the attachment interface, we found that if the robot
attachment cuff were placed around the paw rather than the shank, then
the rat would frequently initiate a flexion reflex, hold the limb in
the flexed posture, and stop stepping (Timoszyk et al.
2002
). In the present study, the rats shortened stance but
continued stepping, moving the hindlimb along approximately the same
swing trajectory as when the force was not applied and preserving
interlimb coordination.
Another unique feature of the loading technique was that elevated loads were applied at the lower shank throughout stance, thus imposing a prolonged perturbation but only to the paw and ankle, rather than to the entire hindlimb. Thus the loading technique applied force to a unique subset of load sensors in the hindlimb. In addition, the loading technique changed the configuration of the hindlimb during stance in a way different from would be expected for a load applied through the torso (i.e., the ankle was dorsiflexed, while the knee and hip were not flexed). Thus information from muscle length or joint angle sensors was also uniquely altered and may have contributed to the observed reduction in stance duration.
Based on the results of the present and other experiments noted above,
we suggest that the spinal circuitry does not interpret isolated
parameters such as joint position, muscle length, muscle force, or
cutaneous input but rather an instantaneous "state" that is defined
by the patterns of sensory input from all mechanoreceptors in both
hindlimbs, i.e., as an ensemble. If such is the case, then a modulation
of a single parameter such as muscle force or joint angle for example
will be interpreted in the context of all other input and may evoke
unique responses. A systems-level response is also consistent with the
encoding of global responses by more than half of the spinocerebellar
neurons responding to limb position in the cat (Bosco et al.
2000
). From this perspective it seems feasible that loading of
the plantarflexors could trigger swing in some experimental paradigms,
but not in others. Based on the present experiments, it appears that
the strategy of the control circuitry of the hindquarters during
stepping when the paw (not the whole hindquarters) is loaded only
during stance is to shorten the stance phase and cycle duration.
A limitation of the present study is that EMG recordings were taken
from only two animals and from only two muscles in each leg of each
animal. These recordings may be have been atypical given the increased
variability of EMG within and between spinalized rats. The MG and TA
burst patterns observed in the present study were similar in timing and
amplitude to our previous studies of EMG patterns during normal rat
locomotion (de Leon et al. 1994
; Roy et al.
1991
). Moreover, there appears to be a similar response to
loading during normal rat locomotion, i.e., increasing loading on the
hindlimbs of intact rats by raising the incline level of the treadmill
significantly increased EMG amplitudes in extensor (MG) but not flexor
(TA) muscles.
Persistence of a loading effect for several steps after the elevated loading was stopped
Whereas the spinal cord adjusted its step duration at the onset of
additional loading, it normalized the step cycle timing over the course
of several steps following the removal of the loading perturbation.
Similar "aftereffects" have been observed when an obstacle is
placed in the path of the leg trajectory during the swing phase of
stepping of spinal transected cats (Nakada et al. 1994
).
On evoking a stumbling response, there was an increase in flexion in
the same and the next step. When the obstacle subsequently was removed,
the spinal cord still "anticipated" the obstacle for several steps,
continuing to produce an enhanced flexion during the swing phase. The
swing kinematics did not return to the prestumble features until
several steps after removal of the obstacle. These results underscore
the possibility that load-related control of stepping in spinal animals
can accommodate over the course of several steps reflecting a type of
"short-term memory" of a complex, dynamic motor task in the
lumbosacral spinal cord.
A normal kinematic pattern was maintained during the swing phase when greater than normal loading was applied to the shank during stance
The stance field perturbed the hindlimb trajectory so that the
position of the ankle at the beginning of swing was shifted downward.
During swing, however, the hindlimb quickly returned to the unperturbed
swing path. The early increase in swing velocity may have been due, in
part, to a mechanical spring-like effect arising in the ankle
musculature. In particular, ankle plantarflexors may have stored energy
during stance as they were stretched by the robot-applied load and then
released this energy during early swing. An alternative explanation may
be that the ankle plantarflexors generated greater force at toe-off
because they were stretched to a more advantageous operating point on
their force-length curve. Arguing against a purely mechanical effect,
however, is the observed increase in hip flexor (RF) EMG at swing
initiation, suggesting that the spinal cord modifies upcoming flexor
activity based on load-related input obtained during stance. Also the
accommodation of the timing of the swing phase of the contralateral leg
during unilateral application of the stance force field suggests a
systems-level response that cannot be attributable to a mechanical
recoil phenomenon of the ankle extensors. In particular, the swing
duration of the nonperturbed hindlimb was adjusted during exposure to
the unilateral force field, thereby reducing step duration in the
contralateral leg. Generation of coordinated bilateral motor responses
in response to unilateral perturbations also has been observed during
stepping in decerebrate cats (Hiebert et al. 1994
) and
human infants (Pang and Yang 2001
; Yang et al.
1998
).
Robotic devices can communicate with and thus potentially assess and teach the injured spinal cord through modulation of afferent input to the spinal cord
There has recently been a surge of interest in developing robotic
devices for assessing and retraining upper and lower extremity movement
following neurologic injury (Colombo et al. 2000
;
Hesse and Uhlenbrock 2000
; Lum et al.
2002
; Matjacic et al. 2000
; Reinkensmeyer et al. 2002a
,b
; Volpe et al. 2001
). The present
results demonstrate the feasibility of modulating afferent input
patterns during locomotion with robotic technology. This control
includes the possibility of enhancing load-related afferent information
during stance to enhance muscle output during both stance and swing.
Identification of the patterns of sensory inputs that optimize
locomotor recovery and a clearer understanding of how these sensory
signals are processed to generate a systems-level response in a given
physiological state e.g., following spinal cord injury and/or
regeneration, will be necessary to decipher the neural strategies used
in in vivo systems to control locomotion.
| |
ACKNOWLEDGMENTS |
|---|
The authors thank H. Zhong and M. Herrera for assistance in the surgeries.
This work was supported by Christopher Reeve Paralysis Foundation Grant DA1-9902-2 and by National Institute of Neurological Disorders and Stroke Grants NS-16333 and NS-42951.
| |
FOOTNOTES |
|---|
Address for reprint requests: D. J. Reinkensmeyer, UCI Department of Mechanical Engineering, 4200 Engineering Gateway, Irvine, CA 92697-3975 (E-mail: dreinken{at}uci.edu).
| |
REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. Heng and R. D. de Leon The Rodent Lumbar Spinal Cord Learns to Correct Errors in Hindlimb Coordination Caused by Viscous Force Perturbations during Stepping J. Neurosci., August 8, 2007; 27(32): 8558 - 8562. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Giszter, M. R. Davies, and V. Graziani Motor Strategies Used by Rats Spinalized at Birth to Maintain Stance in Response to Imposed Perturbations J Neurophysiol, April 1, 2007; 97(4): 2663 - 2675. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. McVea and K. G. Pearson Long-Lasting, Context-Dependent Modification of Stepping in the Cat After Repeated Stumbling-Corrective Responses J Neurophysiol, January 1, 2007; 97(1): 659 - 669. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Lam, M. Anderschitz, and V. Dietz Contribution of Feedback and Feedforward Strategies to Locomotor Adaptations J Neurophysiol, February 1, 2006; 95(2): 766 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Y. C. Pang, T. Lam, and J. F. Yang Infants Adapt Their Stepping to Repeated Trip-Inducing Stimuli J Neurophysiol, October 1, 2003; 90(4): 2731 - 2740. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||