INNOVATIVE METHODOLOGY
1Department of Biological Sciences and Department of Applied Physics, Stanford University, Stanford, California 94305; 2Department of Pharmacology, Oxford University, Oxford OX1 3QT, United Kingdom; 3Bell Laboratories, Lucent Technologies, Murray Hill, New Jersey 07974; and 4Department of Molecular Biology, Princeton University, Princeton, New Jersey 08544
Submitted 8 March 2004; accepted in final form 22 April 2004
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
Unfortunately, high-resolution fluorescence imaging studies of neurons remain largely confined to in vitro preparations of cultured cells or tissue slices (Yuste et al. 2000
), because of the experimental difficulties in accessing many neuron types in vivo. These difficulties chiefly concern light scattering, which leads to an exponential attenuation of light intensity with distance traveled through solid tissue. In mammalian brain, even with near infrared wavelengths of 750900 nm that scatter less than visible light, the characteristic scattering length is only
100 µm in mature animals (Oheim et al. 2001
), 140240 µm in young adults (Kleinfeld et al. 1998
; Oheim et al. 2001
), and limits fluorescence imaging depths to several hundred microns within tissue. Nonetheless, a seminal minority of fluorescence microscopy studies of individual neurons has been performed in live mammals, suggesting the possibility of forging direct links between subcellular building blocks and systems level parameters. To date, fluorescence microscopy studies of neuronal properties in live mammals have been limited to superficial or easily accessible tissues, such as trigeminal nerve terminals (Harris and Purves 1989
), parasympathetic ganglia (Harris and Purves 1989
; Pomeroy and Purves 1988
; Purves and Lichtman 1987
; Purves et al. 1987
), and neuromuscular junctions (Akaaboune et al. 2002
; Balice-Gordon and Lichtman 1994
; Walsh and Lichtman 2003
), or to within
500 µm of the neocortical (Grutzendler et al. 2002
; Helmchen et al. 1999
; Stosiek et al. 2003
; Svoboda et al. 1997
, 1999
; Trachtenberg et al. 2002
) or olfactory bulb (Charpak et al. 2001
) surface. Recently, use of a Ti:sapphire regenerative amplifier as an excitation source for ultra-deep two-photon microscopy has been explored (Beaurepaire et al. 2001
; Theer et al. 2003
). This source has enabled a demonstration in which fluorescent neuronal cell bodies were visualized up to 8501,000 µm deep within the neocortex of 3-wk-old mice (Theer et al. 2003
). However, most mammalian brain regions, which are not superficial and lie deeper than 5001,000 µm within tissue, have remained inaccessible to cellular-level fluorescence imaging in vivo.
Our goal was to develop high-resolution fluorescence imaging within deep brain areas that have been unreachable by optical microscopy, because this would open entirelynew possibilities for studying cellular biology in vivo. We constructed fluorescence microendoscopes that exhibit a smaller range of diameters (3501,000 µm) and higher resolution (as good as 0.86 µm) than previous miniature fluorescence endoscopes, which exhibit 18 mm diam and 1.83.1 µm resolution at best (Knittel et al. 2001
; Rouse and Gmitro 2000
; Sabharwal et al. 1999
). Toward this end, we introduced microendoscope probes comprised of multiple gradient refractive index (GRIN) micro-lenses that are 3501,000 µm in diameter and that provide micron-scale resolution (Jung and Schnitzer 2003
). These probes enable minimally invasive in vivo imaging of individual cells in deep tissues (Jung et al. 2003
; Levene et al. 2003
, 2004
). Unlike conventional lenses that refract light at curved surfaces, GRIN lenses are not ground to shape and so are cheaply fabricated in sub-millimeter sizes (e.g., 1001,000 µm) suitable for insertion into solid tissue (Emkey and Jack 1987
; Messerschmidt et al. 1997
; Reed et al. 2002
). In microscopy, it is well known that one- and two-photon fluorescence imaging modalities have complementary strengths and weaknesses, and we sought to develop both modalities in the microendoscopy format. Although use of only a single GRIN lens for microendoscopy suffers from several optical drawbacks (Levene et al. 2002
, 2004
), early work exploring use of compound doublet GRIN micro-lenses for in vivo one-photon microendoscopy (Jung and Schnitzer 2002
; Jung et al. 2002
) and use of doublet (Jung et al. 2002
, 2003
) and triplet GRIN micro-lenses (Jung and Schnitzer 2003
; Jung et al. 2003
; Levene et al. 2003
, 2004
) for in vitro and in vivo two-photon microendoscopy appears promising.
Here we describe in vivo one- and two-photon fluorescence microendoscopy. The strengths and weaknesses of these two modalities resemble those of one- and two-photon fluorescence microscopy, respectively. One-photon fluorescence imaging offers the advantages of full-frame image acquisition, lower cost, and greater ease of use. Although one-photon microendoscopy lacks optical sectioning capabilities, as with one-photon microscopy lack of sectioning can also be used to advantage, because it allows signals originating from different depths to be aggregated toward achieving faster frame-rates. Two-photon imaging involves laser-scanning and provides localized fluorescence excitation that yields inherent optical sectioning and minimal phototoxicity outside the focal waist (Hopt and Neher 2001
). Because essentially all fluorescence photons originate from the focal spot, even photons that scatter in tissue en route to the photodetector can contribute to image signals (Denk and Svoboda 1997
). Thus two-photon imaging is especially well suited for imaging through hundreds of microns of biological tissue.
Using one-photon microendoscopy we obtained video-rate movies of individual red blood cells moving within neocortical, hippocampal, and thalamic capillaries of anesthetized rats and mice. We also visualized multiple hippocampal pyramidal cells within single image frames acquired in the CA1 area of anesthetized mice. Using two-photon microendoscopy in anesthetized rats and mice, we imaged individual neurons and dendrites in the CA1 hippocampal area, up to
270 µm away in tissue from the face of the endoscope.
| METHODS |
|---|
|
|
|---|
GRIN lenses exhibit a spatially varying refractive index, because the glass is inhomogeneously doped with a cation species, such that cation density is typically highest at the central axis of the lens and declines toward the periphery (Gomez-Reino et al. 2002
; Messerschmidt et al. 1997
). This leads to a refractive index profile that declines approximately quadratically with radial distance from the central axis of the lens (Gomez-Reino et al. 2002
; Reed et al. 2002
). In a ray description, total internal reflection occurs in a gradual manner as light passes from central regions to peripheral regions of lower index, causing light rays to travel down the cylindrical lens axis in an approximately sinusoidal path (Gomez-Reino et al. 2002
). The axial length for one full sinusoidal cycle is called the pitch length, P, of the GRIN medium and is determined by the refractive index profile. Two lenses made from the same GRIN substrate but of different pitch (e.g., -pitch versus -pitch) will have distinct lensing properties (Emkey and Jack 1987
). Decreasing P leads to an increased numerical aperture (NA). Using combinations of GRIN micro-lenses with different NA, we have developed endoscope probes for both one- and two-photon fluorescence imaging.
Endoscope probes were custom fabricated thallium-doped GRIN relay lenses (NSG) and thallium-doped (NSG) and silver-doped objectives (GrinTech) glued together with optical epoxy (Fig. 1A). Such lenses use cylindrical GRIN substrate with an index of refraction, n, that varies approximately quadratically with radius, r: n(r) = n0(1 g2r2/2), where n0 is the refractive index on the lens axis. A substrate's pitch length, P, is determined by the quadratic coefficient g, P = 2
/g.
|
To relate the NA, working distance (WD), focal distance (f), field of view (FOV), and lateral magnification (ML) of our endoscope probes to the properties of the GRIN substrates, we used the matrix formulation of paraxial ray optics (Kogelnik 1965
). Although more detailed design can be performed using numerical ray tracing, this analytical treatment provides insight and satisfactory agreement with experimental values for many purposes. A GRIN objective lens that is immersed in a sample medium of refractive index nm and that possesses a diameter d, length L, and radial refractive index profile nob(r) = nob,0(1
r2/2) will focus collimated rays entering one face of the objective to a focal spot a working distance
![]() |
![]() |
![]() |
in the paraxial approximation. For the doublet designs of Fig. 1, in which a GRIN objective lens of 0 < L < P is combined with a relay lens that is P modulo P (i.e., P, P, etc.; Fig. 1, C and D) and that possesses a refractive index profile nR(r) = nR,0(1
r2/2), a real image of the specimen is focused at the external face of the relay lens with lateral magnification
![]() |
![]() |
With two-photon imaging, we sometimes used a triplet design comprised of an endoscopic objective lens, a -pitch relay lens, and a -pitch coupling lens made of the same GRIN substrate as the objective (Fig. 6C). In this case, the relay lens acts as a telescope of magnification minus one. Although the use of the -pitch relay and the addition of the coupling lens do not change the above formulae for WD and FOV, the expression for ML simplifies: ML = sin(gobL).
|
One-photon microendoscopy
The one-photon fluorescence endoscope was constructed on a rail (X95, Newport, Irvine, CA) and was mounted on a translation stage (XYR-8080, New England Affiliated Technologies, Lawrence, MA) driven by a stepper motor controller (MP285, Sutter, Novato, CA). Key optical components include a binocular (452934, Zeiss) and eyepieces (444034, Zeiss). Light from a mercury arc lamp (HBO 100, Zeiss) is coupled into a 1.0-mm-diam fiber (Multimode Fiber Optics, East Hanover, NJ) as part of an illumination scrambler (Inoue and Spring 1986
). After passing through a wheel (Linos) of fluorescence excitation filters (Chroma, Rockingham, VT), the excitation light is collected by a Leitz H 32x 0.6 NA objective and a Hastings triplet lens (Edmund Optics, Barrington, NJ) and deflected onto the main optical axis by one of several exchangeable dichroic mirrors (Chroma) such that the light overfills the back aperture of a microscope objective. The objective delivers the light to the endoscope probe (Fig. 1B).
Fluorescence emissions return back through the endoscope probe, microscope objective, dichroic mirror, and a wheel (Linos) of emission filters (Chroma). Images can be viewed through the eyepieces of the binocular or projected onto a camera. For imaging neurons stained with Di-I, we used the fluorescence filter set HQ535/50-Q565LP-HQ610/75 (Chroma) and a cooled CCD camera (Coolsnap HQ, Roper) controlled by commercial imaging software (V++, Digital Optics; IPLab, Scanalytics, Fairfax, VA). Red blood cell imaging involved the filter set HQ470/40-Q495LP-HQ525/50 or -HB535/30 (Chroma) and a monochrome CCD camera (XC-ST70, Sony). Movies were stored on DV tape using a digital video recorder (DCR-TRV900, Sony). Imaging in YFP-H mice involved a HQ470/40-Q495LP-HQ535/30 filter set and the cooled CCD camera.
There are two mechanisms for focusing the image, implemented through a pair of modular focus units (81705, Nikon) that hold the microscope objective and endoscope probe. One unit adjusts the position of the endoscope probe within the animal. The second unit adjusts the position of the microscope objective relative to the endoscope probe; therefore the image focus within tissue can be altered without any mechanical movements within the animal.
Optical resolution measurements
Spatial calibration of the endoscope's motorized X-Y stage was performed using a stage micrometer with a 10-µm grid imaged under transmitted light provided by a light box (455136, Zeiss). To estimate the fluorescence intensity point spread function, we prepared microscope slides with individual 170-nm-diam fluorescence beads (P7220, Molecular Probes, Eugene, OR) fixed to the slide surface. A clean slide was washed in 1 M NaCl and a coverslip (no. 1.5, Corning, Corning, NY) was fixed on the slide using double-sided tape as an adhesive spacer. We filled the chamber with a solution of fluorescence beads and waited about 1 min for beads to adhere to the coverslip. We then flushed the chamber with water to remove any beads remaining in solution and allowed the chamber to dry in air.
Fluorescence beads were epi-illuminated under the one-photon endoscope, and images were filtered using a band-pass (HQ535/30) and acquired at 12-bit resolution with the cooled CCD camera. The images were imported into image analysis software (NIH Image) and one-dimensional cross sections through the bead center were normalized in amplitude and then fit with an Airy disk, f(x) = [2J1(kx)/(kx)]2 (Diaspro 2002
). The full width at half-maximum (FWHM) of the Airy disk was obtained from its inverse proportionality to the fitting parameter, k. Standard errors were obtained using the distribution of FWHM values measured over a population of beads.
Tests of optical resolution using grid lines were performed using a Ronchi ruling (600 lines/mm, Edmund Optics) or a Marine Biological Laboratory/National Nanofabrication Facility (MBL/NNF) transmittance test slide (Oldenbourg et al. 1993
). The MBL/NNF slide exhibited a Siemens star and grid lines at a variety of spacings. Resolution was estimated using the Siemens star by determining the radius at which image contrast between adjacent wedges disappears. Errors in radial measurements yielded errors in resolution values.
Two-photon microendoscopy
The two-photon endoscope was constructed on an X95 rail (Newport) mounted on a stepper motor driven X-Y translation stage (New England Affiliated Technologies). This instrument was built as a modification to a previously existing two-photon microscope (Svoboda et al. 1997
). A wavelength-tunable ultra-short pulsed Ti:sapphire laser (Tsunami, Spectra-Physics, Mountain View, CA) pumped by a 10 W Nd:YVO4 laser (Millenia, Spectra-Physics) provides
100-fs pulses of near infrared light for fluorescence excitation. A pair of galvanometer-driven deflector mirrors (Cambridge Technologies, Cambridge, MA) steer the excitation beam in two angular dimensions as controlled by custom written imaging software (Confoc, Raymond Stepnoski). The beam passes through a scan lens (4524268033, Zeiss), a dichroic mirror, and a binocular tube lens (Zeiss), and overfills the back aperture of a microscope objective lens that focuses the beam just above the top face of either a doublet or triplet GRIN endoscope probe. Scanning the laser focal spot in this plane leads to scanning of the focal spot in the specimen image plane.
Anesthetized animals were placed underneath the endoscope probe. The focal plane of the excitation beam within the brain can be adjusted using a custom-built dual-focus mount based on two translation stages (Newport) and two stepper motors. A photomultiplier tube (R3896, Hamamatsu) was used to detect fluorescence returning through the endoscope probe (Figs. 6 and 7). A computer controls the scanning of the galvanometer mirrors and constructs the digital image. In vitro tests of lateral optical resolution had shown that with probes of near 0.5 NA two-photon microendoscopy attains 12 µm resolution (Jung and Schnitzer 2003
). For imaging Di-Ilabeled neurons, the laser wavelength was tuned to 815 nm, and for imaging YFP mice, the laser was tuned to 910925 nm. Typical laser excitation power at the sample was 2550 mW.
|
All animal procedures were performed in accordance with National Institutes of Health guidelines and were approved by the Bell Laboratories and Stanford University Institutional Animal Care and Use Committees.
Adult female Sprague-Dawley rats were anesthetized with urethane (1.5 mg/g body weight, ip) or ketamine and xylazine (0.13 and 0.01 mg/g body weight, ip, respectively). Body temperature was monitored with a rectal probe and was kept stable at 36°C with a feedback-controlled heat blanket (Harvard Apparatus, Holliston, MA). The head was shaved, and the eyes were covered with ointment (Puralube Vet, Pharmaderm, Melville, NY). An incision was made through the scalp along the midline for a length of 11.5 cm. The skin was retracted, and the skull was exposed. A craniotomy was performed over the brain area of study.
We used two general methods for reaching deep brain tissue with the microendoscope probe. The first method involved gradually lowering the probe through brain matter until it reached the area of interest. Thus for our initial studies in rat somatosensory cortex, a 2-mm diam craniotomy was performed at stereotaxic coordinates 4 mm lateral, 3 mm from bregma. After dye labeling (see next section), the endoscope probe was gradually lowered 1 mm ventral into tissue perpendicular to the cortical surface. The brain surface was washed with artificial cerebral spinal fluid (ACSF) and imaging commenced. ACSF was composed of (in mM) 125 NaCl, 5 KCl (Baker), 10 glucose, 10 HEPES, 2 (MgSO4 · 7 H2O) (Fisher Scientific), and 2 (CaCl2 · 2 H2O) (all chemicals are from Sigma except where noted). Although this method for inserting the microendoscope into tissue is similar to that used previously (Levene et al. 2004
) and yielded crisp images of both neurons and blood cell dynamics, we were concerned about the possibility of compression of brain tissue that might depress neuronal activity (Bolay et al. 2002
). Thus, during studies of the CA1 hippocampal area and of lateral dorsal thalamus, we employed a second method that is an intermediate between our first approach and a surgical method recently developed for imaging hippocampus in vivo by conventional microscopy (Mizrahi et al. 2004
).
After aspirating overlying neocortex in the anesthetized mouse, it is possible to use a conventional microscope objective to image fluorescent hippocampal pyramidal cells (Mizrahi et al. 2004
). Although this approach does obviate concerns about compression of overlying tissue, aspiration of large portions of neocortical matter is highly invasive. We developed an intermediate approach in both rats and mice in which prior to insertion of the endoscope probe, we removed a thin column of tissue overlying the brain area of interest. We then inserted the endoscope through this opening. The brain area under observation was left intact. With experience, we were able to image through smaller craniotomies (about 1.5 mm) and with correspondingly less tissue removed, without noticeable effects on image quality. This method allowed us to image the rat laterodorsal thalamus in vivo, slightly less than 4 mm below the neocortical surface, which is prohibitive with conventional microscopy even after large-scale surgical removal of neocortical matter.
For studies of rat hippocampus, a 1.5- or 3.0-mm-diam craniotomy centered at coordinates 1.9 mm lateral, 4.0 mm from bregma was opened. The dura was removed with forceps. Bent and blunted needles (23 and 26 gauge) were used to aspirate a column of tissue from overlying neocortex and corpus callosum that had a diameter slightly smaller than that of the craniotomy. The dorsal alveus layer of hippocampus remained untouched and intact. During aspiration, fine control of suction strength was performed by partially covering a 1-mm hole drilled in the plastic fitting of the suction needle. The corpus callosum and any clotted blood was removed by washing with ACSF while aspiration suction was reduced.
Endoscopic imaging studies of rat laterodorsal (LD) thalamus were performed using a 2.0-mm-diam craniotomy centered at 2.4 mm lateral, 3.5 mm bregma or at 1.9 mm lateral, 3.0 mm bregma. Overlying neocortical and hippocampal tissue within the diameter of the craniotomy were aspirated. LD was untouched and left intact.
Female mice were anesthetized with injection of ketamine (0.13 mg/g ip) and xylazine (0.01 mg/g ip). The head was shaved and eyes were covered with eye ointment. For imaging in CA1 a 1.5- or 2.0-mm-diam craniotomy at coordinates 2.0 mm lateral, 2.0 mm from bregma was opened. Neocortex and corpus callosum were aspirated over a diameter the same as that of the craniotomy. The alveus layer of hippocampus was untouched and remained intact.
Fluorescence labeling
For red blood cell imaging, fluorescein (0.8 mg/kg body weight dissolved in ACSF, Sigma-Aldrich, St. Louis, MO) was injected into the femoral vein of rats and mice with a 30-gauge needle. The injection was done under a surgical stereoscope to ensure that dye was only injected into the circulation and not into adjacent tissue.
Rat neocortical neurons were labeled using Di-I paste (Molecular Probes). The tip of a glass microelectrode of
3 M
impedance was dipped into the paste and inserted up to 1 mm into the cortex. The microelectrode was then removed. After waiting
30 min or longer for the dye to diffuse, individual neurons at the periphery of the microelectrode's path through tissue were visible. The endoscope was then inserted into tissue adjacent to the insertion point of the microelectrode.
Imaging studies of fluorescent CA1 hippocampal neurons in live mice involved the YFP-H transgenic line that expresses yellow fluorescent protein (YFP) under control of neuron-specific elements from the thy1 gene (Feng et al. 2000
).
| RESULTS |
|---|
|
|
|---|
Despite important differences regarding instrument size and lens design, the instrument that we built for one-photon fluorescence microendoscopy exhibits many of the strengths and weaknesses of conventional epi-fluorescence microscopes (Fig. 1). Epi-fluorescence imaging typically employs incoherent illumination, which we provided using a Mercury-arc lamp, and enables direct viewing through eyepieces or image capture on a pixilated detector such as a CCD or photodiode array (Yuste et al. 2000
) (Fig. 1B). Our instrument uses conventional epi-fluorescence filter sets and a microscope objective that couples fluorescence excitation light into a compound doublet microlens probe (3501,000 µm diam; Fig. 1A) comprised of a GRIN endoscopic objective lens and a GRIN relay lens (Fig. 1, C and D). This microendoscope probe delivers the illumination to the tissue and projects a real epi-fluorescence image to the external face of the relay lens (METHODS; Fig. 1, C and D), at the front focal plane of the microscope objective (Fig. 1B). A benefit of our design is that there are two mechanisms for focusing the image. One mechanism adjusts the position of the endoscope probe within the animal. After the initial position of the endoscope probe is set in tissue, it is often advantageous to keep the probe stationary. The second mechanism adjusts the position of the microscope objective relative to the probe, so that the image focus can be altered without any mechanical movements within the animal.
To estimate endoscope resolution, we performed three different tests. First, we imaged grids that were nanofabricated by electron lithography (Oldenbourg et al. 1993
). With three endoscopes of NA from 0.380.47, we found that grids with 1.66- or 1.2-µm spacing could be sharply resolved (Fig. 2A). These same three endoscopes barely resolved a grid of 1.0-µm spacing, and the images of this grid exhibited poor contrast (Fig. 2A, bottom inset). This indicates that the optical resolution of these endoscopes according to the Sparrow criterion of zero contrast (Diaspro 2002
) is only slightly better than a micron.
|
Third, we imaged subresolution 170-nm fluorescent beads (Fig. 2B), which mimic a point source of fluorescence and allow one to study the intensity point spread function of an optical instrument (Diaspro 2002
). We determined FWHM values of 0.89 ± 0.01, 0.98 ± 0.02, and 1.16 ± 0.1 µm for the endoscopes of 0.47, 0.46, and 0.38 NA, respectively. The FWHM is
7% broader than the minimum separation at which two identical Airy disks sum so that their distinct peaks can just barely be distinguished, as in the Sparrow criterion (Diaspro 2002
). Thus these values are consistent with those obtained using the Siemens star.
Video-rate microendoscopy of red blood cell dynamics in deep brain areas
Our first in vivo experiments with one-photon microendoscopy in the mammalian brain involved visualization of cerebral blood flow and red blood cell dynamics. Cerebral blood flow and metabolism constitute important biomedical topics, not only because of the clinical relevance to stroke and ischemia (Dijkhuizen et al. 2001
) and migraine (Bolay et al. 2002
), but also because further study of the coupling between neural activity and vascular response is needed toward understanding the physiological underpinnings of brain imaging modalities that rely on vascular and metabolic effects (Fox et al. 1986
; Martin et al. 2002
; Ogawa et al. 1992
; Shtoyerman et al. 2000
). Prior imaging studies have examined mammalian blood flow in the pia (Hudetz et al. 1997
; Ma et al. 1974
) and in the upper 250 µm (Dirnagl et al. 1992
; Villringer et al. 1994
) or upper 600 µm (Kleinfeld et al. 1998
) of neocortex. As we describe below, in vivo video-rate microendoscopy can now be used to visualize blood flow in individual vessels deep within the live rodent brain.
We examined blood cell dynamics by injecting the fluorescent dye fluorescein into the blood stream. This provides a background label of blood plasma, causing the red blood cells, which are
8 µm in diameter, to appear dark by comparison (Kleinfeld et al. 1998
). Video-rate movies of blood flow revealed trajectories of individual red blood cells in larger vessels and capillaries in rat neocortex (data not shown), mouse hippocampus (Fig. 3A, Movie 1), rat hippocampus (Fig. 3B, Movie 2), and rat thalamus (Fig. 3C, Movie 3). Blood cell speeds in capillaries varied widely, generally over the range of
30350 µm/s, which is within the broad range of speeds observed in prior imaging studies of capillary blood flow in the pia and neocortex (Dirnagl et al. 1992
; Hudetz et al. 1997
; Ivanov et al. 1985
; Kleinfeld et al. 1998
; Ma et al. 1974
; Villringer et al. 1994
). In capillaries, individual red blood cells were particularly apparent and often progressed forward in an irregular manner, occasionally even reversing direction momentarily. Such irregular red blood cell motion has been observed previously in neocortex (Kleinfeld et al. 1998
). Because video-rate movies convey a much stronger sense of individual blood cell motions than do still frames, readers are encouraged to view movie data (available in compressed form in Supplementary Information, Movies 13)1that formed the basis for the still-frame sequences shown in Fig. 3. Endoscopic images of individual red blood cells exhibited the biconcave structure of these cells (Fig. 4), showing our capacity to resolve subcellular micron-scale details in vivo. Because of the lack of optical sectioning in one-photon fluorescence endoscopy, fluorescent blood vessels that were out of the image focal plane were often apparent. This did not prove to be a major obstacle to achieving video-rate imaging of blood cell dynamics in brain areas that have been inaccessible to conventional microscopy.
|
|
To demonstrate our ability to visualize neurons deep within live mammals, we examined neocortical neurons in anesthetized rats and hippocampal neurons in anesthetized mice. To label rat neurons in somatosensory cortex, we used the lipophilic fluorescent dye, Di-I, which binds to and diffuses along lipid membranes (METHODS). After delivery of dye into infragranular layers of somatosensory cortex, allowance of sufficient time for the dye to label neuronal processes, and insertion of the microendoscope probe into neocortex 1 mm ventral to the pia, we were able to visualize individual fluorescent cells and dendrites (e.g., Fig. 5A).
|
These data show that one-photon microendoscopy can be used to visualize individual neurons in vivo. As demonstrated in the YFP-H animals, the method provides a convenient means of detecting fluorescent proteins within both dendritic and somatic cellular compartments in deep brain neurons in vivo, which should be applicable to the study of other transgenic animal lines. As with conventional epi-fluorescence microscopy, when imaging a specimen embedded deep within tissue, there is a fluorescence background from out of focus and scattered fluorescence emissions (Fig. 5, C and D). Thus, to provide three-dimensional sectioning and to increase image contrast, we developed in vivo laser-scanning two-photon microendoscopy.
Laser-scanning two-photon fluorescence microendoscopy
In our previous work, we developed two-photon microendoscopy using a GRIN triplet endoscope probe (Jung and Schnitzer 2003
). In this study, we explored several variants of our original strategy for laser scanning (Fig. 6). One variant uses a doublet endoscope probe comprised of an endoscope objective lens and a -pitch relay lens (Fig. 6A), nearly identical to the probe depicted in Fig. 1C for one-photon imaging. (Minor differences concern the distinct wavelengths used.) A low NA achromatic coupling lens focuses the laser beam just above the face of the endoscope relay lens (Fig. 6A), and a pair of galvanometer-driven deflector mirrors scans the focal spot laterally in two dimensions (Fig. 7). The endoscope probe translates the demagnified focal spot and its scan pattern onto the sample focal plane (Fig. 6A). A portion of the fluorescence emissions returns through the endoscope probe, is separated from the excitation beam by a dichroic mirror, and is captured by a photodetector (Figs. 6A and 7). Another variant uses a doublet probe with a -pitch relay (Fig. 6B) for longer reach into deeper tissues. A third variant uses a GRIN micro-lens triplet probe (Fig. 6C) (Jung and Schnitzer 2003
). The endoscope objective lens has a length less than -pitch and a working distance ranging from 1301,040 µm. The relay lens is an integral multiple of -pitch length and enables insertion into deep tissues. A third GRIN lens is an endoscope coupling lens of 0.5 NA, which matches the NA of a microscope objective lens that couples the laser beam into the endoscope probe (Fig. 6C). These design variants provide some flexibility in the optical and mechanical layout of a two-photon endoscope. Scanning is achieved by pivoting the collimated laser beam about the back aperture of the microscope objective (Fig. 6C) or achromatic lens (Fig. 6, A and B) that focuses the excitation beam onto the endoscope probe. This leads to lateral translation of the laser focal spot. An alternative strategy involves a collimated beam of narrow diameter that pivots about the external face of the endoscope probe (Fig. 6D). In this case, the endoscope consists of a GRIN objective lens and a -pitch GRIN relay, and the focal spot is again scanned laterally in the sample focal plane. Although initial (Jung and Schnitzer 2003
) and subsequent (Levene et al. 2004
) work involving compound GRIN micro-lenses for two-photon imaging has used triplet GRIN probes, after exploration of the various scanning strategies depicted in Fig. 6, we currently favor use of doublet GRIN lens probes (Fig. 6, A and B), which offer similar fields of view as the triplet probes but at about two-thirds the cost and with greater image magnification (46x) and lower NA for incoupling the laser beam. These optical advantages allow one to use a simple low NA achromat to couple the laser into a doublet endoscope probe, rather than a microscope objective of
0.5 NA. Furthermore, the resulting weaker focus of the incoupled beam can help to reduce autofluorescence at the endoscope surface.
In vivo neuronal imaging using two-photon microendoscopy
Initial tests of in vivo two-photon microendoscopy involved anesthetized rats. We introduced Di-I into the ventral layers of CA1 and positioned the endoscope probe atop the hippocampal alveus. After sufficient time for the dye to diffuse dorsally, two-photon images revealed neuronal dendrites (Fig. 8A) in s. oriens. The sectioning capability of two-photon endoscopy was readily apparent during in vivo imaging, because micron-scale movements of the brain or of the endoscope probe brought individual dendritic features in s. oriens in and out of view with an axial sensitivity far surpassing that of one-photon microendoscopy. Further tests of imaging in CA1 involved anesthetized YFP-H mice expressing YFP in a subset of hippocampal pyramidal neurons (Feng et al. 2000
). With an endoscope probe positioned just dorsal to the hippocampal alveus, two-photon microendoscopy images of s. oriens revealed the basal dendrites of the fluorescent pyramidal cells (Fig. 8B).
|
25 pyramidal cell bodies at low magnification (Fig. 8C) or fewer cell bodies and greater dendritic details at higher magnifications (Fig. 8, D and E). At more ventral focal depths (
260 µm below alveus), we were able to visualize the larger apical dendritic trunks in s. radiatum (Fig. 8F). Without the optical sectioning capabilities of two-photon microendoscopy, it would otherwise be very difficult to isolate signals from an individual dendrite in s. radiatum without being overpowered by background fluorescence from the dense layer of fluorescent cell bodies dorsally overlying in s. pyramidale. | DISCUSSION |
|---|
|
|
|---|
1 mm or more from the cranium, well beyond the imaging range of conventional in vivo microscopy. We employed these microendoscopes for in vivo fluorescence imaging of individual cells in hippocampus and in thalamus, deep brain areas that have been inaccessible to cellular imaging in live mammals. A major advantage of our imaging approach is that use of compound micro-lens probes allows us to place a mechanism for focal adjustment on the side of the probe that is external to the animal. In comparison, endoscopes based on individual or bundles of optical fiber must either operate at a fixed working distance (Dickensheets and Kino 1996
0.85 µm (Fig. 2), which we employed to discern subcellular details (Figs. 3, 4, 5, and 8). However, to date, we have been unable to visualize dendritic spines. Further improvements in resolution achieved by either raising the NA of the endoscope probes or by correcting for their optical aberrations will be necessary to resolve spines. In this work, we have introduced one-photon fluorescence microendoscopy and its use in vivo. Our one-photon microendoscopy images (Figs. 3, 4, and 5) illustrate the potential offered by this technology for visualization of a wide variety of cell types with micron-scale resolution deep within tissues of live animals. One-photon microendoscopy is simpler to use and cheaper to construct than two-photon microendoscopy, and enables full-frame image capture for direct viewing of fluorescent cells through eyepieces or with a camera. Such full-frame detection opens the possibility of achieving fast-frame rate imaging of cellular dynamics. The lack of optical sectioning inherent with this modality implies that signals from adjacent depths within tissue may be collected simultaneously, but the accompanying disadvantage is the presence of scattered and out of focus background fluorescence within the image.
We have exploited these properties of one-photon microendoscopy to produce video-rate movies of individual red blood cells flowing through hippocampal and thalamic capillaries in live rodents. Prior optical imaging studies of microcirculation were limited to vessels in the pia (Ma et al. 1974
) and in the upper 250 µm (Villringer et al. 1994
) to 600 µm (Kleinfeld et al. 1998
) of mammalian neocortex. Using two-photon excitation, GRIN lenses, and one-dimensional laser line scans, blood cell speeds have been estimated deep within the neocortex (Levene et al. 2004
), but this approach does not provide full frame movies or the individual trajectories of single blood cells. Microendoscopy studies of red blood cell dynamics in response to neural stimulation may help provide insight into mechanisms that couple nervous activity to hemodynamics and that underlie functional MRI (Ogawa et al. 1992
), PET (Fox et al. 1986
), and intrinsic signal imaging modalities (Malonek and Grinvald 1996
).
Using one- and two-photon microendoscopy we have visualized deep brain neuronal cell bodies and dendrites in live rats and mice. The laser-scanning two-photon endoscope is particularly well suited for in vivo studies of subcellular structures such as individual neuronal dendrites (Fig. 8, A and F), because the instrument combines the ability of a two-photon microscope to sample thin optical sections with the ability of an endoscope to reach tissue areas millimeters deep below the surface. This combination should prove useful for in vivo imaging of fluorescent markers in a wide range of transgenic mouse lines. During both one- and two-photon microendoscopy (Figs. 5, C and D, and 8, CE), multiple fluorescent cell bodies were visible within single image frames acquired in the CA1 area of transgenic YFP-H mice that express YFP within a subset of pyramidal neurons. As in our preliminary accounts (Jung et al. 2003
), we report here two-photon imaging depths of
270 µm from the face of a GRIN microendoscope, sufficient to visualize individual dendrites in CA1 hippocampal s. radiatum in an adult mouse when a microendoscope probe is positioned just dorsal to the CA1 hippocampal alveus (Fig. 8E). The ability to image through such tissue depths has allowed us to visualize multiple anatomical layers without repositioning the microendoscope probe and compares favorably to a report of only
95 µm of imaging depth from the face of a GRIN lens probe (Levene et al. 2004
). Nonetheless, our two-photon imaging depth of
270 µm is inferior to the >500 µm depths provided by conventional in vivo two-photon microscopy (Kleinfeld et al. 1998
; Oheim et al. 2001
). This difference almost certainly arises in large part due to the lower NA values of our microendoscope probes compared with those of commercially available water immersion microscope objective lenses, since the efficiency of fluorescence photon capture rises nonlinearly with NA (Yuste et al. 2000
).
A small number of other studies in the live mammalian brain have used optical fibers or endoscopes. These studies either involved single optical fibers that did not provide images (Duff Davis and Schmidt 2000
; Ikeda and Matsushita 1980
; Kudo et al. 1992
) or coarse-resolution endoscopes that were incapable of resolving individual cells and that relied on scattered light rather than fluorescence (Poe et al. 2003
; Rector et al. 2001
). In clinical practice, neuroendoscopes are used during aneurysm treatments (Wang et al. 2003
) and intraventricular procedures (Anandh et al. 2001
), but these endoscopes of 14 mm diam are designed for viewing millimeter-sized fields at low resolution. Miniature two-photon microscopes that can resolve individual cortical neurons currently remain too large to be used for microendoscopy (Helmchen et al. 2001
).
The use of microendoscopes for imaging deep brain tissues in vivo differs from another recently introduced in vivo imaging methodology that relies on conventional microscope objectives and surgical removal of overlying brain tissue. Katz and colleagues have found that surgical removal of a portion of neocortical matter allows examination of the dorsal tissue of hippocampus by conventional two-photon microscopy (Mizrahi et al. 2004
). After this surgical intervention, they have tracked the stability of hippocampal dendritic spines over several hours. Although such methodology employs water immersion objectives that offer superior resolution and higher NA than currently available GRIN micro-lenses provide, this invasive surgical approach may be limited in use to hippocampus or other structures within the working distance of the microscope objective from the cranium. For example, in vivo imaging of rat thalamus, which is about 4 mm ventral to the neocortical surface, would apparently require surgical removal of overlying neocortical and hippocampal tissue within a diameter broad enough to insert the microscope objective, thereby causing massive tissue disruption. An attempt to skirt this problem by use of a long working distance microscope objective and a small diameter tissue resection would limit the usable NA of the objective lens to that afforded by the geometry of the resection. Furthermore, surgical resection may preclude examination of certain issues that require study in a chronic preparation, because large-scale removal of afferents and efferents may induce long-term disturbance of neuronal properties. We have developed general methodologies for in vivo optical imaging of deep tissue such as thalamus (Fig. 3C) with high NA (
0.5) but without large-scale tissue resection.
Our methods rely on insertion of compound GRIN microlenses into tissue. Such microendoscopes of 3501,000 µm diam are the smallest probes used to date for in vivo cellular imaging of which we are aware (Jung and Schnitzer 2003
; Jung et al. 2003
; Levene et al. 2004
), but this diameter range exceeds that of microelectrodes. The micro-lenses are closer in size to microdialysis probes that are commonly used to study neurochemistry in vivo and that range from
250 µm and upward in diameter. Although the use of microendoscopes appears to be less invasive and more generally applicable than the surgical approaches developed by Mizrahi et al. 2004
, future studies must consider the potential effects of the microendoscope probe on the biological phenomenon under consideration. Several potential strategies exist for inserting the endoscope probe in as minimally invasive a manner as possible, and two of these strategies have been explored to date.
The first strategy involves gradual insertion of the probe into the brain area of interest. We used this approach for studies of deep neocortical capillaries and neurons (Fig. 5A), and Levene et al. 2004
also adopted this tactic, employing stepwise increments sized a few tens of microns. However, insertion of a blunt optical probe will cause some degree of tissue distortion and compression, which may create a spreading depression of neural activity (Bolay et al. 2002
). Thus we also developed a second strategy that is intermediate between gradual probe insertion and gross resection and that seeks to minimize compression of the brain area under study. In this intermediate approach, a column of tissue overlying the brain area of interest is removed over a diameter comparable to that of the microendoscope probe. The microendoscope probe may be inserted through this opening without compressing the tissue to be visualized. The brain area under observation is left intact, and the probe is positioned just above. When this method was applied in the mouse brain, our imaging depth of up to
270 µm from the face of the endoscope probe allowed us to visualize cellular details in multiple hippocampal tissue layers (Fig. 8), without physically entering the hippocampus with the microendoscope probe. Furthermore, we have found that positioning the endoscope probe right above the structure of interest can have a stabilizing mechanical effect. We have often seen a discrete reduction in image distortion that accompanies an incremental mechanical advance of the probe, and we interpret this event as the endoscope probe just touching down on the brain tissue. In future in vivo imaging studies, the choice of whether to employ conventional microscopy and surgical resection, or microendoscopy with gradual probe insertion or minimal tissue removal, will largely depend on the particular brain area under study and the scientific requirements of the experiment. Further pursuit of all three approaches is needed to explore their relative strengths and weaknesses in detail. In microdialysis studies, a common tactic involves positioning the probe within a ventricle, and a similar approach might be fruitful for microendoscopy. Improvement of methods employing minimal tissue removal might involve use of a hypodermic guide tube, within which tissue is aspirated followed by insertion of the microendoscope through the tube.
The methods introduced here open possibilities for several distinct classes of in vivo experiments in deep brain tissues involving microendoscopy. One-photon microendoscopy should be particularly well suited for experiments that combine fast frame-rate imaging and the use of synthetic (Stosiek et al. 2003
) or genetically encoded (Guerrero et al. 2002
; Kuner and Augustine 2000
; Truong et al. 2001
) activity-dependent fluorescent probes, allowing neural activity to be monitored within multiple cell bodies simultaneously. Fluorescent gene reporters (Brown et al. 2001
; Spergel et al. 2001
) could be fruitfully visualized with either endoscopic modality, enabling studies of how environment, therapeutic agents, or specific patterns of brain activity affect gene expression. For determination of how such factors may alter fluorescent protein distributions within specific tissue laminae and cellular compartments, the sectioning capabilities and greater imaging depth offered by two-photon microendoscopy should make this modality the preferred choice. Our two-photon images of hippocampal pyramidal cells in the YFP-H mice revealed the presence of YFP within basal dendrites in s. oriens, cell bodies in s. pyramidale, and apical dendritic trunks in s. radiatum. Future in vivo imaging studies regarding laminar and cellular protein distributions would likely involve fluorescent protein fusions, such as those used to study somatodendritic patterns of cytoskeletal dynamics (Furuyashiki et al. 2002
; Stepanova et al. 2003
) and ion-channel localization (Antonucci et al. 2001
) in vitro. Both microendoscopy modalities should also work well for combined optical and electrophysiological studies in which electrodes are targeted toward particular classes of neurons expressing fluorescent markers (Margrie et al. 2003
). Finally, two-photon microendoscopy would generally be superior to its one-photon counterpart for tracking Ca2+ ion dynamics within individual dendrites using fluorescent Ca2+-indicators, as performed previously using two-photon microscopy near the mammalian brain surface in combined imaging and electrophysiological studies (Charpak et al. 2001
; Helmchen et al. 1999
; Svoboda et al. 1997
, 1999
; Waters et al. 2003
).
Further development of microendoscopy methodologies may enable <