1 The Salk Institute for Biological Studies, San Diego, CA, USA
2 Faculty of Physics and Applied Computer Science, AGH University of Science and Technology, Krakow, Poland
3 University of California, Santa Cruz, Santa Cruz, CA, USA
* To whom correspondence should be addressed. E-mail: ej{at}salk.edu.
Existing epiretinal implants for the blind are designed to electrically stimulate large groups of surviving retinal neurons using a small number of electrodes with diameters of several hundred µm. To increase the spatial resolution of artificial sight, electrodes much smaller than those currently in use are desirable. In this study we stimulated and recorded ganglion cells in isolated pieces of rat, guinea pig, and monkey retina. We utilized micro-fabricated hexagonal arrays of 61 platinum disk electrodes with diameters between 6 and 25 µm, spaced 60 µm apart. Charge-balanced current pulses evoked one or two spikes at latencies as short as 0.2 ms, and typically only one or a few recorded ganglion cells were stimulated. Application of several synaptic blockers did not abolish the evoked responses, implying direct activation of ganglion cells. Threshold charge densities were typically below 0.1 mC/cm2 for a pulse duration of 100 µs, corresponding to charge thresholds of less than 100 pC. Stimulation remained effective after several hours and at high frequencies. To demonstrate that closely spaced electrodes can elicit independent ganglion cell responses, we utilized the multi-electrode array to stimulate several nearby ganglion cells simultaneously. From these data we conclude that electrical stimulation of mammalian retina with small-diameter electrode arrays is achievable and can provide high temporal and spatial precision at low charge densities. We review previous epiretinal stimulation studies and discuss our results in the context of 32 other publications, comparing threshold parameters and safety limits.
This article has been cited by other articles:
![]() |
D. J. Margolis, G. Newkirk, T. Euler, and P. B. Detwiler Functional Stability of Retinal Ganglion Cells after Degeneration-Induced Changes in Synaptic Input J. Neurosci., June 18, 2008; 28(25): 6526 - 6536. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. de Balthasar, S. Patel, A. Roy, R. Freda, S. Greenwald, A. Horsager, M. Mahadevappa, D. Yanai, M. J. McMahon, M. S. Humayun, et al. Factors Affecting Perceptual Thresholds in Epiretinal Prostheses Invest. Ophthalmol. Vis. Sci., June 1, 2008; 49(6): 2303 - 2314. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Sekirnjak, P. Hottowy, A. Sher, W. Dabrowski, A. M. Litke, and E. J. Chichilnisky High-Resolution Electrical Stimulation of Primate Retina for Epiretinal Implant Design J. Neurosci., April 23, 2008; 28(17): 4446 - 4456. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. J. Tehovnik and W. M. Slocum What Delay Fields Tell Us About Striate Cortex J Neurophysiol, August 1, 2007; 98(2): 559 - 576. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Niebur Generation of synthetic spike trains with defined pairwise correlations. Neural Comput., July 1, 2007; 19(7): 1720 - 1738. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. DeMarco Jr, G. L. Yarbrough, C. W. Yee, G. Y. McLean, B. T. Sagdullaev, S. L. Ball, and M. A. McCall Stimulation via a Subretinally Placed Prosthetic Elicits Central Activity and Induces a Trophic Effect on Visual Responses Invest. Ophthalmol. Vis. Sci., February 1, 2007; 48(2): 916 - 926. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. J. Tehovnik, A. S. Tolias, F. Sultan, W. M. Slocum, and N. K. Logothetis Direct and Indirect Activation of Cortical Neurons by Electrical Microstimulation J Neurophysiol, August 1, 2006; 96(2): 512 - 521. [Abstract] [Full Text] [PDF] |
||||