|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
TRANSLATIONAL PHYSIOLOGY
1Department of Neurology, Rambam Medical Center, Haifa; and 2Bruce Rappaport Faculty of Medicine and Research Institute, Haifa, Israel
Submitted 13 May 2006; accepted in final form 29 November 2006
|
|
ABSTRACT |
|---|
|
|
|
INTRODUCTION |
|---|
|
30% of all patients with epilepsy suffer from drug-resistant epilepsy and continue to experience seizures despite adequate AED treatment. Because of the ongoing recurrence of seizures, these patients suffer from a devastating disease with severe implications on the longevity and quality of life (Brodie and French 2001
Previous studies in hippocampal brain slices in vitro and in human patients with temporal lobe epilepsy have shown that cortical electrical stimulation can eliminate epileptiform discharges, including epileptic seizures. Most previous studies have used open-loop stimulation paradigms (Barbarosie and Avoli 1997
; Bikson et al. 2001
; Cuellar-Herrera et al. 2004
; Khosravani et al. 2003
; Kinoshita et al. 2004
; Krauss and Gordon 1999
; Lian et al. 2003
; Theodore and Fisher 2004
; Velasco et al. 2000
, 2001
; Vonck et al. 2002
). However, recently, few preliminary studies tested the ability of closed-loop stimulation to prematurely terminate seizures in epilepsy patients (Fountas et al. 2005
; Kossoff et al. 2004
; Osorio et al. 2005
). In addition, Lesser et al. (1997)
have reported neocortical electrical stimulation eliminated evoked afterdischarges in patients with intractable extratemporal neocortical epilepsy.
Despite growing evidence regarding the antiepileptic effect of electrical stimulation, the use of electrical stimulation for treating intractable epilepsy is only in its early stages of development. The studies performed thus far tested a small number of patients in a nonblinded manner and used mostly acute short-term rather than chronic long-term stimulation protocols. Moreover, the efficacy of stimulation reported thus far in the literature was probably not sufficient for clinical use (Cuellar-Herrera et al. 2004
; Vonck et al. 2002
), possibly because of the fact that the stimulation parameters were not optimized in these studies. In addition, despite growing interest in the potential use of electrical stimulation for treating various neurological diseases including intractable epilepsy, the mechanisms underlying the antiepileptic effect of stimulation remain unclear. Three major candidate mechanisms have been proposed: reduction in the excitability of neurons, increased inhibitory neurotransmission, and depression of excitatory neurotransmission (for review, see Durand and Bikson 2001
; McIntyre et al. 2004b
).
Neocortical brain slices treated with bicuculline (BCC) or magnesium-free solutions serve as acute models of neocortical epilepsy in vitro, because they produce both interictal-like depolarizing paroxysmal shift (PDS) discharges and seizure-like events (Avoli et al. 1991
; Hablitz 1987
; Schiller 2002
, 2004
; Valenzuela and Benardo 1995
). In this study, we used extracellular field potential measurements and whole cell intracellular voltage recordings from neocortical brain slices treated with either BCC or magnesium-free solution to investigate the antiepileptic effects of two stimulation paradigms: sustained low-frequency stimulation (0.15 Hz for 5 min or longer), which is best suited for prolonged open-loop stimulation, and short trains of high-frequency stimulation (25200 Hz for 15 s), which is best suited for closed-loop stimulation. We will concentrate on three main questions. First, can neocortical epileptiform discharges be eliminated by two stimulation paradigms sustained low-frequency and short trains of high-frequency electrical stimulation? Second, what are the optimal stimulation parameters for obtaining maximal antiepileptic effects with the stimulation paradigms? Third, what are the cellular mechanisms underlying the antiepileptic effects of cortical electrical stimulation?
|
|
METHODS |
|---|
|
Parasagittal neocortical brain slices (300400 µm in thickness) were prepared from 13- to 35-day-old Wistar rats, as previously described (Schiller 2004
). The slice and single neurons were visualized using infrared illumination and differential interference contrast optics (IR-DIC) video microscopy. The microscope used was a fixed stage BX-51WI (Olympus). Extracellular field potential recordings and whole cell voltage recordings from the soma of single neurons were obtained using the Multi-clamp 700A amplifier (Axon Instruments, Foster City, CA). Extracellular recordings, which were performed with glass pipettes (0.10.5 MOhm) filled with the extracellular solution, were amplified 1,000-fold and filtered with a low-pass filter of 3 KHz and high-pass filter of 1 Hz. Intracellular whole cell recordings were performed from the soma of layer 5 and layer 2/3 pyramidal neurons as previously described (Schiller 2002
; Schiller et al. 2000
). The slice was bathed in artificial cerebrospinal fluid (ACSF) that contained (in mM) 125 NaCl, 25 NaHCO3, 25 glucose, 4 KCl, 1.25 NaH2PO4, 1.5 CaCl2, and 1 MgCl2, 0.01 BCC, or 0 MgCl2; pH 7.4. Somatic (3- to 6-MOhm resistance) whole cell recording pipettes were filled with (in mM) 115 K-gluconate, 20 KCl, 2 Mg-ATP, 2 Na2-ATP, 10 Na2-phosphocreatine, 0.3 GTP, 10 HEPES, pH 7.2, and either 0.1 calcium green-1 or 0.1 1,2-bis(2-aminophenoxy)ethane-N,N,N,N-tetraacetic acid tetrakis (BAPTA). The experiments were performed at 35 ± 0.5°C. All chemicals were purchased from Sigma except for calcium green-1 (Molecular Probes, Portland, OR), 2-amino-5-phosphonopentanoic acid (APV), 6-cyano-nitroquinoxaline-2,3 dione (CNQX), and BCC (Tocris, Bristol, UK).
In all intracellular experiments, we continuously monitored the resting membrane potential and repeatedly tested the series resistance and the amplitude and shape of action potentials evoked by depolarizing somatic current injections. The recording was discontinued when the resting membrane potential changed by >3 mV, when the series resistance was >25 M
or changed by >30%, or when the amplitude or shape of action potentials evoked by somatic current injection changed. In addition, the position of the intra- and extracellular electrodes was monitored visually using the IR-DIC optics.
In this study, brain slices treated with either BCC or magnesium-free extracellular solution produced both interictal-like discharges and seizure-like events. Interictal-like discharges were defined as events of intense firing (usually consisting of 12 PDS discharges), lasting <0.5 s. Seizure-like events were defined as events of intense firing lasting >2 s. Events lasting 0.52 s were left undefined.
Electrical stimulation
Electrical stimulation was administered using a Master-8 stimulator and Iso-flex isolators (AMPI, Jerusalem, Israel). We used two different types of stimulating electrodes: double barrel theta patch pipettes containing two silver wires, one in each barrel and filled with ACSF (composition of ACSF described above) (Polsky et al. 2004
; Schiller et al. 2000
), and platinum/iridium metal microelectrodes with impedance of 0.20.5 MOhm (Microprobe, Gaithersburg, MD). In both cases, a negative current was administered against a ground electrode. In the case of the double barrel theta pipettes, one wire served as active electrode while the other wire served as the ground electrode, whereas in the case of metal microelectrodes, a separate chlorinated silver wire was inserted into the bath and served as the ground electrode. The results obtained with the two types of stimulating electrodes were analyzed together, because no differences between them were observed. In all experiments, we stimulated the slice with depolarizing square pulses lasting 0.20.4 ms. In all experiments, we first characterized the threshold for initiation of the PDS response in the slice, and the stimulus intensity used for stimulation was 2.5- to 3-fold the PDS threshold unless specifically stated. In all experiments, we applied only a single sustained low-frequency stimulation train per slice aside from experiments in which multiple data points were compared in the same slice. These included experiments examining the effects of stimulus frequency, stimulus intensity, and stimulation site, where no more than five consecutive stimulation sessions were applied per slice. In these experiments, we waited
30 min between stimulation sessions, and the order of stimulation sessions was randomly chosen. In experiments where two stimulating electrodes were used, the interelectrode distance between the two stimulating electrodes was measured separately in the horizontal and vertical directions.
All experiments were performed after slices were prebathed in the magnesium-free solution or the BCC-containing solution for
1.5 h, and spontaneous epileptiform discharges were recorded at a minimal rate of 1.5 discharges/min. Slices that produced spontaneous discharges at lower rates (<1.5 discharges/min) were excluded. For experiments that examined the effect of stimulation on excitatory postsynaptic potentials (EPSPs), we chose slices with a relatively low frequency of spontaneous epileptiform discharges (1.52.5 epileptiform discharges/min).
Data analysis
The data were analyzed using Igor (WaveMetrics, Lake Oswego, OR) and Excel software and presented in the form of average and SD values. To measure the amplitude of epileptiform discharges or EPSPs with over-riding action potentials, the waveform was filtered with a low-pass filter of 300 Hz to filter out the over-riding fast action potentials. Statistical analysis was performed using the Student's t-test and ANOVA statistical tests using Excel and SPSS software.
|
|
RESULTS |
|---|
|
To study the effect of sustained low-frequency electrical stimulation on epileptiform discharges, we stimulated BCC-treated neocortical brain slices at 0.15 Hz for 545 min. Figures 1 and 2 show two typical examples where the slice was stimulated at 1 Hz. During the control prestimulation period, the slice spontaneously produced both interictal-like discharges and seizure-like events (Hablitz 1987
; Schiller 2002
, 2004
). At the onset of the 1-Hz electrical stimulation, PDS discharges were evoked. However, as stimulation progressed, the epileptiform discharges gradually attenuated, and with time, epileptiform discharges disappeared altogether (Figs. 1 and 2). The effect of stimulation was reversible. On discontinuation of stimulation, evoked and spontaneous epileptiform discharges gradually recovered (Fig. 1C, data not shown for spontaneous events).
|
|
45 min) in all recorded slices. In all recorded slices, evoked epileptiform discharges gradually recovered within the first 12 min after discontinuation of the 1-Hz stimulation, whereas spontaneous PDS discharges and seizure-like events reappeared within 10 min after discontinuation of sustained stimulation in 32 of the 36 slices examined (89%). We next characterized the stimulation parameters influencing the antiepileptic effect of sustained low-frequency electrical stimulation. We were especially interested in examining the effect of stimulation frequency, intensity, and location on the antiepileptic effect of stimulation.
The location of the stimulating electrode did not influence the antiepileptic efficiency of the 1-Hz sustained stimulation. No differences were observed in eliminating epileptiform discharges when we changed the location of the stimulating electrode between neocortical layer 5, layer 2/3, and subcortical white matter (n = 5, data not shown). In contrast to the location of the stimulating electrode, the frequency of simulation markedly influenced the magnitude and time-course of suppression of epileptiform discharges (Fig. 3A). In all seven slices examined, attenuation of epileptiform discharges was first observed at stimulation frequencies of 0.330.5 Hz. The magnitude of attenuation increased with stimulation frequency until responses completely disappeared at stimulation frequency of 0.52 Hz. The time-course of attenuation was faster as the stimulation frequency was increased (examined
5 Hz). Similar to our findings in neocortical slices, in hippocampal slices, stimulation frequencies <0.3 Hz were ineffective in prevention of seizures (Khosravani et al. 2003
).
|
It is interesting to note that, despite the spatial constraints of electrical stimulation, sustained 1-Hz stimulation at a single site was sufficient to prevent spontaneous epileptiform discharges in the vast majority of slices. Hence initiation of spontaneous epileptiform discharges probably involved a larger network of neurons, and the antiepileptic effect induced by stimulation in part of this network was sufficient to prevent spontaneous PDS discharges and seizure-like events.
As pointed out earlier, seizure-like events were evoked at the onset of sustained low-frequency stimulation in 42% of cases. We attempted to overcome this potential drawback by gradually increasing the stimulus intensity at the onset of stimulation. In these experiments, slices were stimulated initially at intensities equal to one half the threshold for PDS initiation. Later the stimulus intensity was gradually increased by 30% every 1 min until it reached 2.5-fold the PDS threshold. Under these conditions, no seizure-like events were observed at the onset of stimulation. Instead, only interictal-like PDS discharges were evoked during the initial stages of stimulation. These findings were observed in eight slices. It is important to stress that, in human patients, interictal discharges usually are of no clinical significance, and the main concern is from initiation of seizures by stimulation.
We further studied the antiepileptic effects of sustained electrical stimulation in a second model of acute epilepsy, neocortical brain slices bathed in a magnesium-free extracellular solution. Similar to BCC-treated slices, sustained 1-Hz stimulation suppressed both evoked and spontaneous epileptiform discharges in magnesium free-treated neocortical slices. Figure 4, A and B, shows one such typical experiment. At the onset of stimulation, an epileptiform discharge was evoked. However, as the stimulation progressed, responses gradually attenuated until only EPSPs were evoked. Concomitantly spontaneous epileptiform discharges, which were abundant before the sustain stimulation, disappeared altogether after 280 s of stimulation. Similar experiments were performed in 10 additional neurons. In all these 11 experiments, epileptiform discharges were evoked at the onset of stimulation and were gradually replaced by either sub- or suprathreshold EPSPs. Concomitantly spontaneous epileptiform discharges disappeared altogether in 6 of 11 slices, and in the remaining 5 slices occurred infrequently throughout the 15-min sustained stimulation. On discontinuation of the 1-Hz stimulation, epileptiform discharges rapidly recovered (Fig. 4B).
|
The antiepileptic effects of sustained stimulation in magnesium freetreated slices were location dependent. In these experiments, we used two stimulating electrodes. One electrode stimulated the slice continuously at 1 Hz, and the other electrode was used to stimulate the slice infrequently (2 stimuli/min). To study the spatial efficacy of sustained stimulation, the distance between the two stimulating electrodes was changed during the experiments (for more details, see description of this paradigm in BCC-treated slices). As shown in Fig. 4C, the efficacy of stimulation decreased as the distance between the two stimulating electrodes increased. Sustained stimulation effectively eliminated epileptiform discharges at interelectrode distances of
100 µm. At larger interelectrode distances, the antiepileptic effects gradually decreased, and at interelectrode distances of 400600 µm or greater, sustained 1-Hz stimulation with one electrode did not have noticeable antiepileptic effects on epileptiform discharges evoked by the second, infrequently stimulating electrode (Fig. 4C). It is interesting to note that the spatial efficacy of stimulation in magnesium freetreated slices was smaller than that observed in BCC-treated slices (1,160 ± 305 µm for BCC, n = 5, compared with 490 ± 74 µm for magnesium free, n = 6).
Termination of seizure-like events by short closed-loop high-frequency electrical stimulation
In the previous section, we showed that sustained open-loop low-frequency electrical stimulation (0.335 Hz for 545 min) prevented interictal-like discharges and seizure-like events. A second potential stimulation paradigm for treating epilepsy is a closed-loop stimulation. Ideally, these devices impending seizures will be detected and the appropriate stimulation will be applied to prevent initiation of full-blown clinical seizures. In our experiments, we were unable to detect impending seizures. Hence, we wanted to examine whether we are able to terminate ongoing seizures with short trains of high-frequency electrical stimulation (25200 Hz for 15 s). We used high-frequency stimulation trains in these experiments for two main reasons. First, in contrast to low-frequency stimulation, with its slow onset time and initial excitatory effect, high-frequency stimulation seems more likely to be more appropriate for closed-loop stimulation. Second, high-frequency stimulation paradigms have been used in previous and ongoing human stimulation trials (Fountas et al. 2005
; Kossoff et al. 2004
; Lesser et al. 1997
; Osorio et al. 2005
). We limited the duration of high-frequency stimulation to 5 s, because longer stimulation hampered our ability to electrophysiologically monitor the duration of seizures. To study the antiepileptic effects of short trains of high-frequency stimulation, we used two different experimental designs. In the first, we evoked seizure-like events with one electrode, and 12 s later we stimulated the slice at 100 Hz for 13 s with a second electrode (Fig. 5A). In the second experimental design, we waited for initiation of spontaneous seizure-like events. Once such an event was visually identified, we manually applied the 1- to 3-s 100-Hz stimulation train (Fig. 5B). To ensure that stimulation was applied during seizure-like events rather than interictal-like PDS discharges, we only included cases in which stimulation was applied 0.73 s after seizure initiation.
|
We next wanted to study how different stimulation parameters influenced the ability of high-frequency stimulation to prematurely terminate seizure-like events. In our experiments, we studied four different stimulation parameters: intensity, duration, frequency, and location of stimulation. Figure 6 A shows the average percent of seizures terminated during a 2-s, 100-Hz stimulation train as a function of the stimulus intensity. To average the results of different slices, stimulus intensities were normalized to the PDS discharge threshold. Termination of seizure-like events was dependent on the stimulus intensity. Increasing the stimulus intensity
2.5-fold of the PDS threshold gradually increased the fraction of seizures terminated during the stimulation period. However, additional increase of the stimulus intensity from 2.5- to 7-fold of the PDS threshold did not further enhance termination of seizure-like events.
|
We next examined the effect of frequency on the antiepileptic effect of high-frequency stimulation. Four different stimulation frequencies were tested ranging from 25 to 200 Hz. Increasing the frequency from 25 to 50 Hz markedly increased the fraction of seizures that terminated during stimulation. However, further increasing the frequency from 50 to 100200 Hz did not further increase the fraction of seizures terminated during stimulation (Fig. 6C).
In contrast to the intensity, frequency, and duration of stimulation, termination of seizure-like events was insensitive to the site of high-frequency stimulation. We observed no significant differences in the percent of seizures terminated by a 2-s, 100-Hz stimulation when we changed location of the stimulating electrode between neocortical layer 2/3, layer 5, and subcortical white matter (20 seizures in layer 2/3 and 18 seizures in layer 5 in 6 slices; data not shown). Moreover, changing the interelectrode distance between the electrode that evoked seizure-like events and the electrode that applied the 2-s, 100-Hz stimulation (
1,500 µm horizontally) did not significantly change the percent of seizures terminated (Fig. 6D). The fact that seizure termination was unaffected by the interelectrode distance probably reflected the fact that, once seizure-like events initiated, they involved the entire neuronal network of the slice rather than the site of initiation.
To further study the antiepileptic effects of high-frequency stimulation, we repeated our experiments in magnesium freetreated neocortical brain slices. Magnesium freetreated neocortical brain slices only infrequently produced discharges lasting >3 s. Thus to study the antiepileptic effects of short trains of high-frequency stimulation, we evoked epileptiform discharges with one electrode and 200 ms later applied a 0.6-s, 100-Hz stimulation train with a second electrode. The distance between the two stimulating electrodes was 50100 µm in the horizontal direction. When we compared the duration of epileptiform discharges with and without the 100-Hz stimulation train, we found that epileptiform discharges were significantly shorter when the 0.6-s, 100-Hz stimulation train was administered. In these experiments, the average duration of epileptiform discharges decreased from 1.49 ± 0.72 s under control conditions (n = 48) to 1.08 ± 0.46 s (n = 42) during stimulation (experiments were performed in 10 slices, P < 0.01). Moreover, 22 of 42 epileptiform discharges terminated during the 0.6-s stimulation, in contrast to only 14 of 48 discharges that were shorter than 0.8 s under control conditions. Hence, our findings indicated that, similar to BCC-treated slices, short high-frequency stimulation trains prematurely terminated epileptiform discharges in magnesium freetreated neocortical brain slices.
Cellular mechanisms underlying the antiepileptic effect of low- and high-frequency electrical stimulation
In the previous sections, we showed that seizure-like events can be prevented by sustained low-frequency stimulation on the one hand and prematurely terminated by short trains of high-frequency stimulation on the other hand. We next studied the cellular mechanisms underlying the antiepileptic effect of electrical stimulation. In our experiments, we considered two potential mechanisms: reduction of neuronal excitability and depression of excitatory neurotransmission. In our experimental model, a third potential antiepileptic mechanism of enhanced inhibition was not relevant because GABAA receptors were pharmacologically blocked.
Effect of low- and high-frequency electrical stimulation on neuronal excitability
One mechanism by which electrical stimulation can exert its antiepileptic effect is by reducing the excitability of neurons or axons. Neurons in the stimulated neocortical slice can be divided into two main groups with respect to the effect of stimulation. The first group consists of neurons directly stimulated by the stimulating electrode, whereas the remaining neurons are indirectly activated by electrical stimulation through mono- or polysynaptic connections. The antiepileptic effects of stimulation can be mediated by decreased excitability of one or of both subgroups of directly stimulated and synaptically activated neurons. To study the effects of stimulation on excitability of the different neurons, we used four different experimental paradigms. First, we examined the ability of axons to sustain firing during stimulation. Second, we examined the ability of directly stimulated neurons (by sustained axonal stimulation) to fire action potentials in response to somatic depolarizing waveforms (inputoutput response curves). Third, we studied the ability of indirectly (synaptically) activated neurons to fire action potentials in response to somatic depolarizing waveforms. Fourth, we studied the antiepileptic effect of partial pharmacological blockade of voltage-gated sodium channel that in turn artificially decreased the excitability of neurons.
Ability of axons to sustain firing during low- and high-frequency electrical stimulation
In these experiments, we performed whole cell recordings from a neocortical pyramidal neuron and visually identifying the axon of the recorded neuron using fluorescent and DIC imaging. To directly stimulate the axon, we placed the stimulating electrode in close proximity to the identified axon (Fig. 7A). To eliminate ionotropic synaptic neurotransmission, we added CNQX (20 µM), APV (100 µM), and BCC (10 µM) to the bath solution and confirmed that no EPSPs or inhibitory postsynaptic potentials (IPSPs) were evoked in response to extracellular synaptic stimulation.
|
45 s, neurons reliably generated axonal action potentials after each stimulus. However, when the stimulation rate was increased to 100200 Hz (and in some neurons, 50 Hz as well), a fraction of action potentials failed to initiate within the first second of stimulation, as determined by somatic recordings. On average, during the third second of 100-Hz axonal stimulation, neurons generated only 45 ± 5 action potentials (n = 13, 7 layer 5 neurons and 6 layer 2/3 neurons). Further raising the stimulus intensity transiently increased the firing frequency, but it rapidly decreased again to the baseline value (data not shown). It is interesting to note that a recent modeling study raised the possibility that, during high-frequency axonal stimulation, action potentials failed to propagate to the soma while successfully propagating along the axonal arborization (McIntyre et al. 2004aAbility to generate action potentials in response to depolarizing waveforms in directly stimulated neurons
When the axon of the neuron is directly stimulated, sustained action potential firing can reduce the neuron's ability to generate additional action potentials in response to incoming synaptic potentials. To study this possibility, we directly stimulated the visually identified axon in a similar manner to that described in the previous experiment and examined the response of the neurons to depolarizing somatic current injections before, during, and after the axonal stimulation. When we compared the number of action potentials evoked by 100-ms somatic depolarizing current injections of different amplitudes (inputoutput response curve) before and during the 1-Hz axonal stimulation, we found a small, yet insignificant, reduction in the number action potentials generated in response to the depolarizing current injection (Fig. 7, B and C; n = 16, 9 layer 5 and 7 layer 2/3 pyramidal neurons). All these small effects were reversible on discontinuation of electrical stimulation (data not shown). Hence sustained action potential firing at low frequencies did not significantly affect the ability of neurons to generate action potentials in response to incoming depolarizing waveforms.
In addition, we examined the effect of 100-Hz stimulation on the ability to generate action potentials in response to depolarizing somatic current injections. To test this question, 100-ms somatic depolarizing current injections were applied either during or immediately after a short train of 100-Hz stimulation. When the 100-ms depolarizing current injection was applied 2 s after the initiation of a 4-s, 100-Hz stimulation train, it was very difficult to evoke additional action potentials by somatic depolarizing current injections. Even when we injected large 100-ms somatic depolarizing currents, only one to two additional action potentials were produced (n = 15, 8 layer 5 and 7 layer 2/3 neurons). When the 100-ms depolarizing current injection was applied immediately after a 2-s, 100-Hz stimulation, the number of action potentials generated was markedly decreased (for 220 pA, the average number of action potentials decreased from 2.5 ± 0.9 before the 100-Hz stimulation to 1.7 ± 0.6 immediately after the 2-s, 100-Hz stimulation, P < 0.01).
Taken together, our findings indicated that, in directly stimulated neurons, sustained low-frequency stimulation did not significantly affect excitability of neurons, whereas short high-frequency stimulation trains impaired the ability of neurons to further respond to depolarizing waveforms.
Effects of sustained synaptic stimulation on the ability to generate action potentials in response to depolarizing waveforms
We next examined whether sustained 1-Hz extracellular synaptic stimulation impaired the ability of neurons to generate action potentials in response to depolarizing current injections. In these experiments, we examined the inputoutput response curve evoked by 100-ms depolarizing current injections of different amplitudes under control conditions and 1 and 10 min into sustained 1-Hz extracellular synaptic stimulation. After 1 min of sustained 1-Hz stimulation, one to two action potentials usually accompanied the synaptic responses, whereas 10 minutes into sustained 1-Hz stimulation, only subthreshold EPSPs remained (e.g., Figs. 2 and 4). Our experiments indicated that sustained 1-Hz stimulation had no effect on the number of action potentials evoked by the depolarizing current injections (inputoutput curve) either after 1 (data not shown) or 10 min of sustained 1-Hz stimulation (Fig. 7, D and E; 9 layer 5 neurons and 7 layer 2/3 neurons).
We next examined the effect of short trains of 100-Hz synaptic stimulation on generation of action potentials. We compared the number of action potentials generated by a 100-ms depolarizing current injection under control conditions 2 s after initiation of a 4-s, 100-Hz synaptic stimulation and immediately after a 2-s, 100-Hz stimulation train. We found that the 100-Hz stimulation slightly (but insignificantly) increased the number of action potentials generated, whereas immediately after the 2-s, 100-Hz stimulation train, no significant difference was observed in the number of action potentials generated compared with prestimulated control values. The 100-ms, 100- and 220-pA depolarizing current injections evoked 1.4 ± 0.5 and 2.8 ± 0.4 action potentials under control conditions, 1.6 ± 0.5 and 3.2 ± 0.8 action potentials during the 100-Hz stimulation (the 100-ms, 100- and 220-pA depolarizing current injections were applied 2 s after initiation of a 4-s, 100-Hz stimulation), and 1.3 ± 0.6 and 2.9 ± 1.0 immediately after the 100-Hz stimulation ended, respectively (n = 6, 4 layer 5 neurons and 2 layer 2/3 neurons; P > 0.15 for all 3 cases).
Taken together, our findings indicated that the excitability of neurons indirectly activated by synaptic inputs are unaffected by both sustained low-frequency stimulation and short trains of high-frequency stimulation.
Antiepileptic effects of partial pharmacological blockade of voltage-gated sodium channels
To study the potential antiepileptic effects of increasing the threshold for action potential initiation, we pharmacologically reduced the excitability of neurons by adding low concentrations (0.10.2 µM) of extracellular TTX to the bath solution. In these experiments, we first measured the rate of spontaneous epileptiform discharges and the threshold for axo-somatic action potentials initiation under control conditions. The threshold for axo-somatic action potential initiation in the recorded neuron was measured during 100-ms depolarizing current injections. Next we added 0.1 µM TTX to the bath solution and again measured the threshold for action potential initiation in the recorded neuron. If the threshold was increased by <2.5 mV, we further increased the concentration of TTX to 0.2 µM. Later we again measured the frequency of spontaneous epileptiform discharges and the response to a 5-min, 1-Hz sustained stimulation in the presence of TTX. The average frequency of spontaneous epileptiform discharges was 3.7 ± 0.7 under control conditions and 3.4 ± 0.9 after the addition of 0.10.2 µM TTX (Fig. 7F; n = 5, P > 0.25). In addition, suppression of epileptiform discharges during the 5-min, 1-Hz stimulation was not affected by 12 µM TTX (data not shown, n = 3). In these experiments, the average threshold for action potential initiation in the recorded neuron was increased by 3.9 ± 1.5 mV (n = 5). On further increasing the concentration of TTX to 1 µM to eliminate action potentials altogether, spontaneous and evoked epileptiform discharges disappeared altogether (Fig. 7F; n = 3). Hence mild reduction in excitability, as manifested by increased initiation threshold of axo-somatic action potentials, had no significant antiepileptic effects.
Taken together, the findings of our experiments in directly stimulated neurons, indirectly synaptically activated neurons, and with low concentration of TTX indicated that the antiepileptic effects of electrical stimulation were unlikely to be caused by a reduction in the excitability of neurons.
Effect of low- and high-frequency electrical stimulation on excitatory synaptic transmission
We next examined the effect of electrical stimulation on excitatory neurotransmission. In these experiments, we decreased the stimulus intensity below the threshold for epileptiform discharge initiation, such that either EPSPs or action potentials were evoked, and measured the effect of electrical stimulation on the amplitude of EPSPs or the probability to fire action potentials. Under these stimulation parameters, spontaneous epileptiform activity persisted. Thus we chose for these experiments slices with relative low rates of spontaneous epileptiform discharges (below 2 discharges/min) and performed measurements
20 s after epileptiform discharges ended. Figure 8 shows a typical experiment that examined the effect of sustained low-frequency stimulation on excitatory synaptic transmission. In this experiment, the slice was stimulated at 1 Hz with two different stimulus intensities: one evoked EPSPs (Fig. 8A) and the other evoked action potentials (Fig. 8B). Sustained low-frequency stimulation significantly attenuated excitatory synaptic transmission. The amplitude of EPSPs gradually decreased with stimulation (Fig. 8A), and at the higher stimulus intensity, above the threshold for action potential initiation, the 1-Hz sustained stimulation resulted in gradual failure of action potential firing with only infrequent action potentials observed as the stimulation progressed (Fig. 8B). After stimulation ended, the EPSP amplitude partially recovered and reached 93% of the prestimulus control value 30 min after stimulation was discontinued (P < 0.01 compared with the prestimulus control value). The fact that EPSPs only partially recovered after stimulation probably resulted from the induction of long-term depression (LTD) by the 15-min, 1-Hz stimulation (Froc et al. 2000
; Perrett et al. 2001
).
|
When the stimulus intensity was increased above the threshold for action potential initiation, only 2 ± 4% of stimuli evoked action potentials after 10 min of 1-Hz sustained stimulation compared with 100% during the prestimulus control period (n = 9).
To study the possibility that depression of EPSPs by sustained low-frequency stimulation was related to BCC, we performed similar experiments in brain slices bathed in ACSF without BCC. Similar to BCC-treated brain slices, 5 min of 1-Hz stimulation decreased the amplitude of EPSPs by 53 ± 14% (n = 4).
Depression of EPSPs by sustained low-frequency stimulation was dependent on the stimulation frequency. The higher the stimulation frequency (examined between 0.5 and 5 Hz), the greater the depression of EPSP amplitude. On average, after 5 min of stimulation at 0.5, 1, 2, and 5 Hz, the amplitude of EPSPs decreased by 32 ± 8 (n = 6), 61 ± 10 (n = 16), 67 ± 8 (n = 5), and 82 ± 12% (n = 5; P < 0.01 for comparison of 0.5 Hz with all other frequencies and comparison of 1 and 5 Hz; P < 0.05 for comparison of 2 and 5 Hz).
Similar to BCC-treated slices, in magnesium freetreated brain slices, sustained low-frequency stimulation also markedly depressed the amplitude of EPSPs. At 1-Hz stimulation, the average amplitude of EPSPs decreased by 59 ± 19 and 67 ± 16% after 5 and 10 min of stimulation (n = 5). Depression of the EPSP amplitude by sustained stimulation was frequency dependent. Very little depression of the EPSP amplitude was observed at 0.1-Hz stimulation, whereas at stimulation frequencies of 0.55 Hz, the amplitude of the EPSP was significantly depressed. Moreover, the higher the stimulation frequency, the bigger the depression of the average EPSP amplitude we observed (Fig. 9A).
|
We next examined the effect of higher concentrations of CNQX (10 µM) and APV (50 µM). In these experiments, the pharmacological blockers were washed in gradually, and thus glutamate receptors were progressively blocked over a period of 1520 min. In all eight slices examined, addition of 10 µM CNQX and 50 µM APV to the bath solution gradually eliminated spontaneous epileptiform discharges, and at a later stage, evoked epileptiform discharges as well. In three slices where whole cell recordings were performed from layer 5 pyramidal neurons, the average EPSP amplitude decreased by 46 ± 15% when spontaneous epileptiform discharges disappeared and by 58 ± 18% when evoked spontaneous discharges disappeared. Taken together, our findings indicated that depression of EPSPs was sufficient to eliminate epileptiform discharges.
We next characterized the effect of high-frequency stimulation trains on excitatory neurotransmission. Figure 10 shows the results of one such experiment. The stimulus intensity was decreased such that EPSPs rather than epileptiform discharges were evoked. During the 100-Hz stimulation, the EPSP amplitude rapidly and markedly decreased. Already after the fifth stimulus (40 ms of stimulation), the EPSP amplitude decreased by 77% compared with the prestimulus control EPSP, whereas after 2 s of 100-Hz stimulation, it decreased by 89 ± 2% (average of 10 consecutive EPSPs). The effect of the 100-Hz stimulation on EPSP amplitude was reversible, as shown in Fig. 10B. Fifteen minutes after stimulation ended, the amplitude of EPSPs returned to 96 ± 8% that of the prestimulus control EPSP (P = 0.3). Similar results to those shown in Fig. 10, A and B, were observed in all seven additional experiments performed. The average amplitude of EPSPs decreased by 86 ± 6 and 89 ± 4% after 1 and 2 s of 100-Hz stimulation, respectively (n = 8, 5 layer 5 and 3 layer 2/3 pyramidal neurons). Similar results were obtained in two additional experiments performed on brain slices bathed in ACSF without BCC. In addition, we repeated these experiments in magnesium freetreated brain slices with similar results. On average, after 1 s of 100-Hz stimulation, the amplitude of EPSPs decreased by 81 ± 12% (n = 3) in magnesium freetreated slices. We did not examine the effect of high-frequency stimulation in the presence of 2 µM CNQX and 10 µM APV, because the rate of spontaneous and evoked seizure-like events markedly decreased. Hence these experiments were virtually impossible to perform.
|
|
|
DISCUSSION |
|---|
|
We also examined the effects of stimulation on several aspects of excitability, including the ability of directly stimulated axons to sustain action potential firing, the ability to generate action potentials in response to depolarizing waveforms in directly and synaptically activated neurons, and the effects of pharmacologically compromising excitability by low doses of TTX on epileptiform discharges. We found that sustained low-frequency stimulation had no significant effects on excitability, whereas high-frequency stimulation only reduced the ability to generate high-frequency firing in directly stimulated neurons (not in synaptically activated neurons). Additional experiments with low doses of TTX showed that a mild reduction in the ability to generate axo-somatic action potentials had no significant antiepileptic effects, and generation of epileptiform discharges was compromised only when firing of axo-somatic action potentials was markedly suppressed. Taken together, our findings suggested that the antiepileptic effects of electrical stimulation were mediated by depression of excitatory neurotransmission, whereas reduction of excitability may somewhat contribute in high-frequency stimulation.
It is interesting to note that our findings are different from those reported for sinusoidal high-frequency electrical field stimulation in hippocampal brain slices where the antiepileptic effects were attributed to potassium efflux, depolarization of the membrane potential, and in turn, depolarization block of action potential firing (Bikson et al. 2001
). In our experiments, we found no evidence for depolarization of the membrane potential during low-frequency stimulation.
The antiepileptic effect of sustained open-loop low-frequency stimulation paradigms has been examined in the past in hippocampal brain slices in vitro and in human patients with mesial temporal and neocortical epilepsy. Application of both constant DC and sinusoidal electric fields has been shown to suppress epileptiform discharges in the hippocampus in vitro (Bikson et al. 2001
; Gluckman et al. 1996
; Lian et al. 2003
; Warren and Durand 1998
). In addition, low-frequency stimulation applied either continuously or intermittently also prevented seizure-like events in hippocampal brain slices in vitro (Albensi et al. 2004
; Barbarosie and Avoli 1997
; Jerger and Schiff 1995
; Khosravani et al. 2003
).
In patients with epilepsy, brief trains of high-frequency stimulation have been shown to terminate stimulus evoked afterdischarges in the neocortex (Lesser et al. 1999). In addition, continuous high-frequency stimulation administered to the hippocampus for 23 wk (Cuellar-Herrera et al. 2004
; Velasco et al. 2000
, 2001
) or 5 mo (Vonck et al. 2002
) decreased the number of seizures in a small number of patients suffering from intractable temporal lobe epilepsy.
Our study adds important additional information to that reported previously in the literature. The main new findings of this study are as follows. First, to the best of our knowledge, all previous in vitro studies were performed on hippocampal and entorhinal brain slices, and this is the first study that investigated the antiepileptic effect of electrical stimulation in extratemporal neocortical brain slices in vitro. Second, to the best of our knowledge, our study is the first to investigate the antiepileptic effects of closed-loop high-frequency stimulation in vitro. Third, in this study, we carefully characterized the parameters affecting the antiepileptic effects of stimulation. Hence our findings can be used to optimize the stimulation parameters. Fourth, and most importantly, our study shows for the first time that the antiepileptic effects of stimulation are mediated mostly by synaptic depression of glutaminergic synapses.
One of the main goals of this study was to define the stimulation parameters most effective in seizure prevention. We found that efficient antiepileptic low-frequency stimulation required stimulation frequencies of 0.5 Hz and above and placement of the stimulating electrodes in close proximity to the site of seizure onset. Similarly, Lian et al. (2003)
also reported that the antiepileptic effect of monopolar electrical stimulation in hippocampal slices was limited to a region surrounding the stimulation electrode. The limited spatial efficacy of sustained low-frequency stimulation in seizure prevention may present a significant problem for clinical use. Possible ways to overcome this problem are to increase the stimulus intensity or use multiple stimulating contacts. In our experiments, the duration of sustained low-frequency stimulation was limited to 45 min. Although under our experimental condition in vitro, the effect stabilized after 510 min of low-frequency stimulation, further experiments are needed in vivo to examine the effect of open-loop sustained low-frequency stimulation lasting days. It is important to stress that other open-loop stimulation paradigms can be used, for example, intermittent high-frequency trains as used in vagal nerve stimulation. This study was limited to continuous low-frequency stimulation paradigms. Additional studies are needed to address the antiepileptic efficacy of other potential open-loop stimulation paradigms.
This study was carried out in two acute models of epilepsy: BCC- and magnesium freetreated neocortical brain slices in vitro. There are several apparent differences between our models and human epilepsy. Seizure frequency in our models was much higher than human epilepsy or chronic animal models of epilepsy. Neocortical slices contain only a fraction of the cortical network and lack all subcortical and brain stem structures that can either amplify or attenuate seizures. Synaptic transmission was different in our model either because inhibition was compromised (BCC) or NMDA currents were enhanced (magnesium free). Moreover, the different physical environment in vitro and in vivo can affect the efficacy of stimulation. For all these reasons, further studies are needed to verify our results in chronic models of epilepsy in intact rats in vivo.
In addition to open-loop sustained low-frequency stimulation, a second more elegant approach to treat intractable epilepsy is with computer-controlled closed-loop devices. Such devices will automatically detect impending or ongoing seizures, and in response, generate the appropriate short stimulation trains to terminate emerging seizures. In contrast to open-loop stimulators, closed-loop devices will only generate short stimulation trains, while most of the time, the cortex will remain uninterrupted.
In our study, we were unable to detect the prodromal phase of seizures. Thus we were limited to examining the efficacy of various closed-loop stimulation protocols in prematurely terminating ongoing seizures. Our findings indicated that short high-frequency stimulation protocols had a clear antiepileptic effect. However, they only succeeded in terminating
50% of seizures during the 1- to 3-s, 100-Hz stimulation trains and shortening the remaining one half of the seizures by
50%. It is likely that if and when we have the capability to on-line identify impending seizures and apply stimulation during the prodromal phase of seizures, the antiepileptic efficacy of stimulation will increase.
In this study, we examined the underlying mechanisms responsible for the antiepileptic effect of electrical stimulation. We found that electrical stimulation marked depressed EPSPs and somewhat attenuated the excitability of neurons. Synaptic depression is one form of activity-dependent short-term synaptic plasticity (Von Gersdorff and Borst 2001
; Zucker and Regehr 2002
). Synaptic depression has previously been described in various peripheral and central synapses including excitatory synapses innervating layer 2/3 and layer 5 neocortical pyramidal neurons, the synapses examined in this study (Markram and Tsodyks 1996
; Thomson et al. 1993
; Varela et al. 1997
). The main mechanism responsible for synaptic depression is probably depletion of vesicles from the readily available pool of synaptic vesicles (Zucker and Regehr 2002
). However, other pre- and postsynaptic mechanisms probably also contribute to synaptic depression including desensitization of AMPA receptors, saturation of glutamate receptors, inactivation of presynaptic voltage-gated calcium channels, and activation of presynaptic metabotropic glutamate receptors (Schneggenburger et al. 2002
; Von Gersdorff and Borst 2001
; Zucker and Regehr 2002
).
In addition to short-term synaptic depression, sustained low-frequency stimulation also induced long-term synaptic depression. This finding is consistent with findings of previous studies that described long-term synaptic depression in the neocortex (Bear 1999
; Froc et al. 2000
; Perrett et al. 2001
). The magnitude long-term synaptic depression is much smaller than short-term synaptic depression. However, it may have a long-standing antiepileptic effect (Albensi et al. 2004
). It is interesting to note that, despite the induction of LTD in our preparation, we observed no significant long-term effects of sustained low-frequency stimulation on the magnitude of PDS discharges or on the frequency of spontaneous seizure-like events. This was possibly caused by the relative small magnitude of LTD.
In addition to synaptic depression, electrical stimulation also reduced the excitability of neurons during high-frequency stimulation. The reduction of neuronal excitability during high-frequency stimulation possibly resulted from inactivation of voltage-gated sodium channels (Goldin 2003
).
In conclusion, in this study, we showed that both low- and high-frequency cortical electrical stimulation can eliminate seizure-like events in BCC-treated neocortical brain slices. Moreover, we identified the relevant parameters influencing the antiepileptic effect of cortical electrical stimulation and showed that synaptic depression was the main mechanism responsible for the antiepileptic effect of electrical stimulation. In the future, studies are needed to further investigate the antiepileptic effect and safety profile of open- and closed-loop cortical stimulation in chronic animal models of epilepsy in vivo.
|
|
GRANTS |
|---|
|
|
|
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: Y. Schiller, Dept. of Neurology, Rambam Medical Center, 1 Efron St., Haifa 31096, Israel (E-mail: y_schiller{at}yahoo.com)
|
|
REFERENCES |
|---|
|
Avoli M, Drapeau C, Louvel J, Pumain R, Olivier A, Villemure JG. Epileptiform activity induced by low extracellular magnesium in the human cortex maintained in vitro. Ann Neurol 30: 589596, 1991.[CrossRef][Web of Science][Medline]
Barbarosie M, Avoli M. CA3-driven hippocampal-entorhinal loop controls rather than sustains in vitro limbic seizures. J Neurosci 17: 93089314, 1997.
Bear MF. A synaptic basis for memory storage in the cerebral cortex. Proc Natl Acad Sci USA 93: 1345313459, 1999.[CrossRef]
Bikson M, Lian J, Hahn PJ, Stacey WC, Sciortino C, Durand DM. Suppression of epileptiform activity by high frequency sinusoidal fields in rat hippocampal slices. J Physiol 531: 181191, 2001.
Brodie MJ, French JA. Management of epilepsy in adolescents and adults. Lancet 356: 323329, 2001.[CrossRef][Web of Science]
Browne TR, Holmes GL. Epilepsy. N Engl J Med 344: 11451151, 2001.
Cockerell OC, Johnson AL, Sander JW, Shorvon SD. Prognosis of epilepsy: a review and further analysis of the first nine years of the British National General Practice Study of Epilepsy, a prospective population-based study. Epilepsia 38: 3146, 1997.[CrossRef][Web of Science][Medline]
Cohen-Gadol AA, Stoffman MR, Spencer DD. Emerging surgical and radiotherapeutic techniques for treating epilepsy. Curr Opin Neurol 16: 213219, 2003.[CrossRef][Web of Science][Medline]
Cuellar-Herrera M, Velasco M, Velasco F, Velasco AL, Jimenez F, Orozco S, Briones M, Rocha L. Evaluation of GABA system and cell damage in parahippocampus of patients with temporal lobe epilepsy showing antiepileptic effects after subacute electrical stimulation. Epilepsia 45: 459466, 2004.[CrossRef][Web of Science][Medline]
Duncan JS, Sander JW, Sisodiya SM, Walker MC. Adult epilepsy. Lancet 367: 10871100, 2006.[CrossRef][Web of Science][Medline]
Durand DM, Bikson M. Suppression and control of epileptiform activity by electrical stimulation: a review. Proc IEEE 89: 10651082, 2001.[CrossRef]
Forsgrena L, Beghib E, Oun A, Sillanpaa M. The epidemiology of epilepsy in Europea systematic review. Eur J Neurol 12: 245253, 2005.[CrossRef][Web of Science][Medline]
Fountas KN, Smith JR, Murro AM, Politsky J, Park YD, Jenkins PD. Implantation of a closed-loop stimulation in the management of medically refractory focal epilepsy. Stereotact Funct Neurosurg 83: 153158, 2005.[CrossRef][Web of Science][Medline]
Froc DJ, Chapman CA, Trepel C, Racine RJ. Long-term depression and depotentiation in the sensorimotor cortex of the freely moving rat. J Neurosci 20: 438445, 2000.
Gluckman BJ, Neel EJ, Netoff TI, Ditto WL, Spano ML, Schiff SJ. Electric field suppression of epileptiform activity in hippocampal slices. J Neurophysiol 76: 42024205, 1996.
Goldin AL. Mechanisms of sodium channel inactivation. Curr Opin Neurobiol 13: 284290, 2003.[CrossRef][Web of Science][Medline]
Hablitz JJ. Spontaneous ictal-like discharges and sustained potential shifts in the developing rat neocortex. J Neurophysiol 58: 10521065, 1987.
Jerger K, Schiff SJ. Periodic pacing an in vitro epileptic focus. J Neurophysiol 73: 876879, 1995.
Khosravani H, Carlen PL, Velazquez JL. The control of seizure-like activity in the rat hippocampal slice. Biophys J 84: 687695, 2003.[Web of Science][Medline]
Kinoshita M, Ikeda A, Matsumoto R, Begum T, Usui K, Yamamoto J, Matsuhashi M, Takayama M, Mikuni N, Takahashi J, Miyamoto S, Shibasaki H. Electric stimulation on human cortex suppresses fast cortical activity and epileptic spikes. Epilepsia 45: 787791, 2004.[CrossRef][Web of Science][Medline]
Kossoff EH, Ritzl EK, Politsky JM, Murro AM, Smith JR, Duckrow RB, Spencer DD, Bergey GK. Effect of an external responsive neurostimulator on seizures and electrographic discharges during subdural electrode monitoring. Epilepsia 48: 15601567, 2004.
Krauss G, Gordon B. Brief bursts of pulse stimulation terminate after-discharges caused by cortical stimulation. Neurology 53: 20732081, 1999.
Kwan P, Sander JW. The natural history of epilepsy: an epidemiological view. J Neurol Neurosurg Psychiatry 75: 13761381, 2004.
Lesser RP, Kim SH, Beyderman L, Miglioretti DL, Webber WR, Bare M, Cysyk B, Krauss G, Gordon B. Brief bursts of pulse stimulation terminate after discharges caused by cortical stimulation. Neurology 53: 20732081, 1997.
Lian J, Bikson M, Sciortino C, Stacey WC, Durand DM. Local suppression of epileptiform activity by electrical stimulation in rat hippocampus in vitro. J Physiol 547: 427434, 2003.
Loscher W, Schmidt D. New horizons in the development of antiepileptic drugs: the search for new targets. Epilepsy Res 60: 77159, 2004.[CrossRef][Web of Science][Medline]
Markram H, Tsodyks M. Redistribution of synaptic efficacy between neocortical pyramidal neurons. Nature 382: 807810, 1996.[CrossRef][Medline]
McIntyre CC, Grill WM, Sherman DL, Thakor NV. Cellular effects of deep brain stimulation: model-based analysis of activation and inhibition. J Neurophysiol 91: 14571469, 2004a.
McIntyre CC, Savasta M, Kerkerian Le Goff L, Vitek JL. Uncovering the mechanisms of action of deep brain stimulation: activation, inhibition, or both. Clin Neurophysiol 115: 12391248, 2004b.[CrossRef][Web of Science][Medline]
Osorio I, Frei MG, Sunderam S, Giftakis J, Bhavaraju NC, Schaffner SF, Wilkinson SB. Automated seizure abatement in humans using electrical stimulation. Ann Neurol 58: 258268, 2005.[CrossRef][Web of Science][Medline]
Perrett SP, Dudek SM, Eagleman D, Montague PR, Friedlander MJ. LTD induction in adult visual cortex: role of stimulus timing and inhibition. J Neurosci 21: 23082319, 2001.
Polsky A, Mel BW, Schiller J. Computational subunits in thin dendrites of pyramidal cells. Nature Neurosci 7: 621627, 2004.[CrossRef][Web of Science][Medline]
Schiller J, Major G, Koester HJ, Schiller Y. NMDA spikes in basal dendrites of cortical pyramidal neurons. Nature 404: 285289, 2000.[CrossRef][Medline]
Schiller Y. Activation of a calcium-activated cation current during epileptiform discharges and its possible role in sustaining seizure-like events in neocortical slices. J Neurophysiol 92: 862872, 2002.
Schiller Y. Inter-ictal- and ictal-like epileptic discharges in the dendritic tree of neocortical pyramidal neurons. J Neurophysiol 88: 29542962, 2004.[CrossRef][Web of Science]
Schneggenburger R, Sakaba T, Neher E. Vesicle pools and short-term synaptic depression: lessons from a large synapse. Trends Neurosci 25: 206212, 2002.[CrossRef][Web of Science][Medline]
Schmidt D, Loscher W. Drug resistance in epilepsy: putative neurobiologic and clinical mechanisms. Epilepsia 46: 858877, 2005.[CrossRef][Web of Science][Medline]
Sperling MR. The consequences of uncontrolled epilepsy. CNS Spectr 9: 98109, 2004.[Web of Science][Medline]
Stocker M. Calcium-activated potassium channels: molecular determinants and function of the SK family. Nat Rev Neurosci 5: 758770, 2004.[CrossRef][Web of Science][Medline]
Teyler TJ. Comparative aspects of hippocampal and neocortical long-term potentiation. J Neurosci Methods 28: 101108, 1989.[CrossRef][Web of Science][Medline]
Theodore WH, Fisher RS. Brain stimulation for epilepsy. Lancet Neurol 3: 111118, 2004.[CrossRef][Web of Science][Medline]
Thomson AM, Deuchars J, West DC. Large deep layer pyramid-pyramid single axon EPSPs in slices of rat motor cortex display paired pulse and frequency dependent depression, mediated presynaptically and self facilitation, mediated postsynaptically. J Neurophysiol 70: 23542369, 1993.
Valenzuela V, Benardo LS. An in vitro model of persistent epileptiform activity in neocortex. Epilepsy Res 21: 195204, 1995.[CrossRef][Web of Science][Medline]
Varela JA, Sen K, Gibson J, Fost J, Abbott LF, Nelson SB. A quantitative description of short-term plasticity at excitatory synapses in layer 2/3 of rat primary visual cortex. J Neurosci 17: 79267940, 1997.
Velasco F, Velasco M, Velasco AL, Menez D, Rocha L. Electrical stimulation for epilepsy: stimulation of hippocampal foci. Stereotact Funct Neurosurg 77: 223227, 2001.[CrossRef][Web of Science][Medline]
Velasco M, Velasco F, Velasco AL, Boleaga B, Jimenez F, Brito F, Marquez I. Subacute electrical stimulation of the hippocampus blocks intractable temporal lobe seizures and paroxysmal EEG activities. Epilepsia 41: 158169, 2000.[CrossRef][Web of Science][Medline]
Vonck K, Boon P, Achten E, De Reuck J, Caemaert J. Long-term amygdalo-hippocampal map stimulation for refractory temporal lobe epilepsy. Ann Neurol 52: 556565, 2002.[CrossRef][Web of Science][Medline]
Von Gersdorff H, Borst JGG. Short-term plasticity at the calyx of held. Nat Rev Neurosci 3: 5364, 2001.[Web of Science]
Warren RJ, Durand DM. Effects of applied currents on spontaneous epileptiform activity induced by low calcium in the rat hippocampus. Brain Res 806: 186195, 1998.[CrossRef][Web of Science][Medline]
Zucker RS, Regehr WG. Short-term synaptic plasticity. Annu Rev Physiol 64: 355405, 2002.[CrossRef][Web of Science][Medline]
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |