INNOVATIVE METHODOLOGY
Department of Physiology and Biophysics, Georgetown University Medical Center, Washington, DC
Submitted 1 November 2006; accepted in final form 8 May 2007
| ABSTRACT |
|---|
|
|
|---|
500 individual detectors, each covering an area of cortical tissue 160 µm in diameter (total imaging field
4 mm in diameter) and a temporal resolution of 1,600 frames/s, without multiple-trial averaging. We can record 80100 trials of intermittent 10-s trials from each imaging field before the VSD signal reduces to one half of its initial amplitude because of bleaching and wash-out. Taken together, the methods described in this report provide a useful tool for visualizing evoked and spontaneous waves from rodent cortex. | INTRODUCTION |
|---|
|
|
|---|
105102 of the resting absorption/fluorescence intensity per 100 mV of membrane potential change. In biological experiments, such small signals are quite vulnerable to noise and artifacts. For imaging in vitro preparations such as brain slices and cultured cells, noise and artifact can be controlled well and the sensitivity of VSD imaging rivals that of local field potential recordings (Jin et al. 2002
New "blue" dyes, developed by Amiram Grinvald's group (Shoham et al. 1999
), have brought a great advance for in vivo imaging of mammalian cortex. The excitation wavelength of blue dyes has minimal overlap with the absorption of hemoglobin and hence has minimal pulsation artifact (Shoham et al. 1999
). Although research reports using blue dyes have been published in recent years (Ferezou et al. 2006
; Petersen et al. 2003a
,b
; Slovin et al. 2002
), many technical issues have not been examined. These include methods for reaching a high sensitivity, optimum staining, the reduction of residual pulsation artifact, and limitations in recording time caused by bleaching and photo-toxicity. In this study, we show methods to achieve high signal-to-noise ratio recordings using VSD, with sensitivity as good as that of local field potential recording. This sensitivity allows for recording of spontaneous and evoked activity in single trials without spatial or temporal averaging. We will also discuss methodological issues such as elimination of residual pulsation artifacts, the total possible duration of optical recording experiments, and how deep in the cortex the dye can reach, when stained transdurally.
| METHODS |
|---|
|
|
|---|
Sprague-Dawley and Long Evans rats (250400 g) were used in the experiments. Surgical procedures were approved by Georgetown University Animal Care and Use Committee following National Institutes of Health guidelines.
Animals were pretreated with atropine sulfate (40 µg/kg, ip)
30 min before anesthetic induction to reduce mucus secretion. The animals were induced with 23% isoflurane in air, and tracheostomy was performed. A 16-G over-the-needle catheter was inserted into the trachea, and animals were ventilated by a small animal respirator (Harvard Apparatus) with isoflurane in room air. Respiration rate (60100/min) and volume (24 ml) were adjusted such that inspiratory pressure was
5 cmH2O and the end-tidal carbon dioxide (ETCO2) was maintained at
2628 mmHg, with a calibrating multigas monitor (BCI 9100). Next, the animal was placed in a stereotaxic frame with a regulated heating pad. As soon as the animal was secure, isoflurane was reduced to 1.52% for surgery and during recordings unless specified otherwise. In some experiments, the anesthesia level was lowered, and xylazine was used to augment the level of anesthesia. The animals were closely monitored to ensure sufficient anesthetic plane, using heartbeat response to tail-pinch. Because the depth of isoflurane anesthesia is sensitive to hypothermia (Regan and Eger 1967
), care was taken to ensure normothermia throughout recording. Lactated Ringer solution was periodically infused (
2.5 ml/kg/h, sc) to compensate for urination and insensitive fluid loss, and bland ophthalmic ointment was applied periodically to the eyes to prevent corneal desiccation.
The cranial surface was cleaned thoroughly of soft tissue, and a thin layer of surgical adhesive (Vetbond, 3M) was applied to prevent dye leakage during staining. Next, a cranial window (
5 mm diam) was drilled on the left hemisphere over visual cortex (bregma, 4 to 9 mm; lateral, 16 mm) or barrel field (bregma, 0 to 5 mm; lateral, 27 mm). Bone was carefully separated from the dura, which was left intact. Leaving dura intact significantly reduces the movement artifact during optical recording (London et al. 1989
). Care was taken to avoid irritation of the cortex by heat or pressure during this procedure. The drill tip was immersed in artificial cerebrospinal fluid (ACSF) during drilling to diffuse heat from the drilling. Nontraumatic craniotomy was found to be important for better staining. Irritated brain can appear reddish (because of increased blood flow), or CSF pressure can increase the potential subdural space, which leads to poor staining. In some experiments, dexamethasone sulfate (1 mg/kg, ip) was given 24 h before the experiment to reduce inflammatory response of the dura.
Staining
VSD RH1691 (Optical Imaging) was dissolved at 2 mg/ml in ACSF solution, and staining was done through the intact dura mater.
To increase the dural permeability to the dye, we dried the dura before staining. The surface of the dura mater was washed with ACSF, after which the fluid was removed thoroughly by suction. The dura mater was allowed to dry completely for 34 min, with gentle air flow, until the tissue became very transparent and "glassy." This glassy appearance is essential for good staining. Occasionally, longer drying (
15 min) was needed. The drying procedure did not alter local field potential profiles of the cortex.
A temporary staining chamber was constructed around the craniotomy with a thin layer of silicone valve grease. We used
200 µl of dye solution (
2 mg/ml of RH1691) to stain an area 5 mm in diameter. During staining, the dye solution was continuously circulated by a custom-made perfusion pump (London et al. 1989
). The pump had a battery-powered gear motor that gently pressed the rubber nipple of a Pasteur pipette once every few seconds. The tip of this pipette was placed in the staining solution and performed a gentle, back-and-forth circulation of a small amount of dye (
100 µl). Such circulation was necessary because CSF slowly exudes from the dura and dilutes the dye concentration locally at the dye/dura interface. Staining lasted for 90 min, followed by washing with dye-free ACSF for >15 min.
Physical stabilization of the cortex
By leaving the dura mater intact, we greatly increased the physical stability of the brain, and a sealed chamber was not necessary. In some preparations, we applied high viscosity silicon oil (60,000 centistokes, Sigma DMPS-60M) on top of the dura to further dampen cortical pulsation. This silicon oil was replaced every 30 min to 1 h, because exuded CSF would accumulate underneath and hinder the dampening effect.
Firm fixation within the stereotaxic frame and maintenance of proper body posture, hydration, and ETCO2 can minimize the movement of the skull caused by heartbeat and respiration. In some experiments, we paralyzed the animal with pancuronium bromide (1 mg/kg, im) and paused ventilation during optical recording to completely eliminate the respiratory movement. When ventilation was paused during recording, longer intertrial intervals (12 min) were given.
Imaging apparatus
The cortex was imaged by a x5 macroscope with a field of view
4 mm in diameter (Fig. 1), following the designs of Kleinfeld (Kleinfeld and Delaney 1996
) and Cohen (Prechtl et al. 1997
). The macroscope was assembled from commercial video camera lens (Navitar, 25 mm F0.95) and provides high numerical aperture to effectively collect fluorescent emission from the cortex. The macroscope has a numerical aperture of 0.45, which gathers
100 times more light than an ordinary x4 microscope objective.
|
The fluorescence of the dye (
700 nm, Optical Imaging) from the stained cortex was collected through the macroscope, filtered through a 695-nm long-pass filter (RG-695, Edmund Scientific), and projected onto the fiber optic aperture of a 464-channel photodiode array (WuTech Instruments). Each channel (detector) of the array received light from a cortical area of
160 µm in diameter. The photocurrent from each channel was individually amplified through a two-stage amplifier (Wu and Cohen 1993
). The first stage was an I-V converter of 50 V/µA (50-Mohm transimpedance). The fluorescent resting light intensity (RLI) of the stained cortex was about 1 V (20 nA of photo current) at the output of the I-V converter. The output of the I-V converter was high-pass filtered at 0.1 or 1.4 Hz to remove the DC component of the light or resting florescent light (RLI). The signal was fed into the second amplifier and amplified 500 times. The second stage amplifier only amplifies the differential component of the light and is therefore not saturated by the large RLI. An optical signal of typical sensory-evoked cortical activity was
103 of the RLI or
500 mV (peak-to-peak) at the output of the second stage amplifier. This signal was low-pass filtered at 400 Hz and digitized at 1.6 kHz by an A/D converter card (MicroStar DAP-4000) on a PC. Because the RLI is subtracted before the second stage amplifier, the full 12-bit range of the A/D converter is available to digitize the optical signal. The effective dynamic range can reach
19 bits (see DISCUSSION). The apparatus was mounted on a high-performance isolation stage (Minus K Technology) to remove floor vibrations.
Stimuli
For visual stimulation, we used a flash from a white LED, positioned in front of the contralateral eye. The stimulus lasted for 10 ms and covered
40° of the visual field of the animal. The light path was shielded to avoid contamination of the stimulation light into the optical recordings. For whisker stimulation, we used a 200-ms ramp-and-hold deflection in the anterior direction. The stimulator consisted of a galvanometer with a small loop on the needle tip, holding the whisker at 10 mm from the base, and provided rapid deflections of 3 mm (
15°).
Local field potential recording
Electrodes were placed on the edge of the optical recording field to record the epidural local field potential. Figure 2 uses a tungsten microelectrode;
Fig. 4 uses a silver ball electrode. The signal was amplified 1,000 times using a custom-made amplifier with frequency range of 1100 Hz. Local field potential was digitized concurrently with optical channels, along with ECG and driving voltage of the stimuli. The ECG was used to remove pulsation artifacts off-line.
|
|
|
Although the blue dyes have greatly reduced the artifact caused by hemoglobin absorption (Shoham et al. 1999
), movement artifact from brain pulsation is still a major contributor to recording noise. Because pulsation artifact is time locked to the ECG, it can be largely subtracted from each individual trial to observe single trials of VSD activity (Arieli et al. 1995
). We used an off-line algorithm for this subtraction based on previous methods from our laboratory (Ma et al. 2004
). In this subtraction algorithm, the peak of QRS waves was obtained from the ECG recorded simultaneously with the imaging trial. An average QRS interval was obtained for each trial, and ECG-triggered averages for each optical detector were obtained spanning three average QRS intervals, 1.5 intervals before the peak and 1.5 intervals after. This "averaged pulsation artifact" was repeatedly concatenated in synchrony with each QRS peak to reconstruct an average trace for each detector. Because the QRS interval varies subtly from heartbeat to heartbeat, simply concatenating the average waveform would produce small jumps at the point of each concatenation. To eliminate such jumps, the concatenation was done with a linear-weighted transitionat QRS peak number n, the averaged waveform aligned to peak n was weighted 100%; midway between QRS peak number n and n + 1, the waveform aligned to peak n and n + 1 were weighted at 50% each; at QRS peak number n ± 1, the waveform aligned to peak n + 1 is weighted 100%. The resulting concatenated average pulsation artifact for each optical detector was subtracted from the raw data for that detector (Fig. 5). The subtracted signals were analyzed with a fast Fourier transform and show virtually no energy peak at the heartbeat frequency. Matlab code for the procedure is available on request.
|
Data were processed in Neuroplex (RedShirt Imaging), Matlab (Mathworks), and Mathematica (Wolfram Research). Scripts are available on request. Pseudocolor movies were generated such that each detector was individually scaled to a linear color scale (Grinvald et al. 1982
; Jin et al. 2002
).
| RESULTS |
|---|
|
|
|---|
With our methods to eliminate pulsation artifact, the sensitivity of VSD recording was comparable to that of local field potential recordings (Fig. 2). In this experiment, we recorded cortical activity by VSD imaging and local field potential simultaneously from the same tissue. Comparing the two signals, we found most of the peaks in the local field potential showed a corresponding event in the VSD trace. Under 1.5% isoflurane anesthesia, infrequent bursts of spontaneous activity occurred on both electrical and optical recordings, and most events within each burst correlated well between local field potential and VSD signals (Fig. 2A). However, local field potential and VSD signals were frequently disproportionate in amplitude. In Fig. 2A, events labeled with dots had higher amplitude in VSD trace compared with that in the local field potential trace, whereas events labeled with triangles had larger amplitude in the local field potential trace compared with that in the VSD trace. Events labeled with diamonds were seen in the local field potential but not in the VSD signals. Adjusting the level of anesthesia allows us to further verify the correlation between local field potential and the VSD signals. When the level of anesthesia was lowered, both VSD and local field potential signals show continuous spontaneous fluctuations (Fig. 2B). The fluctuations in the VSD signals are unlikely to be baseline noise, because the baseline noise is much smaller in the quiescent segments of Fig. 2A, which were obtained from the same location in the same animal. Instead, these fluctuations are likely to be biological signals from local cortical activity, probably sleep-like oscillations or up/down states (Petersen et al. 2003b
). Although the correlation between the local field potential and VSD recordings decreased, many events in local field potential were also seen in the VSD traces (exceptions are marked with squares in Fig. 2B). VSD recordings made from two adjacent locations also show good correlation (bottom 2 traces in each panel of Fig. 2), further indicating that the fluctuations on the VSD traces were not noise. Because almost every event in the local field potential can be seen in the VSD recordings, the sensitivity of the VSD measurement is comparable with that of local field potential recordings. However, VSD and local field potential signals have a low overall correlation of
0.2 between 3 and 30 Hz (which includes 90% of the power). This low overall correlation is probably because some local field potential peaks originated from deep cortical layers or subcortical sources.
The correlation between optical traces is much higher, and each peak in the signal appears at many locations (Fig. 2, A and B, bottom traces). This high correlation between optical traces is not from light scattering or optical blurring between channels, because there is a small timing difference between locations, resulting globally in a propagating wave. The spatiotemporal pattern of the waves will be further discussed in later sections.
Staining and the source of the VSD signal
Cortex was stained well after 90 min of staining with RH1691 through the dura mater (Fig. 3A). We sliced the cortex after staining and found that the fluorescence of the dye reached a depth of 500 µm from the cortical surface (Fig. 3A). The fluorescence intensity (Fig. 3A, right) shows that most of the dye fluorescence originates from layers I and II/III; these layers contain
70% of the fluorescence signal. The highest intensity is in layer I, which contributes
30% to the overall fluorescence. This dye distribution in cortical tissue is similar to that stained with dura removed in rat (Kleinfeld and Delaney 1996
; Petersen et al. 2003a
) and mice (Ferezou et al. 2006
). The distribution of the fluorescence intensity indicates that the VSD signals mostly originate from the layers IIII. In addition, because illumination intensity will decrease with depth while scattering/reabsorption of the dye fluorescence will increase with depth, activity in layer IV and deeper layers have little contribution to the signal. Therefore the VSD signal described here is most likely a combination of synaptic and spiking activity in layer I to II/III. Because the fluorescence of RH1691 is in the near infrared range, visual evaluation from the dural/cortical surface may not reveal the quality of staining. When cortex was visually inspected under room light, only a light blue hue was seen in the tissue (Fig. 3B). However, even with this light visual appearance (as shown in Fig. 3), the stained cortex yields excellent optical signals during imaging (Figs. 4, 5, 7, and 8).
|
|
The pulsation artifact of RH1691 was very small compared with that of RH795 (Fig. 4; Shoham et al. 1999
). In this comparison, we used spontaneous epileptiform spikes as the source of VSD signal. Applying bicuculline methiodide to the cortex induces spontaneous epileptiform spikes at a frequency of 1 spike/s (Gutnick et al. 1982
). The spikes provide a large VSD signal (
102 of resting fluorescent intensity), and the signal is relatively stable from trial to trial and from animal to animal (Ma et al. 2004
). We stained with each dye in independent experiments and found both RH1691 and RH795 had similar signal amplitude of epileptiform spikes (Fig. 4). However, the pulsation artifact of RH1691 was much smaller. The amplitude of the pulsation artifact in RH795 was similar to that of the epileptiform spikes, whereas the artifact of RH1691 was only about one seventh of the signal amplitude. This comparison shows a great potential for in vivo imaging using the new blue dyes, as shown in recent reports (Derdikman et al. 2003
; Ferezou et al. 2006
; Grinvald and Hildesheim 2004
; Petersen et al. 2003a
,b
; Shoham et al. 1999
; Spors and Grinvald 2002
).
Removal of residual pulsation artifact
In our experiments, we frequently observe large VSD signals with negligible pulsation artifact (Fig. 5, optical 1). However, residual pulsation artifact can still be large compared with small signals such as spontaneous and sensory-evoked cortical activity. Figure 5 compares residual pulsation artifact at various locations within the visual cortex. At some locations, the pulsation artifact was almost absent (Fig. 5, optical 1). Other locations had pulsation artifact of varying waveform and polarities (Fig. 5, opticals 2 and 3). These traces suggest that the artifact is mainly contributed by movement because of pulsation, because movement artifact can have either polarity at various locations while the polarity of hemoglobin absorption should be consistent at all cortical locations.
To remove residual pulsation artifact, an off-line subtraction procedure is necessary. The procedure can effectively remove artifact that is time locked with the ECG, regardless of polarity and waveform (Fig. 5). There were
50 heartbeats in each 10-s recording trial; therefore in the averaged artifact trace (see METHODS), signal and noise components that are not correlated to the heartbeat were reduced to
1/50 of their original amplitude, and
98% of the artifact can be removed (Fig. 5, bottom traces).
Total recording time
In brain slices, the recording time of VSD imaging is mainly limited only by dye bleaching (Jin et al. 2002
; Momose-Sato et al. 1999
). Using absorption dyes in brain slices, the total optical recording time is
1,800 s before the VSD signal declines to one half of the initial amplitude (Jin et al. 2002
). Under in vivo conditions, other factors such as washing out of the dye by blood circulation may also reduce the total recording time. Therefore the total recording time must be determined empirically.
To determine the feasible total recording time, we measured the decrease of VSD signal (RH1691) under an illumination intensity of
1.5 x 1011 photo-electrons/mm2 ms (at this intensity, the sensitivity of VSD imaging is comparable with that of local field potential recordings). We used spontaneous epileptiform spikes induced by bicuculline methiodide as the source of VSD signal. Epileptiform spikes provide a large VSD signal (
102), and the signal is relatively stable from trial to trial and from animal to animal (Ma et al. 2004
). The cortex was illuminated intermittently, with 20 s of light exposure each minute. The amplitude of the VSD signal was plotted along with exposure time in Fig. 6.
|
To further distinguish dye wash-out from light-related amplitude reduction, we shielded part of the cortex from light during a light exposure experiment (Fig. 6B). After the signal in the exposed area transitioned to the declining period, the shield was removed. Signal from the previously shielded area had similar amplitude as that in the flat period. (Fig. 6B, solid dots). However, the signal in the shielded area did not show a flat period; instead, the signal declined at a similar rate as the unshielded regions. This suggests that the flat period was related to the dye wash-out, which only occurred within a certain period after the staining.
Our data indicate that one can collect 8001,000 s of data before the VSD signal reduced to one half of the initial amplitude. This estimated time was shorter than previous reports (Shoham et al. 1999
), probably because we use higher illumination intensity for higher sensitivity (Fig. 7).
Sensory-evoked activity
We were able to reliably detect sensory-evoked cortical activity in single trials. Signals in Fig. 7 were recorded from individual detectors (160 µm in diameter) in single trials, showing the ability to achieve a high spatiotemporal resolution. The evoked activity had lower amplitude when evoked at certain phases of spontaneous activity (Fig. 7, A, traces 8 and 10, and B, traces 3 and 6). Variations in evoked cortical activity also occurred in spatial pattern.
Propagating waves
Both spontaneous and evoked cortical activity manifested as propagating waves (Fig. 8). Sensory evoked waves followed a consistent propagation pattern, initiating from the location of the thalamic afferents and spreading to large areas of the cortex. In barrel cortex, the whisker evoked waves initiated from within the corresponding barrel and propagated to the entire barrel cortex (Fig. 8A, top row), with an overall velocity of 0.2 ± 0.1 m/s (n = 6 animals). In visual cortex, flash evoked waves started from V1 and propagated to other visual areas (Fig. 8B, top row). Visually evoked waves had a similar overall velocity as those in the barrel cortex. Spontaneous events originated from various locations and propagated with larger variations in direction and velocity (Fig. 8, A and B, bottom rows). The spontaneous event in Fig. 8B appears as two waves: one propagating from the bottom left to upper right and followed immediately by another wave in the opposite direction. The spontaneous and evoked waves we have observed are consistent with reports from other groups (Derdikman et al. 2003
; Petersen et al. 2003b
).
Trial-to-trial variations
Obtaining data with high signal-to-noise ratio, from hundreds of trials from each animal, allows us to examine trial-to-trial variations of cortical activity evoked by identical sensory stimuli. Large trial-to-trial variations were observed in both visual and barrel cortex (Fig. 7). In the barrel cortex, theevoked wave started from the correspondent whisker barrel in all trials and propagated to the entire barrel field. However, individual trials showed large trial-to-trial variations in signal amplitude (Fig. 7) and propagation pattern (Fig. 9).
|
Overall, the blue dye RH1691 exhibited an excellent signal and low pulsation artifact, allowing the use of a high dynamic range imaging device to achieve sensitivity comparable with that of local field potential recordings. Such ability provides a useful tool to examine spatiotemporal dynamics in the cortex when multitrial averaging is not possible.
| DISCUSSION |
|---|
|
|
|---|
Saturation intensity and dynamic range of the imaging device
Visualizing spatiotemporal dynamics in single trials requires higher sensitivity than mapping active areas in the cortex. To visualize VSD signals of 104103 of the resting fluorescent intensity (e.g., Figs. 2B, 7, and 8), the imaging device needs to have high saturation intensity, i.e., a well size of 100 million photons. In addition, an effective dynamic range of 1719 bits is needed to digitize a 104 signal. Ordinary CCD/CMOS cameras with a well size of 110 million photons and 1014 bits of dynamic range will not be able to record the small signals shown in Figs. 2B, 7, and 8. Diode arrays offer an exceptional effective dynamic range by a parallel amplification design, i.e., each detector (pixel) has a dedicated two-stage amplifier (Cohen and Lesher 1986
; for circuit diagram and additional information see Wu and Cohen 1993
and Jin et al. 2002
.). Using the measurement of Fig. 2B as an example, the resting light intensity was
1 V at the output of the first stage. The signal, at
103 of the resting light intensity, therefore produced a change of
1 mV riding atop this resting 1 V. With a dedicated second stage amplifier for each pixel, the DC component (resting light intensity) can be easily removed by an analog circuit. After removing the DC component, the 1-mV signal is amplified 500 times to 500 mV. This signal is digitized with an A/D converter of 12 bits at a range of ±10 V, resulting in the 500-mV signal being represented by
8 bits. Such high dynamic range is essential for the overall sensitivity of optical recordings (Figs. 2B, 7, and 8). The dark noise of the photodiode array is very small compared with shot noise (<1/10), and therefore instrumentation noise can be ignored. In the measurements of this report, we have achieved a signal-to-noise ratio of
5 from each single detector (0.016 mm2, with total 464 detectors in an area of 4 mm in diameter) at 1,600 frames/s (Fig. 7).
Spatiotemporal resolution
Spatial and temporal resolutions are closely linked and limited by shot noise. Higher spatiotemporal resolution leads to lower photon flux because of the smaller volume of tissue under each detector and because of the shorter sampling time. Lower photon flux has higher proportional shot noise and therefore lower signal-to-noise ratio. Increasing illumination intensity may further increase spatiotemporal resolution; however, phototoxicity might become a concern. Although RH1691 does not show toxicity under moderate light (Petersen et al. 2003a
; Slovin et al. 2002
), our illumination intensity was higher than previous reports, and we did observe rapid signal decline (Fig. 6). To address the concern of phototoxicity, results from later recording trials should be compared with those from the beginning of the experiment for verification.
Differences between local field potential signal and VSD signal
Figure 2 shows that VSD and local field potential recordings show good event-to-event correlation but a poor overall correlation. We speculate that the poor correlation in the waveforms was caused by the difference in origin of the two signals. The VSD signals originate from superficial cortical layers as indicated by the staining profile (Fig. 3). Local field potential signals, on the other hand, derive from local current flow, and therefore may contain volume-conducted signal from deeper cortical layers or strong subcortical sources. Although quantitative analyses of the correlation between the VSD and local field potential signals are beyond the scope of this paper, we believe that the issue may be important for understanding how subcortical input manifests in the cortex. Our data suggest a useful method for separating superficial and deep cortical components in local field potential recordings.
Visualizing propagating waves
Many cortical activities are spatiotemporally organized as propagating waves ("traveling waves"; Ermentrout and Kleinfeld 2001
). Propagating waves have been extensively examined in cortical slices (Albowitz and Kuhnt 1995
; Bao and Wu 2003
; Chagnac-Amitai and Connors 1989
; Chervin et al. 1988
; Demir et al. 1998
; Fleidervish et al. 1998
; Golomb and Amitai 1997
; Huang et al. 2004
; Miyakawa et al. 2003
; Tanifuji et al. 1994
; Tsau et al. 1998
; Wadman and Gutnick 1993
; Wu et al. 1999
, 2001
). Because the velocity and direction of propagation are highly dynamic, the ability to record from single trials is essential for studying wave propagation. Before introduction of the blue dyes, sensory-evoked waves could only be imaged in the brain of turtle (Lam et al. 2000
, 2003
; Prechtl et al. 1997
; Senseman and Robbins 1999
, 2002
), salamander (Cinelli et al. 1995
), and invertebrate (Delaney et al. 1994
; Kleinfeld et al. 1994
), where heartbeat is not present or can be paused with cooling. The pulsation artifact of red dye RH795 had an amplitude of 0.51% of resting light intensity, similar to that of epileptiform spikes induced by bicuculline (Fig. 4A), and
10 times larger than the sensory-evoked cortical activity (London et al. 1989
). With the use of blue dyes, pulsation artifact is greatly reduced and sensory evoked waves can be seen in mammalian cortex (Derdikman et al. 2003
; Ferezou et al. 2006
; Petersen et al. 2003a
,b
).
Pulsation artifact
RH1691 shows virtually no hemoglobin absorption artifact (Figs. 4 and 5; Ferezou et al. 2006
; Shoham et al. 1999
), but pulsating movement artifact may still be large. Imaging with the dura mater intact increased stability, and ECG-triggered subtraction further removed residual movement artifact time-locked to the heartbeat (Fig. 5). In a 10-s recording trial, there are
50 heartbeats; therefore signals that are not time-locked to the ECG would be distorted by <2%. The results are similar to other advanced mathematical methods, such as independent component analysis and advanced frequency domain analysis (Mitra and Pesaran 1999
; Pesaran et al. 2005
; Reidl et al. 2007
).
Limitations of the method
The main limitation of VSD imaging methods is that the signals are from the superficial layers of the cortex (Fig. 3). Further improvements in this regard will await newer staining methods or bioengineered VSDs. Our transdural staining and imaging is also limited in species, because carnivores and primates have thicker and less translucent dura. The Grinvald group and others have described artificial dural implants (Chen et al. 2002
; Slovin et al. 2002
), which may provide sufficient mechanical stability in those species to obtain a large number of stable single trial recordings. Another limitation is that VSD recording is not a preferred way to measure DC signal. This is mainly because the DC component in VSD signal is contaminated by dye bleaching. In the process of making pseudocolor maps in this report, the signals were digitally filtered above 23 Hz to emphasize the propagating wave front. However, our cut-off at low frequency can be as low as 0.06 Hz, which is low enough to capture all low frequency components reflected in conventional EEG signals. The third limitation to this method is the use of anesthetized and firmly fixed animals. Elimination of motion is necessary for high-sensitivity recordings. This could perhaps be alleviated by novel and humane awake fixation methods or by optical-fiber based tether method (Ferezou et al. 2006
).
In conclusion, the methods described in this report offer high signal-to-noise ratio recordings for a number of applications, such as the study of trial-to-trial variability of sensory evoked waves, interaction between evoked and spontaneous activity, event-related activity in higher cortical areas, the interactions between individual neurons and local population, and the complex dynamics of epileptic activity.
| GRANTS |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
|
|
|---|
Present address of M. T. Lippert: Leibniz-Institut for Neurobiology; 39118 Magdeburg; Germany.
| FOOTNOTES |
|---|
The costs of publication of this article were defrayed in part by the payment of page charges. The article must therefore be hereby marked "advertisement" in accordance with 18 U.S.C. Section 1734 solely to indicate this fact.
Address for reprint requests and other correspondence: J.-Y. Wu, The Research Building, WP26, 3970 Reservoir Rd., NW, Washington, DC 20057 (E-mail: wuj{at}georgetown.edu)
| REFERENCES |
|---|
|
|
|---|
Arieli A, Shoham D, Hildesheim R, Grinvald A. Coherent spatiotemporal patterns of ongoing activity revealed by real-time optical imaging coupled with single-unit recording in the cat visual cortex. J Neurophysiol 73: 20722093, 1995.
Bao W, Wu JY. Propagating wave and irregular dynamics: spatiotemporal patterns of cholinergic theta oscillations in neocortex in vitro. J Neurophysiol 90: 333341, 2003.
Chagnac-Amitai Y, Connors BW. Horizontal spread of synchronized activity in neocortex and its control by GABA-mediated inhibition. J Neurophysiol 61: 747758, 1989.
Chen LM, Heider B, Williams GV, Healy FL, Ramsden BM, Roe AW. A chamber and artificial dura method for long-term optical imaging in the monkey. J Neurosci Methods 113: 4149, 2002.[CrossRef][Web of Science][Medline]
Chervin RD, Pierce PA, Connors BW. Periodicity and directionality in the propagation of epileptiform discharges across neocortex. J Neurophysiol 60: 16951713, 1988.
Cinelli AR, Neff SR, Kauer JS. Salamander olfactory bulb neuronal activity observed by video rate, voltage-sensitive dye imaging. I. Characterization of the recording system. J Neurophysiol 73: 20172032, 1995.
Cohen LB, Keynes RD, Hille B. Light scattering and birefringence changes during nerve activity. Nature 218: 438441, 1968.[CrossRef][Medline]
Cohen LB, Lesher S. Optical monitoring of membrane potential: methods of multisite optical measurement. Soc Gen Physiol Ser 40: 7199, 1986.[Medline]
Delaney KR, Gelperin A, Fee MS, Flores JA, Gervais R, Tank DW, Kleinfeld D. Waves and stimulus-modulated dynamics in an oscillating olfactory network. Proc Natl Acad Sci USA 91: 669673, 1994.
Demir R, Haberly LB, Jackson MB. Voltage imaging of epileptiform activity in slices from rat piriform cortex: onset and propagation. J Neurophysiol 80: 27272742, 1998.
Derdikman D, Hildesheim R, Ahissar E, Arieli A, Grinvald A. Imaging spatiotemporal dynamics of surround inhibition in the barrels somatosensory cortex. J Neurosci 23: 31003105, 2003.
Ermentrout GB, Kleinfeld D. Traveling electrical waves in cortex: insights from phase dynamics and speculation on a computational role. Neuron 29: 3344, 2001.[CrossRef][Web of Science][Medline]
Ferezou I, Bolea S, Petersen CC. Visualizing the cortical representation of whisker touch: voltage-sensitive dye imaging in freely moving mice. Neuron 50: 617629, 2006.[CrossRef][Web of Science][Medline]
Fleidervish IA, Binshtok AM, Gutnick MJ. Functionally distinct NMDA receptors mediate horizontal connectivity within layer 4 of mouse barrel cortex. Neuron 21: 10551065, 1998.[CrossRef][Web of Science][Medline]
Golomb D, Amitai Y. Propagating neuronal discharges in neocortical slices: computational and experimental study. J Neurophysiol 78: 11991211, 1997.
Grinvald A, Hildesheim R. VSDI: a new era in functional imaging of cortical dynamics. Nat Rev Neurosci 5: 874885, 2004.[CrossRef][Web of Science][Medline]
Grinvald A, Manker A, Segal M. Visualization of the spread of electrical activity in rat hippocampal slices by voltage-sensitive optical probes. J Physiol 333: 269291, 1982.
Gutnick MJ, Connors BW, Prince DA. Mechanisms of neocortical epileptogenesis in vitro. J Neurophysiol 48: 13211335, 1982.
Huang X, Troy WC, Yang Q, Ma H, Laing CR, Schiff SJ, Wu JY. Spiral waves in disinhibited mammalian neocortex. J Neurosci 24: 98979902, 2004.
Jin W, Zhang RJ, Wu JY. Voltage-sensitive dye imaging of population neuronal activity in cortical tissue. J Neurosci Methods 115: 1327, 2002.[CrossRef][Web of Science][Medline]
Kleinfeld D, Delaney KR. Distributed representation of vibrissa movement in the upper layers of somatosensory cortex revealed with voltage-sensitive dyes. J Comp Neurol 375: 89108, 1996.[CrossRef][Web of Science][Medline]
Kleinfeld D, Delaney KR, Fee MS, Flores JA, Tank DW, Gelperin A. Dynamics of propagating waves in the olfactory network of a terrestrial mollusk: an electrical and optical study. J Neurophysiol 72: 14021419, 1994.
Lam YW, Cohen LB, Wachowiak M, Zochowski MR. Odors elicit three different oscillations in the turtle olfactory bulb. J Neurosci 20: 749762, 2000.
Lam YW, Cohen LB, Zochowski MR. Odorant specificity of three oscillations and the DC signal in the turtle olfactory bulb. Eur J Neurosci 17: 436446, 2003.[CrossRef][Web of Science][Medline]
London JA, Cohen LB, Wu JY. Optical recordings of the cortical response to whisker stimulation before and after the addition of an epileptogenic agent. J Neurosci 9: 21822190, 1989.[Abstract]
Ma HT, Wu CH, Wu JY. Initiation of spontaneous epileptiform events in the rat neocortex in vivo. J Neurophysiol 91: 934945, 2004.
Mitra PP, Pesaran B. Analysis of dynamic brain imaging data. Biophys J 76: 691708, 1999.[Web of Science][Medline]
Miyakawa N, Yazawa I, Sasaki S, Momose-Sato Y, Sato K. Optical analysis of acute spontaneous epileptiform discharges in the in vivo rat cerebral cortex. Neuroimage 18: 622632, 2003.[CrossRef][Web of Science][Medline]
Momose-Sato Y, Sato K, Arai Y, Yazawa I, Mochida H, Kamino K. Evaluation of voltage-sensitive dyes for long-term recording of neural activity in the hippocampus. J Membr Biol 172: 145157, 1999.[CrossRef][Web of Science][Medline]
Pesaran B, Sornborger AT, Nishimura N, Kleinfeld D, Mitra PP. Analysis of dynamic optical imaging data. In: Imaging in Neuroscience and Development: A Laboratory Manual, edited by Yuste R, Konnerth A. Cold Spring Harbor, New York: Cold Spring Harbor Laboratory Press, 2005, p. 815826.
Petersen CC, Grinvald A, Sakmann B. Spatiotemporal dynamics of sensory responses in layer 2/3 of rat barrel cortex measured in vivo by voltage-sensitive dye imaging combined with whole-cell voltage recordings and neuron reconstructions. J Neurosci 23: 12981309, 2003a.
Petersen CC, Hahn TT, Mehta M, Grinvald A, Sakmann B. Interaction of sensory responses with spontaneous depolarization in layer 2/3 barrel cortex. Proc Natl Acad Sci USA 100: 1363813643, 2003b.
Prechtl JC, Cohen LB, Pesaran B, Mitra PP, Kleinfeld D. Visual stimuli induce waves of electrical activity in turtle cortex. Proc Natl Acad Sci USA 94: 76217626, 1997.
Regan MJ, Eger EI. Effect of hypothermia in dogs on anesthetizing and apneic doses of inhalation agents. Determination of the anesthetic index (Apnea/MAC). Anesthesiology 28: 689700, 1967.[Web of Science][Medline]
Reidl J, Starke J, Omer DB, Grinvald A, Spors H. Independent component analysis of high-resolution imaging data identifies distinct functional domains. Neuroimage 34: 94108, 2007.[CrossRef][Web of Science][Medline]
Ross WN, Salzberg BM, Cohen LB, Grinvald A, Davila HV, Waggoner AS, Wang CH. Changes in absorption, fluorescence, dichroism, and Birefringence in stained giant axons: optical measurement of membrane potential. J Membr Biol 33: 141183, 1977.[CrossRef][Web of Science][Medline]
Senseman DM, Robbins KA. Modal behavior of cortical neural networks during visual processing. J Neurosci 19: RC3, 1999.
Senseman DM, Robbins KA. High-speed VSD imaging of visually evoked cortical waves: decomposition into intra- and intercortical wave motions. J Neurophysiol 87: 14991514, 2002.
Shoham D, Glaser DE, Arieli A, Kenet T, Wijnbergen C, Toledo Y, Hildesheim R, Grinvald A. Imaging cortical dynamics at high spatial and temporal resolution with novel blue voltage-sensitive dyes. Neuron 24: 791802, 1999.[CrossRef][Web of Science][Medline]
Slovin H, Arieli A, Hildesheim R, Grinvald A. Long-term voltage-sensitive dye imaging reveals cortical dynamics in behaving monkeys. J Neurophysiol 88: 34213438, 2002.
Spors H, Grinvald A. Spatio-temporal dynamics of odor representations in the mammalian olfactory bulb. Neuron 34: 301315, 2002.[CrossRef][Web of Science][Medline]
Tanifuji M, Sugiyama T, Murase K. Horizontal propagation of excitation in rat visual cortical slices revealed by optical imaging. Science 266: 10571059, 1994.
Tasaki I, Watanabe A, Sandlin R, Carnay L. Changes in fluorescence, turbidity, and birefringence associated with nerve excitation. Proc Natl Acad Sci USA 61: 883888, 1968.
Tsau Y, Guan L, Wu JY. Initiation of spontaneous epileptiform activity in the neocortical slice. J Neurophysiol 80: 978982, 1998.
Wadman WJ, Gutnick MJ. Non-uniform propagation of epileptiform discharge in brain slices of rat neocortex. Neuroscience 52: 255262, 1993.[CrossRef][Web of Science][Medline]
Wu JY, Cohen LB. Fast multisite optical measurement of membrane potential. In: Fluorescent and Luminescent Probes of Biological Activity, edited by Mason WT. San Diego, CA: Academic Press, 1993, p. 389404.
Wu JY, Guan L, Bai L, Yang Q. Spatiotemporal properties of an evoked population activity in rat sensory cortical slices. J Neurophysiol 86: 24612474, 2001.
Wu JY, Guan L, Tsau Y. Propagating activation during oscillations and evoked responses in neocortical slices. J Neurosci 19: 50055015, 1999.
This article has been cited by other articles:
![]() |
B. Ahmed, A. Hanazawa, C. Undeman, D. Eriksson, S. Valentiniene, and P. E. Roland Cortical Dynamics Subserving Visual Apparent Motion Cereb Cortex, December 1, 2008; 18(12): 2796 - 2810. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-Y. Wu, Xiaoying Huang, and Chuan Zhang Propagating Waves of Activity in the Neocortex: What They Are, What They Do Neuroscientist, October 1, 2008; 14(5): 487 - 502. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |