INNOVATIVE METHODOLOGY
1Department of Physics and Astronomy, University of Pennsylvania School of Arts and Sciences, Philadelphia; 2Department of Bioengineering, University of Pennsylvania School of Engineering and Applied Science, Philadelphia; and 3Departments of Neuroscience and 4Physiology, University of Pennsylvania School of Medicine, Philadelphia, Pennsylvania
Submitted 17 August 2007; accepted in final form 2 January 2008
|
|
ABSTRACT |
|---|
|
1–2 µm) mammalian nerve terminals in an intact in vitro preparation, the mouse neurohypophysis. The measurements used two-photon excitation along the "blue" edge of the two-photon absorption spectrum of di-3-ANEPPDHQ (a fluorescent voltage-sensitive naphthyl styryl-pyridinium dye), and epifluorescence detection, a configuration that is critical for noninvasive recording of electrical activity from intact brains. Single-trial recordings of action potentials exhibited signal-to-noise ratios of
5:1 and fractional fluorescence changes of up to
10%. This method, by virtue of its optical sectioning capability, deep tissue penetration, and efficient epifluorescence detection, offers clear advantages over linear, as well as other nonlinear optical techniques used to monitor voltage changes in localized neuronal regions, and provides an alternative to invasive electrode arrays for studying neuronal systems in vivo. |
|
INTRODUCTION |
|---|
|
Imaging techniques employing potentiometric dyes as molecular indicators of membrane voltage (Cohen and Salzberg 1978
; Salzberg 1983
) permit noninvasive measurements of electrical activity from a wide range of excitable structures ranging from individual dendritic spines (Nuriya et al. 2006
) to intact, complex neuronal circuits, in vitro as well as in vivo, with extremely high (
0.002 ms at room temperature) temporal resolution (Salzberg et al. 1993
). Previous voltage-sensitive dye imaging of neuronal networks in vitro (e.g., Ang et al. 2005
; Contreras and Llinás 2001
; Obaid et al. 1999
, 2005
) and in vivo (e.g., Civillico and Contreras 2006
; Fisher et al. 2004
; Petersen et al. 2003
) have used conventional one-photon (linear) excitation. Although one-photon fluorescence microscopy can be used to measure neuronal activity with high signal-to-noise (S:N), these measurements inherently integrate the activity encoded by out-of-focus fluorescence signals. This spatial summation of signals complicates the interpretation of fluorescence patterns from thick tissue samples and precludes the detection of potentially localized events such as activity in individual nerve terminals and dendrites.
Nonlinear optical microscopy techniques such as two-photon microscopy (Denk et al. 1990
) and second harmonic generation (SHG) microscopy (Campagnola et al. 1999
; Hellwarth and Christensen 1974
; Nuriya et al. 2006
; Sheppard et al. 1977
) offer intrinsic optical sectioning at micrometer-scale resolution, virtually eliminating spatial signal summation. However, single-trial measurements of fast electrical events in three-dimensional preparations (as opposed to those in low-density cultured neurons; Kuhn et al. 2004
; Sacconi et al. 2006
) with appreciable S:N have not yet been achieved using nonlinear optical techniques (Sjulson and Miesenboeck 2007
). The small size of one-photon fluorescence changes in response to membrane depolarization, the limited two-photon absorption cross-sections (and, for SHG, 2-photon scattering cross-section) of potentiometric probes, and the constraints on incident light intensity imposed by the need to avoid phototoxicity, all conspire against this goal. In the case of SHG, the nonlinear optical signal propagates primarily in the forward direction, rendering epiillumination measurements (180°) severely shot-noise limited. Furthermore, with the exception of a few studies (Fisher et al. 2005
; Millard et al. 2005
; Moreaux et al. 2003
), the wavelength-dependence of potentiometric, nonlinear optical signals has not been explored.
The neurohypophysis (or posterior pituitary) is a classic in vitro preparation for studying evoked release of peptide hormones (Douglas 1973
). Magnocellular neurons, with their cell bodies in the hypothalamus, project their axons as a bundle of fibers through the median eminence and infundibular stalk, to terminate in the neurohypophysis where they ramify extensively. In the rat,
20,000 neurons give rise to
40,000,000 terminals and secretory swellings, each 1–10 µm in diameter, which account for
99% of the excitable membrane in the neurohypophysis, whereas the short axons contribute only
1% (Nordmann 1977
). When this in vitro preparation is stained with voltage-sensitive dyes and is stimulated electrically, it generates large optical signals. These compound responses, i.e., changes in one-photon absorption (Gainer et al. 1986
) or fluorescence (Muschol et al. 2003
), represent the synchronous firing of APs from a large population of neurosecretory terminals.
Here we present single-trial two-photon imaging of APs from either individual, or a very few (1–2 µm), nerve terminals in a surgically excised, intact mouse neurohypophysis stained with the naphthyl styryl-pyridinium potentiometric probe di-3-ANEPPDHQ (Obaid et al. 2004
). To optimize the recording conditions, we studied the wavelength dependence of the signal with excitation between 800 and 900 nm and emission between 490 and 700 nm. The best two-photon single-trial measurements exhibited high S:N (up to 5:1) and large fractional fluorescence changes (up to
10%). To our knowledge, these are the first single-trial optical recordings of electrical activity at submicrometer resolution from an intact mammalian preparation using epifluorescence detection, and they provide the methodological foundation for in-depth optical recording of electrical activity from neuronal circuits in vivo.
|
|
METHODS |
|---|
|
To record electrical activity from small nerve terminals in the intact neurohypophysis, we used a home-built two-photon microscope (Fisher 2007
). The excitation source was a mode-locked Titanium:Sapphire (Ti:Al2O3) laser (Chameleon, Coherent, Santa Clara, CA). Two-photon absorption-induced fluorescence was collected in the epifluorescence path by the objective and redirected to a photomultiplier (PMT) using a dichroic mirror [90% reflectance (R) at 700 nm] followed by an emission filter (typically interference-coated glass absorption filter). To minimize distortion of the beam profile, we adjusted the average excitation power at the sample to be
10–20 mW using a
/2 plate and a Glan–Thompson polarizer. The PMT used to detect two-photon excited fluorescence was an R943 (Hamamatsu Photonics, Hamamatsu City, Japan) operating in analog mode. (Signals were recorded as analog because the photon flux was too high for reliable photon counting.) To monitor and correct for fluctuations in the square of the laser power (
P(t)
2), a small fraction of the beam was picked off with a microscope coverslip and focused onto a β-barium borate (β-BaB2O4, or BBO) SHG crystal. This SHG signal was filtered with a BG39 Schott glass absorption filter and detected by an auxiliary PMT (R955, Hamamatsu Photonics), also operating in analog mode. The excitation beam was scanned in the sample plane using galvanometer scanning mirrors from Cambridge Technology (Lexington, MA) and a piezoelectric-driven objective translator (P721, Physik Instrumente, Karlsruhe/Palmbach, Germany) scanned the beam along the vertical (illumination) axis. The combination of scan lens and tube lens allowed us to expand the excitation laser beam and back-fill a microscope objective (x20/0.95 NA XLUMPlanFl, Olympus, Melville, NY). The dichroic mirror that selected the two-photon excited fluorescence emission (R < 700 nm), and the various emission interference filters were from Chroma Technology (Rockingham, VT). Analog signals from both PMTs were amplified with a wide bandwidth transimpedance amplifier (13AMP005, Melles Griot, Carlsbad, CA) before digitization by a multifunction data acquisition board (PCI-6052E, National Instruments, Austin, TX). The voltage outputs from our transimpedance amplifiers were oversampled at 150 kHz, and a number of points equal in duration to our integration time were collected and averaged for each sampled point. Device control and data acquisition were performed by custom-written software using the LabVIEW visual programming environment (National Instruments) (Fisher 2007
).
Choice of potentiometric probe for optical recording
Di-3-ANEPPDHQ, a chimeric product of di-8-ANEPPS (Bedlack et al. 1992
) and RH-795 (Grinvald et al. 1994
), has been shown to be less phototoxic than the former and less prone to internalization than the latter (Obaid et al. 2004
). Since phototoxicity compromises the physiology of the tissue under study, and internalization increases background fluorescence, thereby diminishing the S:N, the properties of the di-n-ANEPPDHQ dyes (Obaid et al. 2004
) are among the most favorable of the potentiometric probes currently available. Di-3-ANEPPDHQ and di-4-ANEPPDHQ have been recently licensed to Invitrogen and are now commercially available (catalog numbers D36801 and D36802, respectively).
One- and two-photon spectroscopy
Absorption and emission spectra for di-3-ANEPPDHQ were measured using octanol as a solvent to approximate the environment of the dye when bound to membrane (Sims et al. 1974
). Final sample concentration was
100 µM. Absorption measurements were performed on a spectrophotometer (Ocean Optics, Dunedin, FL), and excitation and emission spectrum measurements were performed on a spectrofluorometer (DeltaRAM, Photon Technology International). Two-photon excitation cross sections (
2 vs.
) were measured for the di-octyl analog of di-3-ANEPPDHQ in octanol as described previously (Fisher et al. 2005
).
Neurohypophysis preparation
The neurointermediate lobe of the pituitary, that is made up of the pars nervosa (neurohypophysis) and pars intermedia, was obtained (n = 6), as previously described (Salzberg et al. 1985
), from 30- to 60-day-old female CD-1 mice that had been anesthetized by CO2 inhalation and decapitated in accordance with institutional guidelines. Briefly, the mouse head was pinned to the bottom of a Sylgard-lined dissection dish, and, following removal of the skin, the skull was opened along the dorsal midline and the bone removed bilaterally. Cutting the optic and olfactory nerves allowed caudal deflection of the brain and rupture of the infundibular stalk, leaving the infundibular stump together with the entire pituitary gland attached to the base of the skull. The gland, gently removed using fine forceps and iridectomy scissors, was transferred to a dish containing oxygenated mouse Ringer solution (in mM: 154 NaCl, 5.6 KCl, 1.0 MgCl2, 2.2 CaCl2, 10 glucose, and 20 HEPES, adjusted to pH 7.4 with NaOH) where the anterior pituitary (pars anterior) was separated from the neurointermediate lobe and discarded. The pars intermedia, which consists of a delicate lacework of cells supporting the neurohypophysis, provided a convenient border through which the specimen could be pinned to the Sylgard bottom of a small petri dish (see
Fig. 2A) while preserving the integrity of the neurohypophysis. The tissue was stained by bathing it for 40 min in a Ringer solution containing 50 µg/ml of the potentiometric probe di-3-ANEPPDHQ together with 0.5% ethanol. (The easy solubility of this dye in ethanol rendered the use of DMSO or any surfactants, including Pluronics or PEG, unnecessary. In our experience, 0.5% ethanol is much less perturbing to the neurohypophysis than these other agents.) After staining, the neurohypophysis was washed in mouse Ringer solution, which was replaced with newly oxygenated medium every 20–30 min to preserve tissue viability.
|
|
150 µm below the surface of the tissue. The focusing routine consisted of advancing the adjustable stage upward until two-photon fluorescence was detected on the surface of the preparation (either by monitoring photomultiplier voltage or by viewing 2-photon fluorescence by eye). Repeated experiments with stained preparations revealed that locating the preparation surface by eye was consistently accurate to within
50 µm. After obtaining a surface image of the preparation, the objective was translated downward
150 µm using the objective translator, which has submicrometer precision and repeatability. Although optical signals could be recorded in response to stimulation from depths between the surface and
300 µm below the surface, a depth of
150 µm below the surface was used as the standard imaging depth to keep that factor as constant as possible while varying the experimental parameters of excitation and emission wavelengths. This depth was chosen because it offered reliable tissue viability, avoiding potentially damaged surface tissue, while incurring little blurring caused by scattering. One- and two-photon action spectra
One-photon action spectrum measurements were generated from the
F/F responses (n = 10) to electrical stimulation of the intact neurohypophysis at 18 different wavelengths. Excitation wavelength dependence was measured by varying excitation filters (band-pass) and comparing the peak fluorescence (
F/F) responses to APs. Because the peak absorption and emission wavelengths for this dye were narrowly shifted using one-photon excitation compared with two-photon excitation, several dichroic mirror and filter combinations were used to study the excitation wavelength range from 360 to 580 nm. The excitation filters were narrow (10-nm FWHM) band-pass filters centered at each excitation wavelength shown in Fig. 4A. The dichroic mirror and emission filter combinations were as follows: from 360 to 440 nm, R < 450 nm dichroic and Schott OG455 glass emission filter; from 460 to 510 nm, R < 515 nm dichroic and Schott OG530 glass emission filter; from 520 to 580 nm excitation, R < 585 nm dichroic and Schott OG590 glass emission filter. Because the system excitation and collection efficiency varied with different filter sets, we measured
F/F instead of S:N as a function of excitation wavelength.
|
F/F peak (see Fig. 4A) were beyond the tuning range of our mode-locked Ti:Al2O3 laser. The shape of the one-photon action spectrum (Fig. 4A) in the wavelength range that corresponds to our two-photon excitation (i.e., approximately one half of the 2-photon excitation wavelength) predicts that the sign of the two-photon signal should be positive for membrane depolarization. This was confirmed (Fig. 6),because the AP is registered optically as an increase in fluorescence.
|
Two-photon imaging was performed with the custom-built microscope shown in Fig. 1 and consisted of short line-scans (typically
3 points along a line
0.5 µm long, as in Fig. 6A) integrating for
100 µs at each point. Each point in the raw data traces (e.g., Fig. 6D, unfiltered) represents the average of the approximately three points along the short red line segment. Electrical stimuli (
1–10 mA) were delivered within
50 ms of the beginning of the line-scan to minimize the effect of photobleaching. When multiple stimuli were delivered (maximum 16 stimuli), the stimulation frequency was 16 Hz. Otherwise, all trials were single-stimulus experiments. The records were 100 ms long with 5-s wait time between trials for single stimuli and 500–1,000 ms long with 1-min wait time between trials for trains of stimuli. Averaging, when needed, was performed off-line using exclusively responses within a single multi-stimulus trial. When signal averaging was used (Fig. 6C), the number of averaged spikes is indicated in the corresponding figure legend. For each point scanned in functional recordings, fractional fluorescence (
F/F) was calculated by subtracting the baseline, F, from each point in the recording, and dividing by this same background. F, in our experiments, was the average of the first 10 data points in the recording. Functional recording data were filtered as indicated in the corresponding figure legends.
|
|
RESULTS |
|---|
|
99% of the excitable membrane in the tissue (Nordmann 1977Choice of optimal two-photon excitation wavelength for di-3-ANEPPDHQ
To optimize the use of di-3-ANEPPDHQ for optical measurements of membrane potential, we first characterized the dye spectral properties in octanol, a solvent that simulates the membrane environment (Fig. 3). Figure 3A shows absorption- and fluorescence-excitation spectra, whereas Fig. 3B shows fluorescence emission with excitation at 425 or 530 nm. It is evident that fluorescence emission peaks beyond 560 nm.
|
F/F]/
Vm vs.
, where Vm is the transmembrane potential) varies widely between phyla and even species (Ross and Reichardt 1979
100 mV; Jackson et al. 1991
F/F at
525 nm (the wavelength customarily used for 1-photon excitation measurements of Vm) and a smaller increase at
400 nm. The sign change occurs because the absorption spectrum of di-3-ANEPPDHQ is blue-shifted in response to depolarization. Thus
F, in response to excitation near the absorption peak, can be either positive or negative, depending on whether the excitation wavelength is on its rising or falling edge.
Optical recording of membrane potential using two-photon fluorescence requires that the molecular probe exhibits a significant (
F/F)/
Vm and possesses an appreciable two-photon excitation cross-section (
2). We have previously measured
2 (Fisher et al. 2005
) as a function of excitation wavelength (
), for the di-octyl analog of di-3-ANEPPDHQ, and this is shown in Fig. 4B. Note the significant cross-section between 780 and 870 nm and note further the rise to 920 nm and, presumably, beyond. The combination of the one-photon fluorescence action spectrum (Fig. 4A) and the two-photon excitation cross-section (Fig. 4B) might suggest that excitation at 920 nm or beyond would be optimal for two-photon fluorescence measurement of membrane potential in the mouse neurohypophysis. However, there are two important caveats. First, the maximum value of the measured
2 coincides with twice the zero-crossing wavelength (460 nm) of the action spectrum, at which (
F/F)/
Vm is near zero (Fig. 4A). Second, H2O absorption in tissue rises dramatically between 920 and 970 nm and does not fall off until
1.025 µm (Collins 1925
), beyond the range of our mode-locked Ti:Al2O3 laser. To circumvent these obstacles within the limitations imposed by our light source, we decided to exploit the "blue" peak of the dye's action spectrum and excite di-3-ANEPPDHQ at 850 nm. The two-photon sensitivity of the measurement should depend on the derivative of the cross-section with respect to wavelength (Pons and Mertz 2006
), so exciting beyond the 920-nm peak, where the cross-section decreases could, in principle, make sense. However, we were unable to record reliably voltage signals at 955 nm. (As an aside, longer wavelength Ti:Al2O3 lasers are now available that reach the local minimum in H2O absorption at
1.06 µm, exactly twice the wavelength of the "red" peak in the 1-photon action spectrum.)
Figure 5A compares the
F/F of the two-photon excitation action spectrum of di-3-ANEPPDHQ in response to an AP in the neurohypophysis (top x-axis), with the
F/F at the "blue" peak of the one-photon action spectrum (bottom x-axis). There appears to be a peak in the
F/F response at 850 nm, but it is not statistically significant. Furthermore, the main statistical trend is a downward slope toward 900 nm. S:N is a more relevant experimental parameter than
F/F. With that in mind, the same two-photon excitation data are replotted in Fig. 5B as S:N. While there was significant statistical variability in single-trial two-photon measurements, a local maximum at 850 nm could be identified. Assuming constant laser intensity over the excitation spectrum, Fig. 5B is consistent with both Figs. 5A and 4B, as the signal-to-noise is proportional to
F/F and the square root of the two-photon excitation cross-section,
2, i.e., S/N
(
F·
)/F. The statistical variability in the response is reflected in the error bars in the two-photon action spectra (Fig. 5). Error bars on the one-photon action spectra are significantly smaller, reflecting the fact that those optical responses were measured from the entire neurohypophysis and, as such, represent the synchronous firing of hundreds of thousands, if not millions of terminals.
|
When di-3-ANEPPDHQ is excited at 425 nm, the one-photon fluorescence emission spectrum, measured in octanol, has a peak at
613 nm and extends much further into the red (Fig. 3B). In the stained neurohypophysis, however, with two-photon excitation between 800 and 900 nm, we determined empirically that the optimal wavelength for detecting voltage-dependent changes in fluorescence was
530 nm. Indeed, band-pass filters centered at longer wavelengths, as well as long-pass filters, yielded measurements with significantly smaller S:N (Fig. 5B).
Two-photon optical recording of APs
The results shown in Fig. 6 establish the two-photon technique as a valuable tool for optical recording of electrically evoked APs from individual, or, at most, a very few (see DISCUSSION) nerve terminals in situ, using a voltage-sensitive dye. Figure 6A shows the two-photon background image and the recording location from which the data in Fig. 6, B and C, were obtained. Functional imaging consisted of short line-scans (typically
3 points along a line
0.5 µm long, as seen in the inset of Fig. 6A) integrating for
100 µs at each location. Each point in the final data traces (e.g., Fig. 6, B and D) represents the average of approximately three points along the short (red) line (line scanning was used instead of static single-point "beam-parking" to minimize phototoxicity).
Figure 6B shows a train of APs (16 stimuli delivered at 16 Hz) recorded optically, in a single trial, from the location indicated in red in Fig. 6A. Figure 6C shows the spike-average of the 16 optical responses from the single trial shown in Fig. 6B. The characteristic shape of the neurohypophysial AP (Gainer et al. 1986
; Muschol et al. 2003
), including its afterhyperpolarization, is clearly visible. Figure 6D shows a single-trial recording of an AP from an individual nerve terminal (or, possibly, as many as 4–6 nerve terminals), unfiltered, on the left, and filtered, on the right, and exemplifies the high S:N and large
F/F (
10%/100 mV) achievable with this technology. Single-trial responses were 2–3 ms wide at half-maximum, and averaged responses were often 4–5 ms wide. This broadening is likely caused by physiological jitter in the timing of the response, since the recordings represent the activity of just a single, or, at most, a very few nerve terminals. As such, these results are the first demonstration of functional two-photon recording with potentiometric probes of electrically evoked APs from micrometer-scale nerve terminals in situ.
|
|
DISCUSSION |
|---|
|
1 µm patches of neuronal membrane on small nerve terminals in an intact preparation. Indeed, the special morphology of the mammalian neurohypophysis (Nordmann 1977
Although experimental measurement of the two-photon focal excitation volume in situ is technically challenging, the lateral and axial dimensions of the focal volume can be calculated from high numerical-aperture formulations of the illumination point spread function (IPSF) (Richards and Wolf 1959
), which is the three-dimensional illumination intensity distribution. Because two-photon excitation is proportional to the square of the excitation intensity, measures of the lateral and axial widths of IPSF2 such as FWHM and 1/e width of Gaussian fits to the IPSF2 function are effective measures of lateral and axial resolution (Zipfel et al. 2003
). For the case of our overfilled 0.95 NA objective at 850-nm excitation, assuming the refractive index of the sample to be
1.4, the lateral and axial 1/e diameters (>FWHM) are
0.4 and
1.8 µm, respectively.
Scattering caused by heterogeneities in the tissue refractive index affects both the excitation and fluorescence photon trajectories. In single-point scanning two-photon microscopy, all collected fluorescence is attributed to the focal volume, thus scattering of the fluorescence signal does not affect spatial resolution. Scattering of the illumination photons, however, can cause a portion of the illumination to be deviated from the focal volume, and thus can change the effective focal volume. Because the excitation power used in these experiments was moderate (
10–20 mW) and the two-photon absorption cross-section of the dye was relatively small at the 850-nm excitation wavelength (<10 GM), it is unlikely that scattered (i.e., nonballistic) excitation photons seriously expanded the focal volume and decreased spatial resolution. Nevertheless, it is certainly reasonable that scattering could have somewhat expanded the focal volume from which the optical recordings were made. Assuming that scattering doubled the 1/e diameters of the IPSF2, the axial and spatial resolutions would be
0.8 and
3.6 µm, respectively. If the line scan excitation point intersected an
1-µm nerve terminal at the interface with another nerve terminal, these expanded resolutions would integrate the virtually synchronous activity of approximately six nerve terminals, as an upper limit. Since care was taken to direct the line scan in the center of nerve terminal fluorescence profiles, however, it is more likely that three or fewer nerve terminals contributed to the fluorescence signal. Finally, assuming that scattering did not lead to such a drastic decrease in axial resolution, it is most probable that the signals shown in Fig. 6 have their origin in the stained membrane of an individual nerve terminal. However, definitive proof that the optical signal reflects the activity of only a single terminal must await further experiments.
Interestingly, two-photon excitation yielded fractional fluorescence changes
5 times larger than one-photon excited
F/F at corresponding (
2Photon/2) excitation wavelengths. Such increases in
F/F following two-photon excitation are consistent with reported measurements in cultured neurons stained with a hemicyanine potentiometric probe (Kuhn et al. 2004
). While this agreement may be fortuitous (since these measurements were made using excitation at opposite sides of the absorption spectrum), it is noteworthy, since the relatively poor S:N of one-photon measurements is caused in no small part by the small fractional fluorescence changes exhibited in this mode, and, in light of our study (and that of Kuhn et al. 2004
), it seems that two-photon excitation recordings may not be limited in this way.
Although our present apparatus did not permit excitation beyond 960 nm, we recognize that infrared functional two-photon imaging of membrane potential offers exciting possibilities that must be pursued. Obvious advantages include greater penetration into highly scattering tissue and diminished likelihood of photodynamic damage. Previous evidence (our Fig. 4B; Fisher et al. 2005
) suggests that several voltage-sensitive dyes have the peak of their two-photon absorption cross-sections beyond 960 nm. Since the one-photon action spectrum clearly indicates that larger signals can be expected for two-photon excitation between 1.000 and 1.150 µm, measurement of
2 at these longer wavelengths, using newly available mode-locked Ti:Al2O3 lasers, should permit optimization of the excitation parameters, thereby increasing even further the sensitivity and, by virtue of increased sensitivity, the spatial resolution of the technique.
The large
F/F signals with high S:N described here are particularly relevant to neuroscience and neurology, where the mammalian brain is of primary interest. Even more significant is the fact that this technique uses epifluorescence detection of the optical signals, i.e., detection in the backward direction, a measurement modality that will permit noninvasive recording of electrical activity from intact brain. Furthermore, the fact that the molecular probes are applied in the bath, rather than by intracellular injection of a single neuron (Antic and Zecevic 1995
; Djurisic et al. 2004
; Grinvald et al. 1987
), and that the recording sites are inherently very small by virtue of the properties of the two-photon excitation, means that multiplexing schemes, including laser scanning and multi-beamlet excitation (Crepel et al. 2007
; Fisher 2007
; Kurtz et al. 2006
), will permit optical recording from nerve terminals or dendrites from multiple cells virtually simultaneously.
|
|
GRANTS |
|---|
|
|
|
ACKNOWLEDGMENTS |
|---|
|
Present address of J.A.N. Fisher: Howard Hughes Medical Institute and Laboratory of Sensory Neuroscience, The Rockefeller University, New York, NY 10021.
|
|
FOOTNOTES |
|---|
Address for reprint requests and other correspondence: B. M. Salzberg, Dept. of Neuroscience, Univ. of Pennsylvania School of Medicine, 234 Stemmler Hall, Philadelphia, PA 19104-6074 (E-mail: bmsalzbe{at}mail.med.upenn.edu)
|
|
REFERENCES |
|---|
|
Antic S, Zecevic D. Optical signals from neurons with internally applied voltage-sensitive dyes. J Neurosci 15: 1392–1405, 1995.[Abstract]
Bedlack RS Jr, Wei M, Loew LM. Localized membrane depolarizations and localized calcium influx during electric field-guided neurite growth. Neuron 9: 393–403, 1992.[CrossRef][Web of Science][Medline]
Campagnola PJ, Wei MD, Lewis A, Loew LM. High-resolution nonlinear optical imaging of live cells by second harmonic generation. Biophys J 77: 3341–3349, 1999.[Web of Science][Medline]
Civillico EF, Contreras D. Integration of evoked responses in supragranular cortex studied with optical recordings in vivo. J Neurophysiol 96: 336–351, 2006.
Cohen LB, Salzberg BM. Optical measurement of membrane potential. Rev Physiol Biochem Pharmacol 83: 35–88, 1978.[Medline]
Collins JR. Change in the infra-red absorption spectrum of water with temperature. Physiol Rev 26: 771–779, 1925.[CrossRef]
Contreras D, Llinás R. Voltage-sensitive dye imaging of neocortical spatiotemporal dynamics to afferent activation frequency. J Neurosci 21: 9403–9413, 2001.
Crepel V, Aronov D, Jorquera I, Represa A, Ben-Ari Y, Cossart R. A parturition-associated nonsynaptic coherent activity pattern in the developing hippocampus. Neuron 54: 105–120, 2007.[CrossRef][Web of Science][Medline]
Denk W, Strickler JH, Webb WW. Two-photon laser scanning fluorescence microscopy. Science 248: 73–76, 1990.
Djurisic M, Antic S, Chen WR, Zecevic D. Voltage imaging from dendrites of mitral cells: EPSP attenuation and spike trigger zones. J Neurosci 24: 6703–6714, 2004.
Douglas WW. How do neurones secrete peptides? Exocytosis and its consequences, including "synaptic vesicle" formation, in the hypothalamo-neurohypophyseal system. Prog Brain Res 39: 21–39, 1973.[Medline]
Fisher JAN. Linear and Non-Linear Fluorescence Imaging of Neuronal Activity (PhD thesis). Department of Physics and Astronomy, University of Pennsylvania, Philadelphia, PA, 2007.
Fisher JAN, Civillico EF, Contreras D, Yodh AG. in vivo fluorescence microscopy of neuronal activity in three dimensions by use of voltage-sensitive dyes. Opt Lett 29: 71–73, 2004.[CrossRef][Web of Science][Medline]
Fisher JAN, Salzberg BM, Yodh AG. Near infrared two-photon excitation cross-sections of voltage-sensitive dyes. J Neurosci Methods 148: 94–102, 2005.[CrossRef][Web of Science][Medline]
Gainer H, Wolfe SA Jr, Obaid AL, Salzberg BM. Action potentials and frequency-dependent secretion in the mouse neurohypophysis. Neuroendocrinology 43: 557–563, 1986.[Web of Science][Medline]
Grinvald A, Lieke EE, Frostig RD, Hildesheim R. Cortical point-spread function and long-range lateral interactions revealed by real-time optical imaging of macaque monkey primary visual cortex. J Neurosci 14: 2545–2568, 1994.[Abstract]
Grinvald A, Salzberg BM, Lev-Ram V, Hildesheim R. Optical recording of synaptic potentials from processes of single neurons using intracellular potentiometric dyes. Biophys J 51: 643–651, 1987.[Web of Science][Medline]
Heer F, Hafizovic S, Ugniwenko T, Frey U, Franks W, Perriard E, Perriard JC, Blau A, Ziegler C, Hierlemann A. Single-chip microelectronic system to interface with living cells. Biosens Bioelectron 22: 2546–2553, 2007.[CrossRef][Web of Science][Medline]
Hellwarth R, Christensen P. Nonlinear optical microscope examination of structures in polycrystalline ZnSe. Opt Commun 12: 318–322, 1974.[CrossRef]
Jackson MB, Konnerth A, Augustine GJ. Action potential broadening and frequency-dependent facilitation of calcium signals in pituitary nerve terminals. Proc Natl Acad Sci USA 88: 380–384, 1991.
Kuhn B, Fromherz P, Denk W. High sensitivity of Stark-shift voltage-sensing dyes by one- or two-photon excitation near the red spectral edge. Biophys J 87: 631–639, 2004.[CrossRef][Web of Science][Medline]
Kurtz R, Fricke M, Kalb J, Tinnefeld P, Sauer M. Application of multiline two-photon microscopy to functional in vivo imaging. J Neurosci Methods 151: 276–286, 2006.[CrossRef][Web of Science][Medline]
Millard AC, Jin L, Wuskell JP, Boudreau DM, Lewis A, Loew LM. Wavelength- and time-dependence of potentiometric non-linear optical signals from styryl dyes. J Membr Biol 208: 103–111, 2005.[CrossRef][Web of Science][Medline]
Moreaux L, Pons T, Dambrin V, Blanchard-Desce M, Mertz J. Electro-optic response of second-harmonic generation membrane potential sensors. Opt Lett 28: 625–627, 2003.[Web of Science][Medline]
Muschol M, Kosterin P, Ichikawa M, Salzberg BM. Activity-dependent depression of excitability and calcium transients in the neurohypophysis suggests a model of "stuttering conduction". J Neurosci 23: 11352–11362, 2003.
Nordmann JJ. Ultrastructural morphometry of the rat neurohypophysis. J Anat 123: 213–218, 1977.[Web of Science][Medline]
Nuriya M, Jiang J, Nemet B, Eisenthal KB, Yuste R. Imaging membrane potential in dendritic spines. Proc Natl Acad Sci USA 103: 786–790, 2006.
Obaid AL, Koyano T, Lindstrom J, Sakai T, Salzberg BM. Spatiotemporal patterns of activity in an intact mammalian network with single-cell resolution: optical studies of nicotinic activity in an enteric plexus. J Neurosci 19: 3073–3093, 1999.
Obaid AL, Loew LM, Wuskell JP, Salzberg BM. Novel naphthyl styryl-pyridinium potentiometric dyes offer advantages for neural network analysis. J Neurosci Methods 134: 179–190, 2004.[CrossRef][Web of Science][Medline]
Obaid AL, Nelson ME, Lindstrom J, Salzberg BM. Optical studies of nicotinic acetylcholine receptor subtypes in the guinea-pig enteric nervous system. J Exp Biol 208: 2981–3001, 2005.
Patolsky F, Timko BP, Yu G, Fang Y, Greytak AB, Zheng G, Lieber CM. Detection, stimulation, and inhibition of neuronal signals with high-density nanowire transistor arrays. Science 313: 1100–1104, 2006.
Petersen CC, Grinvald A, Sakmann B. Spatiotemporal dynamics of sensory responses in layer 2/3 of rat barrel cortex measured in vivo by voltage-sensitive dye imaging combined with whole-cell voltage recordings and neuron reconstructions. J Neurosci 23: 1298–1309, 2003.
Pons T, Mertz J. Membrane potential detection with second-harmonic generation and two-photon excited fluorescence: a theoretical comparison. Opt Commun 258: 203–209, 2006.[CrossRef]
Richards B, Wolf E. Electromagnetic diffraction in optical systems II. Structure of the image field in an aplanatic system. Proc R Soc London A 253: 358–379, 1959.
Ross WN, Reichardt LF. Species-specific effects on the optical signals of voltage-sensitive dyes. J Membr Biol 48: 343–356, 1979.[CrossRef][Web of Science][Medline]
Sacconi L, Dombeck DA, Webb WW. Overcoming photodamage in second-harmonic generation microscopy: real-time optical recording of neuronal action potentials. Proc Natl Acad Sci USA 103: 3124–3129, 2006.
Salzberg BM. Optical recording of electrical activity in neurons using molecular probes. In: Current Methods in Cellular Neurobiology, edited by Barker J, McKelvy J. New York: John Wiley, 1983, p. 139–187.
Salzberg BM, Obaid AL, Bezanilla F. Microsecond response of a voltage-sensitive merocyanine dye: fast voltage-clamp measurements on squid giant axon. Jpn J Physiol 43: S37–S41, 1993.[Web of Science][Medline]
Salzberg BM, Obaid AL, Gainer H. Large and rapid changes in light scattering accompany secretion by nerve terminals in the mammalian neurohypophysis. J Gen Physiol 86: 395–411, 1985.
Sheppard CJR, Kompfner R, Gannaway J, Walsh D. The scanning harmonic optical microscope. IEEE J Quant Electron QE13: 100D, 1977.
Sims PJ, Waggoner AS, Wang CH, Hoffman JF. Studies on the mechanism by which cyanine dyes measure membrane potential in red blood cells and phosphatidylcholine vesicles. Biochemistry 13: 3315–3330, 1974.[CrossRef][Web of Science][Medline]
Sjulson L, Miesenboeck G. Optical recording of action potentials and other discrete physiological events: a perspective from signal detection theory. Physiology (Bethesda) 22: 47–55, 2007.[CrossRef][Medline]
Zipfel WR, Williams RM, Webb WW. Nonlinear magic: multiphoton microscopy in the biosciences. Nat Biotechnol 21: 1369–1377, 2003.[CrossRef][Medline]
This article has been cited by other articles:
![]() |
J. Bradley, R. Luo, T. S. Otis, and D. A. DiGregorio Submillisecond Optical Reporting of Membrane Potential In Situ Using a Neuronal Tracer Dye J. Neurosci., July 22, 2009; 29(29): 9197 - 9209. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Sjulson and G. Miesenbock Rational Optimization and Imaging In Vivo of a Genetically Encoded Optical Voltage Reporter J. Neurosci., May 21, 2008; 28(21): 5582 - 5593. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| Visit Other APS Journals Online |