|
|
||||||||
INNOVATIVE METHODOLOGY
Department of Psychology, The University of Connecticut, Storrs, Connecticut
Submitted 5 November 2004; accepted in final form 7 December 2004
| ABSTRACT |
|---|
|
|
|---|
| INTRODUCTION |
|---|
|
|
|---|
For some experimental aims, however, it would be desirable to leave electrodes within the brain for longer periods and to slowly explore each of the single neurons that lie along several closely spaced microelectrode penetrations. In principal, Reitboeck electrodes are well suited for chronic indwelling use because both the platinum alloy core and the quartz shell are very durable and stable materials. Here, we describe a system for using densely packed, indwelling arrays of independently controlled Reitboeck electrodes for long-term, chronic recording. We describe a compact and simple method for preventing the buckling of these electrodes that is suitable for use with electrode shaft diameters as small as 40 µm, an ultra-miniature, multiple microdrive system that allows the independent control of these electrodes, and procedures for keeping electrodes mobile and under microdrive control for many months by minimizing the entrance and subsequent coagulation of brain fluids within the guide tubes. We have used variants of this system in rabbits for >5 yr. Electrodes typically remain mobile, under microdrive control, and able to record from well-isolated single neurons for periods of many months and, in once case, for >4 yr.
| METHODS |
|---|
|
|
|---|
2100°C) vertical puller (Thomas Recording). This is done within a chamber that is filled with inert gas to prevent oxidization of the tungsten filament. The tips are then sharpened to the desired characteristics using a fine diamond grinding wheel.1 Methods for fabricating these electrodes have been described by Reitboeck (1983a)
|
A primary problem impeding the use of these fine filament electrodes is that the shaft readily "buckles" when the tip encounters any resistance (e.g., the dura) (see DISCUSSION in Eckhorn and Thomas 1993
; Reitboeck 1983b
). The portion of an electrode located within a guide tube is, of course, constrained, but the electrode shaft lying between the top of the guide tube and the point of fixation to the microdrive is susceptible to buckling. The problem, then, is to constrain this portion of the shaft while still retaining the ability to lower the electrode. The "Eckhorn" microdrive system ( Eckhorn and Thomas 1993
) provides an elegant solution to the buckling problem for acute use of fine filaments. This system uses stretchable silastic tubing to both constrain buckling of the portion of the electrode that lies above the guide tube and to provide power to drive the electrode forward. This solution has advantages over the original clutch-based drive mechanism described by Reitboeck ( Mountcastle et al. 1991
; Reitboeck 1983b
). However, neither of these solutions are satisfactory for manipulation of chronically implanted electrodes. Moreover, very-fine-diameter filaments (
40 µm) will often buckle and break within the silastic tubing of the Eckhorn system.
Because of the preceding problems, an alternative method was developed to prevent the electrode from buckling using a "tube-over-tube" principal. Here, buckling is prevented along the length of the electrode by either the stainless steel guide tube or by a second tube, attached to the top of the electrode, that fits over the guide tube. Figure 2 illustrates the steps in the construction of these electrodes. Dimensions given are suitable for use when associated guide tubes are 22 mm long and electrode travel need extend 5 mm beyond the tip of the guide tube. After fabrication of electrodes is complete, impedance measures are made to insure electrical continuity (generally 13 M
at 1,000 Hz for electrodes aimed at recording from well-isolated single units). However, we have found visual inspection of the tips to be more useful than impedance measurements in predicting electrode performance.
|
The guide tube array (Fig. 1A, a) is constructed of 33-gauge stainless steel tubing (100 µm ID), the walls of which have been thinned to
25 µm (
150 µm OD, special order, Small Parts). Distal ends are bound together in a concentric array by tight-fitting polyimide tubing and a heat-curing insulating resin (Epoxylite, St. Louis, MO) is used to secure the tubes together. The combined tips are slightly beveled and cleaned after construction is complete. Cutting and beveling of stainless steel tubing is accomplished using a small diamond grinding wheel (0.1 mm thick, Abrasive Technology, Lewis Center, OH). The upper portion of the guide tubes are separated and inserted through matching holes drilled into a plastic template (Fig. 1A, c). Each of the guide tubes has an associated (somewhat larger) hole that has been predrilled to accommodate the base of a microdrive.2 The tops of the guide tubes are indicated by the dashed arrow in Fig. 1A.
Preventing blockage of the guide tubes
It is crucial that fluids from the brain be prevented from entering the guide tubes. Otherwise, solidification of the fluids would soon prevent the movement of the electrodes. To deal with this problem, we first fill each guide tube with a single application of a sterile antibiotic ointment (Vetropolycin, a bacitracin-neomycin-polymyxin opthalamic ointment). This is accomplished by pressure injection, until the ointment can be seen emerging from the distal ends of the tubes. Next, we create a hydrophobic seal at the tips with melted bone wax. The microelectrodes readily penetrate a few hundred micrometers of the bone wax, but the wax must be securely positioned and held within/over the tips of the guide tubes. Otherwise, the electrodes could displace rather than penetrate the wax. We have used several means to achieve this. Most recently, for studies of cortex, we have secured a length of polyimide tubing over the guide tube array (using Vetbond, a biocompatible cynoacralate glue) to create a space at the tips of
0.25 mm. After filling the guide tubes with antibiotic ointment, with the electrodes retracted within the guide tubes, the space at the tip of the guide tube is filled with melted bone wax (
40 nl of bone wax is required to fill this space). As shown in Fig. 1B, the microelectrodes readily penetrate this amount of bone wax.
Microdrives
Many systems of simple, screw-driven microdrives have been developed for chronic recording purposes. To be suitable for the preceding electrode-guide tube arrays, the microdrive must simply provide the appropriate linear motion along the z axis (36 mm for our purposes) and be narrow enough to allow one microdrive to be aligned with each guide tube without overcrowding. The requirement for such close spacing of microdrives presents a challenge. Two components are generally found in screw-driven microdrives: a screw and a linked shaft assembly, located parallel to the screw, that constrains motion to a single (z) axis. To reduce the x-y dimensions of the microdrive so that drives could be more densely packed, this general design principle was modified and the parallel shaft assembly was eliminated. Figure 3 is a schematic illustration of this microdrive and shows its basic design principle. A fine stainless steel screw (0000160, J.I. Morris, Southbridge, MA) is housed within a cylinder (made from 22-gauge, thin-wall stainless steel tubing, Small Parts) that has a nut (matching the 0000160 screw) soldered to the top. A cross-bar is soldered to the top of the screw so that it can easily be turned. The screw is threaded through the nut, and its distal, flattened tip pushes down on a piston, which has an armature that exits the cylinder at approximately a 90° angle via a groove along the length of the cylinder. This groove serves to constrain the motion of the armature to a single (z) axis. The armature will later be secured to the microelectrode and will control its motion. The piston is held tightly against the base of the screw by a small compression spring that puts an upward pressure on the armature.3 The microdrive is light (
60 mg) and is narrow enough (
1 mm diameter) so that microdrives can be spaced at distances of
1.3 mm.
|
Prior to use, the guide tube array is disinfected in 70% alcohol. Because liquids do not readily enter such fine tubing, the alcohol is pressure injected into each tube and then placed within the bath. After disinfection, the guide tubes are filled, by pressure injection, with the antibiotic ointment, and melted bone wax is applied to the tips of the guide tubes to create the hydrophobic seal. The electrodes are usually inserted and attached to the microdrives prior to the day of surgical implantation. Electrodes are inserted into the guide tubes under microscopic control and manually advanced until the polyimide tubing over the upper portion of the electrode passes over the stainless steel guide tube. The upper portion of the electrode (the polyimide tubing) is then secured to the microdrive armature using a small amount of melted dental impression compound (illustrated in Fig. 3J, right). This juncture is rigid but can be readily broken to re-position or exchange an electrode (electrodes can be exchanged either before or after the array has been implanted). The electrode is then advanced using the microdrive until the electrode tip is observed to emerge from the bone wax at the tip of the guide tubes. The electrode is then retracted a known distance (100 or 200 µm) into the bone wax. This procedure is repeated for each of the seven electrodes. The electrodes can also be placed into the guide tubes after the array has been surgically fixed into position, but the position of the electrode tips are less accurately known when using this procedure.
Implanting the array and fixing it to the skull
The array is mounted on a stainless steel rod that can be held by a stereotaxic carrier. The rod has a break-point near the junction with the array so that it can be released after being cemented to the skull. For cortical recordings, the bone wax at the tips of the guide tubes is placed in contact with the dura. For thalamic recordings, we usually insert the guide tubes through overlying (nonrelated) cortex so that the tips lie 34 mm above the region under investigation. Antibiotic ointment is applied to the surface of the dura, and the array is cemented into place using acrylic cement. This cement is applied liberally over the skull and around the guide tubes and is joined to the cement around the head-bolt assembly to ensure stability. Further stability is achieved by creating a cement bridge that joins the acrylic mass on the skull to the position on the array where the guide tubes are beginning to separate (
10 mm above the skull). Electrical contact between the electrodes and a miniature plug is made by soldering a length of Teflon coated Platinum-iridium wire (25 or 50 µm core diameter) to the platinum rod at the top end of each electrode. Our seven-channel microdrive system (as shown in Fig. 1) generally extends 3035 mm above the level of the skull. A molded plastic cap that attaches to the head bolt protects the array between recording sessions.
| RESULTS |
|---|
|
|
|---|
|
|
|
0.5 µm/yr.
The rabbit described above was killed at the age of 5.75 years, after the guide tube array had been in place for 4 yr and 5 mo. The rabbit was perfused with saline followed by formalin, and the tissue was sectioned and stained for Nissl substance. This array was implanted a little deeper than we had planned, and the tips of the guide tubes inadvertently penetrated the dorsal thalamus (Fig. 6A). Unfortunately, the preparation of this tissue was less than ideal, the plane of our tissue sections were not well-aligned with the angle of the guide tubes, and many freezing (and other) artifacts are present. Nevertheless, aside from the mechanical damage done by the guide tubes (total outer diameter of the array was
0.5 mm, as in Fig. 1B), damage to tissue around the tips of the guide tubes (which had been filled with bone wax) seemed minimal. No unambiguous signs of the electrode tracks could be followed to VB thalamus (Fig. 6B), where recordings were obtained.
|
5 mm beneath the dural surface of the rabbit), the tip of the guide tube array rested on the dural surface. Because extensive tissue reaction and growth often occurs on the dura, we were concerned that the guide tubes might be quickly blocked, with a consequent loss of electrode mobility. This occurred only very rarely. Figure 7 (top) shows receptive fields generated in superficial superior colliculus by five microelectrodes within 1 wk after this electrode array was implanted. These electrodes were of low impedance (<1 M
), designed to be optimal for multiunit and field potential recordings, as well as for microstimulation. Figure 7, bottom, shows the receptive fields recorded from these same electrodes just >1 yr later. Receptive fields were periodically monitored during the intervening months, and between these measures, the electrode tips were usually moved to a position
1 mm above the collicular surface. | DISCUSSION |
|---|
|
|
|---|
0.5 µm/yr. And 3) the very fine maximal outer diameter of the Reitboeck electrodes allow them to be funneled down very fine-diameter guide tubes for very close electrode spacing, yet their stiffness generates a straight and predictable electrode trajectory.
In all of these experiments, the rabbits were awake, but the head was fixed during recordings. Thus these were not freely behaving subjects, and our experimental aims have generally been to record from well-isolated single neurons for several hours and then to move on to record from other such neurons. Because these electrodes can be fabricated with very fine tips, they are well suited to record from neurons of all sizes. Although long-term recordings from the same neuron was not the goal of these experiments, we sometimes did record from a neuron for several days (e.g., see neuron "Hercules" in Swadlow and Gusev 2002
). Low-impedance, fixed microwires are undoubtedly better suited when it is desirable to study the same neuron for very long periods of time (e.g., Porada et al. 2000
; Swadlow 1982
, 1985
).
We have used different variations of this system in rabbits for >5 yr and have varied the seven-channel concentric design described here in several ways. For cortical recordings, we have employed three-channel triangular arrays, seven-channel concentric arrays, or five to seven channel linear arrays of guide tubes and electrodes (most with 150-µm spacing as described in the preceding text). We have also employed 19-channel8 concentric arrays of guide tubes and electrodes for recording both from the thalamus and from the superior colliculus. Somewhat longer guide tubes (2730 mm) are required for such large arrays to allow adequate spacing for the microdrives.
To prevent fluids from entering and solidifying within the guide tubes, we used small quantity of bone wax to create a hydrophobic seal at the tip of the guide tube. In our initial studies, we sealed the tips of the guide tubes so that a small amount of hot bone wax entered the tubes. In later studies, we expanded this hydrophobic seal by creating a wax-filled chamber, containing
40 nl of bone wax, at the tips of the guide tubes (as in Fig. 1B). Whereas bone wax is relatively biocompatible, there are some reports of long-term effects of this substance ( Aksu et al. 2001
; Alberius et al. 1987
; Allison 1994
). Moreover, bone wax is resorbed, albeit very slowly. Undoubtedly, there are better ways of creating a hydrophobic seal at the guide tube-brain interface that is penetrable by fine microelectrodes. Of course, it is generally prudent to position the tips of the guide tubes as far as possible from the tissue under investigation (e.g., on the surface of the dura when this is possible).
Chronically implanted fixed microwires can remain within the cortex and, in some cases, record from the same neurons for periods in excess of 1 yr (e.g., Porada et al. 2000
; Swadlow 1982, 1985
). For many experimental aims, however, microwires do not yield optimal recordings. The present work presents a system that enables the use and control of fine-diameter, precisely configured and mobile Reitboeck microelectrodes ( Reitboeck 1983a
) within the brain for periods of many months and even years. It is important to know if the system described here could be useful in recording from larger brains, which may not be as stable, relative to fixed skull coordinates, as the rabbit brains studied here. Preliminary experiments in a primate ( Chen et al. 2004
), in which a three-channel version of the preceding system was used successfully for a 3-mo period, suggest the affirmative.
| GRANTS |
|---|
|
|
|---|
| FOOTNOTES |
|---|
1 We grind electrode tips using a continuously wetted, "extra fine" diamond abrasive plate (Sutter Instrument, Novato, CA, plate model 104 F), that is rotating toward the electrode tip (at 13 revolutions/s) on a homemade horizontal turntable. The electrode is held within a length of stainless steel tubing at an angle (to the grinding wheel) of
12°. During grinding, the electrode is rotated about its axis (at
0.3 revolutions/s) using a toy motor that is linked to the electrode by a rubber band. A complete grinding system is commercially available from Thomas Recording. ![]()
2 Another means to control the position of the guide tubes is to fit them into slots that are ground into the side of the template (rather than into drilled holes). It is easier to fit the guide tubes into these slots, but the upper ends of the guide tubes are generally more divergent when using this method. ![]()
3 An alternative means of maintaining upward tension on the piston is through the use of a very fine band of elastic material that pulls upward on the armature and is attached to the microdrive near the nut. ![]()
4 One of the seven microdrives ceased to function due to an inadequate solder joint between the nut and the stainless steel tube (Fig 3). ![]()
5 The bent tip in electrode "C" is clearly the result of a mechanical accident. We do not know when this occurred but, based on the fresh appearance of the fractured quartz in the electron micrograph (Fig. 4C2), we guess is that it probably happened after these recordings were obtained as the electrode was removed from the guide tube. ![]()
6 These receptive fields were concentrically organized and only the on-center responses are shown. The fields (and those in Fig. 7) were generated using methods of reverse correlation. Stimuli consisted of 1 or 2° flashing light spots, presented pseudo randomly in a spatial grid of
16 x 16. ![]()
7 By the end of the initial year after implantation, one of the seven microdrives was broken and another of the electrodes was frozen within its guide tube. By the end of the fourth year, two additional electrodes were immobile, but three of the seven electrodes could still be moved and were fully functional. ![]()
8 Due to geometric constraints, concentric arrays of 7 or 19 channels (with 1 ring or 2 rings of guide tubes, respectively, around the center tube) are most easily fabricated when all guide tubes are of the same outer diameter. ![]()
Address for reprint requests and other correspondence: H. A. Swadlow, Dept. of Psychology (U-1020), The University of Connecticut, Storrs, CT 06269 (E-mail: Swadlow{at}psych.psy.uconn.edu)
| REFERENCES |
|---|
|
|
|---|
Alberius P, Klinge B, and Sjogren S. Effects of bone wax on rabbit cranial bone lesions. Craniomaxillofacial Surg 15: 6367, 1987.
Allison RT. Foreign body reactions and an associated histological artefact due to bone wax. Br J Biomed Sci 51: 1417, 1994.[ISI][Medline]
Alonso JM, Usrey WM, and Reid RC. Rules of connectivity between geniculate cells and simple cells in cat primary visual cortex. J Neurosci 21: 40024015, 2001.
Baker SN, Philbin N, Spinks R, Pinches EM, Wolpert DM, MacManus DM, Pauluis Q, and Lemon RN. Multiple single unit recording in the cortex of monkeys using independently moveable microelectrodes. J Neurosci Methods 15: 517, 1999.
Bezdudnaya T, Cano M, Alonso JM, and Swadlow HA. Temporal frequency tuning and arousal in awake LGN. Soc Neurosci Abstr 551.4, 2003
Cano M, Bezdudnaya T, Swadlow HA, and Alonso JM. How changes in arousal affect visual contrast sensitivity. Soc Neurosci Abstr 551.5, 2003
Chen Y, Martinez-Conde S, Macknik S, Swadlow HA, and Alonso JM. Task difficulty modulates neuronal activity in primary visual cortex. Soc Neurosci Abstr 175.7, 2004.
DiCarlo JJ, Johnson KO, and Hsiao SS. Structure of receptive fields in area 3b of primary somatosensory cortex in the alert monkey. J Neurosci 18: 26262645, 1998.
Eckhorn R, Bauer R, Jordan W, Brosch M, Kruse W, Munk M, and Reitboeck HJ. Coherent oscillations: a mechanism of feature linking in the visual cortex? Mulitple electrode and correlation analyses in the cat. Biol Cybern 60: 121130, 1988.[CrossRef][ISI][Medline]
Eckhorn R and Thomas U. A new method for the insertion of multiple microprobes into neural and muscular tissue, including fiber electrodes, fine wires, needles and microsensors. J Neurosci Methods 49: 175179, 1993.[CrossRef][ISI][Medline]
Gail A, Brinksmeyer HJ, and Eckhorn R. Perception-related modulations of local field potential power and coherence in primary visual cortex of awake monkey during binocular rivalry. Cereb Cortex 14: 300313, 2004.
Lee D. Behavioral context and coherent oscillations in supplementary motor area. J Neurosci 24: 44534459, 2004.
Lee D, Port NL, Knuse W, and Georgopoulos AP. Neuronal population coding: multielectrode recording in primate cerebral cortex. In: Neuronal Ensembles: Strategies for Recording and Decoding, edited by Eichenbaum H and Davis J. New York: New York, 1998, p. 117136.
Malpeli JG, Weyand TG, and LaClair R. A new method of mounting and directing chronically implanted microdrives. J Neurosci Methods 44: 1926, 1992.[CrossRef][ISI][Medline]
Mountcastle VB, Reitboeck HJ, Poggio GF, and Steinmetz MA. Adaptation of the Reitboeck method of multiple microelectrode recording to the neocortex of the waking monkey. J Neurosci Methods 36: 7784, 1991.[CrossRef][ISI][Medline]
Porada I, Bondar I, Spatz WB, and Kruger J. Rabbit and monkey visual cortex: more than a year of recording with up to 64 microelectrodes. J Neurosci Methods 95: 1328, 2000.[CrossRef][ISI][Medline]
Reitboeck HJ. Fiber microelectrodes for electrophysiological recordings. J Neurosci Methods 8: 249262, 1983a.[CrossRef][ISI][Medline]
Reitboeck HJ. A 19-channel matrix drive with individually controllable fiber microelectrodes for neurophysiological applications. IEEE Trans Syst Man Cybern 113: 626683, 1983b.
Steinmetz PN, Roy A, Fitzgerald PJ, Hsiao SS, Johnson KO, and Niebur E. Attention modulates synchronized neuronal firing in somatosensory cortex. Nature 404: 187190, 2000.[CrossRef][Medline]
Swadlow HA. Impulse conduction in the mammalian brain: physiological properties of individual axons monitored for several months. Science 218: 911913, 1982.
Swadlow HA. Physiological properties of individual cerebral axons studied in vivo for as long as one year. J Neurophysiol 54: 13461362, 1985.
Swadlow HA and Gusev AG. The influence of single VB thalamocortical impulses on barrel columns of rabbit somatosensory cortex. J Neurophysiol 83: 28032813, 2000.
Swadlow HA and Gusev AG. The impact of "bursting" thalamic impulses at a neocortical synapse. Nat Neurosci 4: 402408, 2001.[CrossRef][ISI][Medline]
Swadlow HA and Gusev AG. Receptive field construction in cortical inhibitory interneurons. Nat Neurosci 5: 403404, 2002.[CrossRef][ISI][Medline]
Swadlow HA, Gusev AG, and Bezdudnaya T. Activation of a cortical column by a thalamocortical impulse. J Neurosci 22: 77667773, 2002.
Usrey WM, Alonso JM, and Reid RC. Synaptic interactions between thalamic inputs to simple cells in cat visual cortex. J Neurosci 20: 54615467, 2000.
This article has been cited by other articles:
![]() |
C. R. Stoelzel, Y. Bereshpolova, A. G. Gusev, and H. A. Swadlow The Impact of an LGNd Impulse on the Awake Visual Cortex: Synaptic Dynamics and the Sustained/Transient Distinction J. Neurosci., May 7, 2008; 28(19): 5018 - 5028. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Soteropoulos and S. N. Baker Bilateral representation in the deep cerebellar nuclei J. Physiol., February 15, 2008; 586(4): 1117 - 1136. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Jackson and E. E. Fetz Compact Movable Microwire Array for Long-Term Chronic Unit Recording in Cerebral Cortex of Primates J Neurophysiol, November 1, 2007; 98(5): 3109 - 3118. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |